510(k) Report

510(k) · Cleared

Cleared

Task Force CORE

K223332510(k) TraditionalCleared Apr 12, 2023163-day reviewClass II21 CFR 870.1130

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Cnsystems Medizintechnik GmbH · DXN System, Measurement, Blood-Pressure, Non-Invasive

Cleared on Apr 12, 2023 as substantially equivalent to K173916 (Finapres Medical Systems B.V., 2018) and 1 other predicate after a 163-day review. Its family tree reaches back 12 generations to 2000. 5 of the 106 devices in its family, counting itself, have a Class I recall.

Decided
Apr 12, 2023
Review
163 days
Class
Class II
Product code
DXN
Recalls
None
Predicates
2

Its longest line of descent runs back 12 generations to K000126 (2000). 105 ancestors in all, the oldest K822055 (1982). No later device cites it as a predicate yet. 5 of the 106 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year cleared 2018: 2 other ancestors2017: 1 other ancestor2015: 1 other ancestor2008: 2 other ancestors2014: 1 other ancestor2012: 1 other ancestor2010: 2 other ancestors, 1 with a Class I recall12007: 3 other ancestors2006: 3 other ancestors2003: 5 other ancestors2002: 6 other ancestors, 1 with a Class I recall11988: 2 other ancestors2011: 1 other ancestor2005: 2 other ancestors2001: 4 other ancestors2000: 2 other ancestors1994: 4 other ancestors1986: 1 other ancestor1982: 1 other ancestor2004: 4 other ancestors1999: 7 other ancestors1996: 7 other ancestors1995: 8 other ancestors1990: 4 other ancestors1998: 7 other ancestors, 2 with a Class I recall21985: 2 other ancestors1993: 1 other ancestor1991: 5 other ancestors1997: 2 other ancestors1992: 2 other ancestorsK000126 · Masimo Set Radical Pulse Oximeter with Satshare, Masimo Lnop-Adt-Adult/Pediatric Disposable Sensor, Masimo Lno-PDT-Pedia · Masimo Corp. · 2000K031330 · Modification to Masimo Set Radical Pulse Oximeter with Satshare and Lnop Series of Sensors and Cables · Masimo Corp. · 2003K033296 · Masimo Set Rad-5 Pulse Oximeter, Model Rad-5; Masimo Set Rad-5V Pulse Oximeter, Model Rad-5V · Masimo Corp. · 2004 · recalledK042536 · Masimo Set Rad 57 Pulse CO-Oximeter · Masimo Corporation · 2005 · recalled×2: 2 generations cleared close together, K053477 (2006), K061204 (2006) K080238 · Masimo Rainbow Set Radical 7 Pulse CO-Oximeter, Masimo Rainbow Set Rad 87 Pulse CO-Oximeter · Masimo Corporation · 2008 · recalledK101123 · Nexfin Model 2 · Bmeye B.V. · 2010K122381 · Nexfin Model 2 · Edwards Lifesciences, LLC · 2013K140312 · EV1000 Clinical Platform with Clearsight TM Finger Cuff or Clearsight TM System · Edwards Lifesciences, LLC · 2014 · recalledK160552 · EV1000 Clinical Platform with ClearSight Finger Cuff or ClearSight System, EV1000 Clinical Platform · Edwards Lifesciences, LLC · 2016 · Class I recallK183521 · CNAP Monitor · Cnsystems Medizintechnik GmbH · 201919801990200020102020
Other ancestors, by generation 1 generation back: 1 other ancestor2 generations back: 4 other ancestors3 generations back: 11 other ancestors, 1 with a Class I recall14 generations back: 14 other ancestors5 generations back: 17 other ancestors6 generations back: 12 other ancestors, 2 with a Class I recall27 generations back: 15 other ancestors, 1 with a Class I recall18 generations back: 17 other ancestors9 generations back: 2 other ancestors K183521 · CNAP Monitor · Cnsystems Medizintechnik GmbH · 2019K160552 · EV1000 Clinical Platform with ClearSight Finger Cuff or ClearSight System, EV1000 Clinical Platform · Edwards Lifesciences, LLC · 2016 · Class I recallK140312 · EV1000 Clinical Platform with Clearsight TM Finger Cuff or Clearsight TM System · Edwards Lifesciences, LLC · 2014 · recalledK122381 · Nexfin Model 2 · Edwards Lifesciences, LLC · 2013K101123 · Nexfin Model 2 · Bmeye B.V. · 2010K080238 · Masimo Rainbow Set Radical 7 Pulse CO-Oximeter, Masimo Rainbow Set Rad 87 Pulse CO-Oximeter · Masimo Corporation · 2008 · recalledK061204 · Masimo Set Radical 7 Pulse CO-Oximeter · Masimo Corporation · 2006K053477 · Masimo Set Rad 57 CM and M Pulse CO-Oximeters · Masimo Corporation · 2006K042536 · Masimo Set Rad 57 Pulse CO-Oximeter · Masimo Corporation · 2005 · recalledK033296 · Masimo Set Rad-5 Pulse Oximeter, Model Rad-5; Masimo Set Rad-5V Pulse Oximeter, Model Rad-5V · Masimo Corp. · 2004 · recalledK031330 · Modification to Masimo Set Radical Pulse Oximeter with Satshare and Lnop Series of Sensors and Cables · Masimo Corp. · 2003K000126 · Masimo Set Radical Pulse Oximeter with Satshare, Masimo Lnop-Adt-Adult/Pediatric Disposable Sensor, Masimo Lno-PDT-Pedia · Masimo Corp. · 2000−10−50
  1. Root · 12 generations backK000126 Masimo Set Radical Pulse Oximeter with Satshare, Masimo…Masimo Corp. · 2000
  2. 8 generations between
  3. 3 generations backK140312 EV1000 Clinical Platform with Clearsight TM Finger Cuff or…Edwards Lifesciences, LLC · 2014 · Recalled
  4. 2 generations backK160552 EV1000 Clinical Platform with ClearSight Finger Cuff or ClearSight…Edwards Lifesciences, LLC · 2016 · Recalled (Class I)
  5. 1 generation backK183521 CNAP MonitorCnsystems Medizintechnik GmbH · 2019
  6. This deviceK223332 Task Force CORECnsystems Medizintechnik GmbH · 2023
Show all 12 generations
  1. Root · 12 generations backK000126 Masimo Set Radical Pulse Oximeter with Satshare, Masimo…Masimo Corp. · 2000
  2. 11 generations backK031330 Modification to Masimo Set Radical Pulse Oximeter with Satshare and…Masimo Corp. · 2003
  3. 10 generations backK033296 Masimo Set Rad-5 Pulse Oximeter, Model Rad-5; Masimo Set Rad-5V…Masimo Corp. · 2004 · Recalled
  4. 9 generations backK042536 Masimo Set Rad 57 Pulse CO-OximeterMasimo Corporation · 2005 · Recalled
  5. 8 generations backK053477 Masimo Set Rad 57 CM and M Pulse CO-OximetersMasimo Corporation · 2006
  6. 7 generations backK061204 Masimo Set Radical 7 Pulse CO-OximeterMasimo Corporation · 2006
  7. 6 generations backK080238 Masimo Rainbow Set Radical 7 Pulse CO-Oximeter, Masimo Rainbow Set…Masimo Corporation · 2008 · Recalled
  8. 5 generations backK101123 Nexfin Model 2Bmeye B.V. · 2010
  9. 4 generations backK122381 Nexfin Model 2Edwards Lifesciences, LLC · 2013
  10. 3 generations backK140312 EV1000 Clinical Platform with Clearsight TM Finger Cuff or…Edwards Lifesciences, LLC · 2014 · Recalled
  11. 2 generations backK160552 EV1000 Clinical Platform with ClearSight Finger Cuff or ClearSight…Edwards Lifesciences, LLC · 2016 · Recalled (Class I)
  12. 1 generation backK183521 CNAP MonitorCnsystems Medizintechnik GmbH · 2019
  13. This deviceK223332 Task Force CORECnsystems Medizintechnik GmbH · 2023

