510(k) Report

510(k) · Cleared

Cleared

OpenSight

K172418510(k) TraditionalCleared Sep 21, 2018407-day reviewClass II21 CFR 892.2050

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Novarad Corporation · LLZ System, Image Processing, Radiological

Cleared on Sep 21, 2018 as substantially equivalent to K170167 (Echopixel, Inc., 2017) and 1 other predicate after a 407-day review. Its family tree reaches back 9 generations to 1993. 5 later devices descend from it.

Decided
Sep 21, 2018
Review
407 days
Class
Class II
Product code
LLZ
Recalls
None
Predicates
2

Its longest line of descent runs back 9 generations to K924630 (1993). 49 ancestors in all, the oldest K890348 (1989). 5 later devices descend from it, 3 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2017: 1 other ancestor2006: 2 other ancestors2004: 6 other ancestors2003: 6 other ancestors2001: 4 other ancestors2000: 4 other ancestors1999: 1 other ancestor1998: 3 other ancestors2002: 2 other ancestors2007: 1 other ancestor1997: 1 other ancestor1996: 1 other ancestor1995: 2 other ancestors1994: 2 other ancestors1993: 3 other ancestors1989: 1 other ancestor2022: 3 descendants cleared2023: 1 descendant cleared2025: 1 descendant clearedK924630 · Surgicad (TM) View (TM) · Surgicad Corp. · 1993K970617 · CT-Modeller System · Materialise NV · 1997K033849 · Simplant System · Materialise NV · 2004K073468 · Mimics · Materialise NV · 2008K093269 · Procedure Rehearsal Studio · Simbionix, Ltd. · 2010K112387 · Procedure Rehearsal Studio (TM) · Simbionix, Ltd. · 2011K123023 · Surgery Rehearsal Platform · Surgical Theater, LLC · 2013K142107 · Echo True 3D Viewer · Echopixel, Inc. · 2015 · recalledK170167 · True 3D Viewer Software · Echopixel, Inc. · 2017 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 1 other ancestor3 generations back: 6 other ancestors4 generations back: 10 other ancestors5 generations back: 7 other ancestors6 generations back: 8 other ancestors7 generations back: 4 other ancestors8 generations back: 3 other ancestors 1 generation forward: 3 descendants2 generations forward: 2 descendants K170167 · True 3D Viewer Software · Echopixel, Inc. · 2017 · recalledK142107 · Echo True 3D Viewer · Echopixel, Inc. · 2015 · recalledK123023 · Surgery Rehearsal Platform · Surgical Theater, LLC · 2013K112387 · Procedure Rehearsal Studio (TM) · Simbionix, Ltd. · 2011K093269 · Procedure Rehearsal Studio · Simbionix, Ltd. · 2010K073468 · Mimics · Materialise NV · 2008K033849 · Simplant System · Materialise NV · 2004K970617 · CT-Modeller System · Materialise NV · 1997K924630 · Surgicad (TM) View (TM) · Surgicad Corp. · 1993−50
  1. Root · 9 generations backK924630 Surgicad (TM) View (TM)Surgicad Corp. · 1993
  2. 5 generations between
  3. 3 generations backK123023 Surgery Rehearsal PlatformSurgical Theater, LLC · 2013
  4. 2 generations backK142107 Echo True 3D ViewerEchopixel, Inc. · 2015 · Recalled
  5. 1 generation backK170167 True 3D Viewer SoftwareEchopixel, Inc. · 2017 · Recalled
  6. This deviceK172418 OpenSightNovarad Corporation · 2018
  7. Descendant · cites it directly · 1 of its ownK221796 Cyware 4H and Cysart 4HAbys Medical · 2022
  8. Descendant · cites it directly · 1 of its ownK213128 IntraOp VSP Software DeviceXironetic, LLC · 2022
  9. Descendant · 2 generations on · 0 of its ownK243069 Lumys (V1)Intradys · 2025
  10. Descendant · 2 generations on · 0 of its ownK232189 OrionXRHoth Intelligence, Inc. · 2023
  11. Descendant · cites it directly · 0 of its ownK220146 VisARNovarad Corporation · 2022
Show all 9 generations
  1. Root · 9 generations backK924630 Surgicad (TM) View (TM)Surgicad Corp. · 1993
  2. 8 generations backK970617 CT-Modeller SystemMaterialise NV · 1997
  3. 7 generations backK033849 Simplant SystemMaterialise NV · 2004
  4. 6 generations backK073468 MimicsMaterialise NV · 2008
  5. 5 generations backK093269 Procedure Rehearsal StudioSimbionix, Ltd. · 2010
  6. 4 generations backK112387 Procedure Rehearsal Studio (TM)Simbionix, Ltd. · 2011
  7. 3 generations backK123023 Surgery Rehearsal PlatformSurgical Theater, LLC · 2013
  8. 2 generations backK142107 Echo True 3D ViewerEchopixel, Inc. · 2015 · Recalled
  9. 1 generation backK170167 True 3D Viewer SoftwareEchopixel, Inc. · 2017 · Recalled
  10. This deviceK172418 OpenSightNovarad Corporation · 2018

