510(k) Report

510(k) · Cleared

Recalled · Class II · 2019

PulsioFlex Monitoring System

K172259510(k) TraditionalCleared Jan 18, 2018175-day reviewClass II21 CFR 870.1435Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Pulsion Medical Systems SE · DXG Computer, Diagnostic, Pre-Programmed, Single-Function

Cleared on Jan 18, 2018 as substantially equivalent to K122121 (Pulsion Medical Systems SE, 2012) after a 175-day review. Its family tree reaches back 1 generation to 2012. 3 later devices descend from it. It has been recalled once.

Decided
Jan 18, 2018
Review
175 days
Class
Class II
Product code
DXG
Recalls
1
Predicates
1

Its longest line of descent runs back 1 generation to K122121 (2012). 3 later devices descend from it, 2 citing it directly.

Descendants, by year cleared 2020: 1 descendant cleared2019: 1 descendant cleared2023: 1 descendant clearedK122121 · Pulsioflex Monitoring System with Picco Module · Pulsion Medical Systems SE · 20122010201520202025
Descendants, by generation 1 generation forward: 2 descendants2 generations forward: 1 descendant K122121 · Pulsioflex Monitoring System with Picco Module · Pulsion Medical Systems SE · 20120
  1. Root · 1 generation backK122121 Pulsioflex Monitoring System with Picco ModulePulsion Medical Systems SE · 2012
  2. This deviceK172259 PulsioFlex Monitoring SystemPulsion Medical Systems SE · 2018
  3. Descendant · cites it directly · 1 of its ownK183521 CNAP MonitorCnsystems Medizintechnik GmbH · 2019
  4. Descendant · 2 generations on · 0 of its ownK223332 Task Force CORECnsystems Medizintechnik GmbH · 2023
  5. Descendant · cites it directly · 0 of its ownK192169 PulsioFlex Monitoring System with ProAQT SensorPulsion Medical Systems SE · 2020

Most influential descendants

Showing 3 of 3 · See all
  • K183521CNAP MonitorCnsystems Medizintechnik GmbH · 2019 · cites it directly1descendant
  • K223332Task Force CORECnsystems Medizintechnik GmbH · 2023 · 2 generations on0descendants
  • K192169PulsioFlex Monitoring System with ProAQT SensorPulsion Medical Systems SE · 2020 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K122121Pulsioflex Monitoring System with Picco ModulePulsion Medical Systems SE2012—
0 (this device)K172259PulsioFlex Monitoring SystemPulsion Medical Systems SE2018—
+1K183521CNAP MonitorCnsystems Medizintechnik GmbH2019—
+2K223332Task Force CORECnsystems Medizintechnik GmbH2023—
+1K192169PulsioFlex Monitoring System with ProAQT SensorPulsion Medical Systems SE2020—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201910
202010
202310
FIG. 1 — Lineage of K172259, 2012–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DXG? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DXG device

  1. #1 K253092 Argos Infinity (Rev. 1.0)score 11.1 · 2 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K122121Pulsioflex Monitoring System with Picco Module
Summary page 4
…computer. Classification: According to 21 CFR 870.1435 the device classification is Class II, Product code DXG. Predicate Device: K122121, PulsioFlex Monitoring System with PiCCO Module…
Pulsion Medical Systems SE2012—

What descends from it? (3)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K183521CNAP MonitorCnsystems Medizintechnik GmbH20191
K192169PulsioFlex Monitoring System with ProAQT SensorPulsion Medical Systems SE20200

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 5, 2019Class IIZ-2077-2019Software designThe monitor displays an error message . The error message states "internal error restart or service".

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has no recall.
  • Of its descendants, K183521 (Cnsystems Medizintechnik GmbH, 2019) has gone furthest, with 1 descendant of its own.

How does its category compare?

All DXG devices

DXG Computer, Diagnostic, Pre-Programmed, Single-Function: FDA decisions, review times, recalls and adverse event reports over 20 years. 2018, this device’s year, is highlighted.

FDA decisions per year · DXG

1in 2025

2007: 12008: 52010: 42011: 32012: 22013: 32014: 62015: 12016: 12017: 12018: 22019: 22020: 12021: 12023: 12025: 12026 (partial year): 1
2007high 6 (2014) · low 0 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DXG

290in 2014

2008: 118 days median2010: 282 days median2011: 104 days median2013: 236 days median2014: 290 days median
2007high 290 (2014) · low 104 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DXG

2in 2020

2007: 12008: 12009: 62010: 12020: 2
2007high 6 (2009) · low 0 (2011)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DXG

8in 2025

2007: 432008: 372009: 82010: 172011: 72012: 292013: 232014: 442015: 342016: 672017: 622018: 832019: 812020: 492021: 292022: 442023: 122024: 212025: 8
2007high 83 (2018) · low 7 (2011)2025

MAUDE reports naming product code DXG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—143
20085118137
20090—68
20104282117
2011310407
20122—029
20133236023
20146290044
20151—034
20161—067
20171—062
20182—083
20192—081
20201—249
20211—029
20220—044
20231—012
20240—021
20251—08
20261—0—

What is it indicated for?

The PulsioFlex Monitoring System is a diagnostic aid for the measurement and monitoring of blood pressure, cardiopulmonary, circulatory and organ function variables. The PulsioFlex Monitoring System is indicated in patients where cardiovascular and circulatory volume status monitoring is necessary. If a patient’s biometric data are entered, the PulsioFlex monitor presents the derived parameters indexed. With the PiCCO Module cardiac output is determined…

Read the full statement (137 words)

The PulsioFlex Monitoring System is a diagnostic aid for the measurement and monitoring of blood pressure, cardiopulmonary, circulatory and organ function variables. The PulsioFlex Monitoring System is indicated in patients where cardiovascular and circulatory volume status monitoring is necessary. If a patient’s biometric data are entered, the PulsioFlex monitor presents the derived parameters indexed.

With the PiCCO Module cardiac output is determined both continuously through pulse contour analysis and intermittently through thermodilution technique. Both are used for the determination of other derived parameters.

With the CeVOX oximetry module connected to a compatible oximetry probe, the PulsioFlex Monitoring System measures continuous venous oxygen saturation to assess oxygen delivery and consumption.

The use of the PulsioFlex Monitoring System is indicated in patients where cardiovascular and organ monitoring is useful. This includes patients in surgical, medical, and other hospital units.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K172259: PulsioFlex Monitoring System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K172259

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026