Class I recalls in this family

View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−12K000126Masimo Set Radical Pulse Oximeter with Satshare, Masimo…Masimo Corp.2000—
−11K031330Modification to Masimo Set Radical Pulse Oximeter with Satshare and Lnop…Masimo Corp.2003—
−10K033296Masimo Set Rad-5 Pulse Oximeter, Model Rad-5; Masimo Set Rad-5V Pulse…Masimo Corp.20041
−9K042536Masimo Set Rad 57 Pulse CO-OximeterMasimo Corporation20051
−8K053477Masimo Set Rad 57 CM and M Pulse CO-OximetersMasimo Corporation2006—
−7K061204Masimo Set Radical 7 Pulse CO-OximeterMasimo Corporation2006—
−6K080238Masimo Rainbow Set Radical 7 Pulse CO-Oximeter, Masimo Rainbow Set Rad 87…Masimo Corporation20081
−5K101123Nexfin Model 2Bmeye B.V.2010—
−4K122381Nexfin Model 2Edwards Lifesciences, LLC2013—
−3K140312EV1000 Clinical Platform with Clearsight TM Finger Cuff or Clearsight TM SystemEdwards Lifesciences, LLC20141
−2K160552EV1000 Clinical Platform with ClearSight Finger Cuff or ClearSight System…Edwards Lifesciences, LLC2016Class I
−1K183521CNAP MonitorCnsystems Medizintechnik GmbH2019—
0 (this device)K223332Task Force CORECnsystems Medizintechnik GmbH2023—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198210
198520
198610
198820
199040
199150
199220
199310
199440
199580
199670
199720
199872
199970
200020
200140
200261
200350
200440
200520
200630
200730
200820
201021
201110
201210
201410
201510
201710
201820
FIG. 1 — Lineage of K223332, 2000–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