Most influential descendants

Showing 5 of 5 · See all
  • K221796Cyware 4H and Cysart 4HAbys Medical · 2022 · cites it directly1descendant
  • K213128IntraOp VSP Software DeviceXironetic, LLC · 2022 · cites it directly1descendant
  • K243069Lumys (V1)Intradys · 2025 · 2 generations on0descendants
  • K232189OrionXRHoth Intelligence, Inc. · 2023 · 2 generations on0descendants
  • K220146VisARNovarad Corporation · 2022 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−9K924630Surgicad (TM) View (TM)Surgicad Corp.1993—
−8K970617CT-Modeller SystemMaterialise NV1997—
−7K033849Simplant SystemMaterialise NV2004—
−6K073468MimicsMaterialise NV2008—
−5K093269Procedure Rehearsal StudioSimbionix, Ltd.2010—
−4K112387Procedure Rehearsal Studio (TM)Simbionix, Ltd.2011—
−3K123023Surgery Rehearsal PlatformSurgical Theater, LLC2013—
−2K142107Echo True 3D ViewerEchopixel, Inc.20151
−1K170167True 3D Viewer SoftwareEchopixel, Inc.20171
0 (this device)K172418OpenSightNovarad Corporation2018—
+1K221796Cyware 4H and Cysart 4HAbys Medical2022—
+1K213128IntraOp VSP Software DeviceXironetic, LLC2022—
+2K243069Lumys (V1)Intradys2025—
+2K232189OrionXRHoth Intelligence, Inc.2023—
+1K220146VisARNovarad Corporation2022—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198910
199330
199420
199520
199610
199710
199830
199910
200040
200140
200220
200360
200460
200620
200710
201710

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202230
202310
202510
FIG. 1 — Lineage of K172418, 1993–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LLZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LLZ device

  1. #1 K230955 syngo Application Softwarescore 21.5 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K170167True 3D Viewer Software
Summary page 4
…Name: Imaging Software Regulation Name: Picture archiving and communications system Regulation Number: 21 CFR 892.2050 Product Code: LLZ Primary Predicate Device K170167 True 3D Viewer Software EchoPixel Inc. Predicate Device K163622 Lu…
Echopixel, Inc.20171
*K163622Lung Vision
Summary page 4
…mber: 21 CFR 892.2050 Product Code: LLZ Primary Predicate Device K170167 True 3D Viewer Software EchoPixel Inc. Predicate Device K163622 LungVision System Body Vision Medical Ltd. Novarad OpenSight 510(k) Submission | 2018…
Bodyvision Medical, Ltd.2017—

What descends from it? (5)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K221796Cyware 4H and Cysart 4HAbys Medical20221
K213128IntraOp VSP Software DeviceXironetic, LLC20221
K220146VisARNovarad Corporation20220

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 25 years and passes through 6 companies; a new generation was cleared every 2.1 years on median.
  • 21 of the 49 ancestors have a recall of any class.
  • Of its descendants, K221796 (Abys Medical, 2022) has gone furthest, with 1 descendant of its own.

How does its category compare?

All LLZ devices

LLZ System, Image Processing, Radiological: FDA decisions, review times, recalls and adverse event reports over 20 years. 2018, this device’s year, is highlighted.

FDA decisions per year · LLZ

58in 2025

2007: 762008: 942009: 702010: 512011: 1062012: 822013: 952014: 782015: 642016: 822017: 782018: 1042019: 742020: 812021: 832022: 652023: 872024: 622025: 582026 (partial year): 45
2007high 106 (2011) · low 51 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LLZ

173in 2025

2007: 36 days median2008: 52 days median2009: 62 days median2010: 98 days median2011: 79 days median2012: 87 days median2013: 84 days median2014: 89 days median2015: 90 days median2016: 90 days median2017: 107 days median2018: 105 days median2019: 107 days median2020: 115 days median2021: 116 days median2022: 137 days median2023: 157 days median2024: 131 days median2025: 173 days median2026 (partial year): 110 days median
2007high 173 (2025) · low 36 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LLZ

11in 2025

2007: 202008: 322009: 212010: 172011: 302012: 122013: 642014: 62015: 162016: 122017: 52018: 72019: 22020: 62021: 32022: 42023: 82024: 62025: 112026 (partial year): 14
2007high 64 (2013) · low 2 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LLZ

116in 2025

2007: 1522008: 1542009: 1452010: 1452011: 4542012: 2662013: 1,1302014: 15,6312015: 2112016: 2902017: 1042018: 722019: 882020: 822021: 702022: 282023: 262024: 422025: 116
2007high 15,631 (2014) · low 26 (2023)2025

MAUDE reports naming product code LLZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007763620152
2008945232154
2009706221145
2010519817145
20111067930454
2012828712266
20139584641,130
20147889615,631
2015649016211
2016829012290
2017781075104
2018104105772
201974107288
202081115682
202183116370
202265137428
202387157826
202462131642
20255817311116
20264511014—

What is it indicated for?

OpenSight is intended to enable users to display, manipulate, and evaluate 2D, 3D, and 4D digital images acquired from CR, DX, CT, MR, and PT sources. It is intended to visualize 3D imaging holograms of the patient, on the patient, for pre-operative localization and pre-operative planning of surgical options. OpenSight is designed for use only with performance-tested hardware specified in the user documentation. OpenSight is intended to enable users to…

Read the full statement (142 words)

OpenSight is intended to enable users to display, manipulate, and evaluate 2D, 3D, and 4D digital images acquired from CR, DX, CT, MR, and PT sources. It is intended to visualize 3D imaging holograms of the patient, on the patient, for pre-operative localization and pre-operative planning of surgical options. OpenSight is designed for use only with performance-tested hardware specified in the user documentation.

OpenSight is intended to enable users to segment previously acquired 3D datasets, overlay, and register these 3D segmented datasets with the same anatomy of the patient in order to support pre-operative analysis.

OpenSight is not intended for intraoperative use. It is not to be used for stereotactic procedures.

OpenSight is intended for use by trained healthcare professionals, including surgeons, radiologists, chiropractors, physicians, cardiologists, technologists, and medical educators. The device assists doctors to better understand anatomy and pathology of patient.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K172418: OpenSight.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K172418

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026