510(k) Report

510(k) · Cleared

Recalled · 2004

LC, Cardiac Positioner #B5078A

K890348510(k) TraditionalCleared Feb 27, 198935-day reviewClass II21 CFR 892.1600Recalls 5

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

General Electric Co. · IZI System, X-Ray, Angiographic

Cleared on Feb 27, 1989 after a 35-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 10 of the 332 devices in its family, counting itself, have a Class I recall. 331 later devices descend from it. It has been recalled 5 times.

Decided
Feb 27, 1989
Review
35 days
Class
Class II
Product code
IZI
Recalls
5
Predicates
0

331 later devices descend from it, 1 citing it directly. 10 of the 332 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2000: 1 descendant cleared2002: 5 descendants cleared2005: 2 descendants cleared2010: 12 descendants cleared2008: 11 descendants cleared2003: 5 descendants cleared2007: 1 descendant cleared2004: 2 descendants cleared2009: 7 descendants cleared2006: 4 descendants cleared2012: 14 descendants cleared2016: 17 descendants cleared, 1 with a Class I recall12011: 10 descendants cleared2015: 15 descendants cleared2013: 16 descendants cleared, 1 with a Class I recall12018: 27 descendants cleared2017: 12 descendants cleared, 2 with a Class I recall22014: 19 descendants cleared2019: 11 descendants cleared, 1 with a Class I recall12022: 20 descendants cleared2021: 17 descendants cleared2024: 19 descendants cleared, 2 with a Class I recall22020: 19 descendants cleared, 1 with a Class I recall12025: 29 descendants cleared, 1 with a Class I recall12023: 15 descendants cleared, 1 with a Class I recall12026: 21 descendants cleared1990200020102020
  1. This deviceK890348 LC, Cardiac Positioner #B5078AGeneral Electric Co. · 1989
  2. Descendant · cites it directly · 330 of its ownK993037 LCV+ Version 2Ge Medical Systems, Inc. · 2000 · Recalled
  3. Descendant · 2 generations on · 218 of its ownK021021 Modular Angiography System Axiom ArtisSiemens Medical Solutions USA, Inc. · 2002 · Recalled
  4. Descendant · 2 generations on · 124 of its ownK020055 Philips Integris Allura 9 with FD OptionPhilips Medical Systems · 2002 · Recalled
  5. Descendant · 3 generations on · 114 of its ownK022899 Philips Integris Allura Flat Detector Release 1.0Philips Medical Systems North America Co. · 2002 · Recalled
  6. Descendant · 4 generations on · 113 of its ownK031333 Philips Integris Allura Flat Detector Release 1.2Philips Medical Systems North America, Inc. · 2003 · Recalled

Showing 5 of 331 descendants. See all descendants

Descendants cleared per year

2000: 12002: 52003: 52004: 22005: 22006: 42007: 12008: 112009: 72010: 122011: 102012: 142013: 162014: 192015: 152016: 172017: 122018: 272019: 112020: 192021: 172022: 202023: 152024: 192025: 292026: 21
2000high 29 (2025) · low 0 (2001)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 331 · See all
  • K993037LCV+ Version 2Ge Medical Systems, Inc. · 2000 · cites it directly330descendants
  • K021021Modular Angiography System Axiom ArtisSiemens Medical Solutions USA, Inc. · 2002 · 2 generations on218descendants
  • K020055Philips Integris Allura 9 with FD OptionPhilips Medical Systems · 2002 · 2 generations on124descendants
  • K022899Philips Integris Allura Flat Detector Release 1.0Philips Medical Systems North America Co. · 2002 · 3 generations on114descendants
  • K031333Philips Integris Allura Flat Detector Release 1.2Philips Medical Systems North America, Inc. · 2003 · 4 generations on113descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K890348LC, Cardiac Positioner #B5078AGeneral Electric Co.1989—
+1K993037LCV+ Version 2Ge Medical Systems, Inc.20006
+2K021021Modular Angiography System Axiom ArtisSiemens Medical Solutions USA, Inc.200221
+2K020055Philips Integris Allura 9 with FD OptionPhilips Medical Systems20022
+3K022899Philips Integris Allura Flat Detector Release 1.0Philips Medical Systems North America Co.20028
+4K031333Philips Integris Allura Flat Detector Release 1.2Philips Medical Systems North America, Inc.200325

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200010
200250
200350
200420
200520
200640
200710
2008110
200970
2010120
2011100
2012140
2013161
2014190
2015150
2016171
2017122
2018270
2019111
2020191
2021170
2022200
2023151
2024192
2025291
2026210
FIG. 1 — Lineage of K890348, 1989–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

10 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new IZI device

  1. #1 K230727 SPY Portable Handheld Imaging (SPY-PHI) Systemscore 23.2 · 20 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it? (331)

See all 331

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K993037LCV+ Version 2Ge Medical Systems, Inc.2000330