5 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DXN device

  1. #1 K222125 Arm Blood Pressure Monitor, models AOJ-30A…score 18.5 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K173916Finapres Nova Noninvasive Hemodynamic Monitor
Summary page 5
…s Medizintechnik GmbH Technical Documentation – Task Force® CORE 5.4 Predicate Devices 5.4.1 Primary predicate device (PPD): Finapres NOVA: K173916 Manufacturer: Finapres Medical Systems B.V. Hogehilweg 8, 1101 CC Amster…
Finapres Medical Systems B.V.2018—
*K183521CNAP Monitor
Summary page 5
…he Finapres NOVA share the same intended use and the same fields of application in the market. 5.4.2 Secondary predicate device (SPD): CNAP® Monitor 500 HD: K183521 Manufacturer: CNSystems Medizintechnik GmbH Graz, Austria The CNAP® Monitor 500HD is a s…
Cnsystems Medizintechnik GmbH2019—

Also cleared against K173916: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (5 Class I)

No FDA recall names this device.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K160552EV1000 Clinical Platform with ClearSight Finger Cuff or ClearSight System, EV1000 Clinical PlatformEdwards Lifesciences, LLC · 2016 · 2 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K093261MOVESThornhill Research, Inc. · 2010 · 3 generations back · ◆ in the 2010 ancestors bar of FIG. 12 Class I recalls
  • K013642AVEABird Products Corp. · 2002 · 6 generations back · ◆ in the 2002 ancestors bar of FIG. 113 Class I recalls
  • K981072Esprit VentilatorRespironics, Inc. · 1998 · 6 generations back · ◆ in the 1998 ancestors bar of FIG. 11 Class I recall
  • K970460Nellcor Puritan Bennett, 840 Ventilator System (Model 840)Puritan Bennett Corp. · 1998 · 7 generations back · ◆ in the 1998 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 23 years and passes through 5 companies; a new generation was cleared every 2.0 years on median.
  • 24 of the 105 ancestors have a recall of any class; 5 have a Class I recall.
  • 1 other device also cites its predicate K173916.

How does its category compare?

All DXN devices

DXN System, Measurement, Blood-Pressure, Non-Invasive: FDA decisions, review times, recalls and adverse event reports over 20 years. 2023, this device’s year, is highlighted.

FDA decisions per year · DXN

28in 2025

2007: 272008: 232009: 462010: 432011: 342012: 362013: 342014: 292015: 302016: 302017: 332018: 222019: 332020: 242021: 232022: 252023: 352024: 172025: 282026 (partial year): 29
2007high 46 (2009) · low 17 (2024)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DXN

142in 2025

2007: 118 days median2008: 82 days median2009: 73 days median2010: 94 days median2011: 72 days median2012: 116 days median2013: 150 days median2014: 133 days median2015: 97 days median2016: 144 days median2017: 184 days median2018: 235 days median2019: 156 days median2020: 198 days median2021: 190 days median2022: 137 days median2023: 182 days median2024: 107 days median2025: 142 days median2026 (partial year): 147 days median
2007high 235 (2018) · low 72 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DXN

3in 2024

2008: 12009: 32011: 12013: 12016: 12019: 12024: 3
2007high 3 (2009) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DXN

239in 2025

2007: 212008: 82009: 172010: 142011: 102012: 192013: 722014: 472015: 622016: 542017: 902018: 6172019: 962020: 1302021: 1622022: 1552023: 1622024: 2702025: 239
2007high 617 (2018) · low 8 (2008)2025

MAUDE reports naming product code DXN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200727118021
2008238218
20094673317
20104394014
20113472110
201236116019
201334150172
201429133047
20153097062
201630144154
201733184090
2018222350617
201933156196
2020241980130
2021231900162
2022251370155
2023351820162
2024171073270
2025281420239
2026291470—

What is it indicated for?

The Task Force® CORE (TFC) is intended for the measurement of non-invasive continuous blood pressure, pulse rate, and the determination of associated derived hemodynamic parameters. The TFC is intended to be used in professional healthcare facilities and operated by qualified healthcare professional staff. The TFC is intended to be used for adults and under uninterrupted surveillance of the operator. The TFC contains software which runs on a separate…

Read the full statement (125 words)

The Task Force® CORE (TFC) is intended for the measurement of non-invasive continuous blood pressure, pulse rate, and the determination of associated derived hemodynamic parameters.

The TFC is intended to be used in professional healthcare facilities and operated by qualified healthcare professional staff. The TFC is intended to be used for adults and under uninterrupted surveillance of the operator.

The TFC contains software which runs on a separate computer and which is used to operate the device. This software includes a basic frontend (Task Force® CORE Viewer) as well as an interface for integration into other software products.

The TFC can provide all signals as well as measured and derived parameters to the analog output interface and can acquire signals from the analog input interface.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K223332: Task Force CORE.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K223332

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026