Has anything in its family been recalled? (10 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 24, 2004Not classifiedZ-0896-04OtherUnintended movements of both the C-arc and the image intensifier have occured due to shearing of the gear motor shaft as a result of the misalignment of the gearbox.
Mar 24, 2004Not classifiedZ-0897-04OtherUnintended movements of both the C-arc and the image intensifier have occured due to shearing of the gear motor shaft as a result of the misalignment of the gearbox.
Mar 24, 2004Not classifiedZ-0898-04OtherUnintended movements of both the C-arc and the image intensifier have occured due to shearing of the gear motor shaft as a result of the misalignment of the gearbox.
Mar 24, 2004Not classifiedZ-0899-04OtherUnintended movements of both the C-arc and the image intensifier have occured due to shearing of the gear motor shaft as a result of the misalignment of the gearbox.
Mar 24, 2004Not classifiedZ-0900-04OtherUnintended movements of both the C-arc and the image intensifier have occured due to shearing of the gear motor shaft as a result of the misalignment of the gearbox.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K130842Allura Xper FD or Table Series; Allura Xper FD10 Series; Allura Xper FD20 Series; Allura Xper FD Biplane…Philips Medical Systems Nederland B.V. · 2013 · 6 generations on · ◆ in the 2013 descendants bar of FIG. 111 Class I recalls
  • K162859Allura Xper FD series and Allura Xper OR Table seriesPhilips Medical Systems Nederland B.V. · 2016 · 7 generations on · ◆ in the 2016 descendants bar of FIG. 112 Class I recalls
  • K163715Allura Xper R9Philips Medical Systems Nederland B.V. · 2017 · 7 generations on · ◆ in the 2017 descendants bar of FIG. 17 Class I recalls
  • K172822Azurion series R1.2Philips Medical Systems Nederland B.V. · 2017 · 8 generations on · ◆ in the 2017 descendants bar of FIG. 19 Class I recalls
  • K200917Azurion R2.1Philips Medical Systems Nederland B.V. · 2020 · 9 generations on · ◆ in the 2020 descendants bar of FIG. 18 Class I recalls
  • K190106Intra-Operative Positioning System; Simple Curve Catheter, Reverse Curve Catheter ; Angled Tip Guidewire…Centerline Biomedical, Inc. · 2019 · 10 generations on · ◆ in the 2019 descendants bar of FIG. 11 Class I recall
  • K242133Intra-Operative Positioning System (IOPS®) (Fiducial Tracking Pad); Intra-Operative Positioning System…Centerline Biomedical, Inc. · 2024 · 11 generations on · ◆ in the 2024 descendants bar of FIG. 11 Class I recall
  • K241243Intra-Operative Positioning System (IOPS®) Viewpoint (Simple Curve Catheters); Intra-Operative Positioning…Centerline Biomedical, Inc. · 2024 · 11 generations on · ◆ in the 2024 descendants bar of FIG. 11 Class I recall
  • K230309Intra-Operative Positioning System (IOPS®) (MC-1); IOPS Simple Curve Catheter (SCC-1) and IOPS Reverse Curve…Centerline Biomedical, Inc. · 2023 · 11 generations on · ◆ in the 2023 descendants bar of FIG. 11 Class I recall
  • K243842Intra-Operative Positioning System (IOPS®)Centerline Biomedical, Inc. · 2025 · 11 generations on · ◆ in the 2025 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • Of its descendants, K993037 (Ge Medical Systems, Inc., 2000) has gone furthest, with 330 descendants of its own.

How does its category compare?

All IZI devices

IZI System, X-Ray, Angiographic: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · IZI

2in 2025

2007: 72008: 62009: 22010: 22011: 42012: 12013: 12014: 42016: 12017: 22018: 22019: 22020: 12021: 12022: 12023: 32024: 22025: 22026 (partial year): 2
2007high 7 (2007) · low 0 (2015)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · IZI

81in 2023

2007: 28 days median2008: 68 days median2011: 104 days median2014: 111 days median2023: 81 days median
2007high 111 (2014) · low 28 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · IZI

1in 2021

2007: 62008: 522009: 182010: 112011: 62012: 62013: 62014: 52015: 142016: 72017: 12018: 22019: 22020: 22021: 12026 (partial year): 1
2007high 52 (2008) · low 0 (2022)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · IZI

2,171in 2025

2007: 922008: 1012009: 1322010: 2522011: 3772012: 772013: 582014: 1232015: 362016: 1022017: 882018: 492019: 302020: 142021: 292022: 2652023: 1,2042024: 1,3322025: 2,171
2007high 2,171 (2025) · low 14 (2020)2025

MAUDE reports naming product code IZI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007728692
200866852101
20092—18132
20102—11252
201141046377
20121—677
20131—658
201441115123
20150—1436
20161—7102
20172—188
20182—249
20192—230
20201—214
20211—129
20221—0265
202338101,204
20242—01,332
20252—02,171
20262—1—
Cite this page

510(k) Report. “K890348: LC, Cardiac Positioner #B5078A.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K890348

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026