510(k) Report

510(k) · Cleared

Cleared

Ecolab POU Water Filter

K173164510(k) TraditionalCleared Jun 29, 2018273-day reviewClass II21 CFR 876.5665

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ecolab, Inc. · NHV System, Water Purification, General Medical Use

Cleared on Jun 29, 2018 as substantially equivalent to K153784 (Evoqua Water Technologies, LLC, 2016) after a 273-day review. Its family tree reaches back 4 generations to 2001. 1 later device descends from it.

Decided
Jun 29, 2018
Review
273 days
Class
Class II
Product code
NHV
Recalls
None
Predicates
1

Its longest line of descent runs back 4 generations to K002045 (2001). 1 later device descends from it, 1 citing it directly.

  1. Root · 4 generations backK002045 Water Purification System for HemodialysisTotal Water Treatment Systems, Inc. · 2001
  2. 3 generations backK012716 Mainstream Water Purification DevicePrismedical Corp. · 2002
  3. 2 generations backK141731 Nephros Dsu-H, Nephros Ssu-HNephros, Inc. · 2014
  4. 1 generation backK153784 Nosogard FiltersEvoqua Water Technologies, LLC · 2016
  5. This deviceK173164 Ecolab POU Water FilterEcolab, Inc. · 2018
  6. Descendant · cites it directly · 0 of its ownK182164 Ecolab POU Water Filter Spray Outlet, Ecolab POU Water Filter Nozzle…Ecolab, Inc. · 2019

Most influential descendants

Showing 1 of 1 · See all
  • K182164Ecolab POU Water Filter Spray Outlet, Ecolab POU Water Filter Nozzle Outlet, Ecolab POU…Ecolab, Inc. · 2019 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K002045Water Purification System for HemodialysisTotal Water Treatment Systems, Inc.2001—
−3K012716Mainstream Water Purification DevicePrismedical Corp.2002—
−2K141731Nephros Dsu-H, Nephros Ssu-HNephros, Inc.2014—
−1K153784Nosogard FiltersEvoqua Water Technologies, LLC2016—
0 (this device)K173164Ecolab POU Water FilterEcolab, Inc.2018—
+1K182164Ecolab POU Water Filter Spray Outlet, Ecolab POU Water Filter Nozzle Outlet…Ecolab, Inc.2019—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201910
FIG. 1 — Lineage of K173164, 2001–2019Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in NHV? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new NHV device

  1. #1 K261470 QF10score 13.2 · 5 generations · no Class I recall
  2. #2 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K153784Nosogard Filters
Summary page 4
…ater purification system for hemodialysis Regulation Number 21 CFR 876.5665 Product Code NHV Predicate Device Evoqua Water Technologies, Nosogard Filters, K153784 Device Description The Ecolab POU Water Filter is a disposable Polyethersulfone (PES) membrane filter seal…
Evoqua Water Technologies, LLC2016—

Also cleared against K153784: 1 other device →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K182164Ecolab POU Water Filter Spray Outlet, Ecolab POU Water Filter Nozzle Outlet, Ecolab POU…Ecolab, Inc.20190

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 17 years and passes through 5 companies; a new generation was cleared every 2.0 years on median.
  • None of the 4 ancestors has a recall.
  • Of its descendants, K182164 (Ecolab, Inc., 2019) has gone furthest, with 0 descendants of its own.
  • 1 other device also cites its predicate K153784.

How does its category compare?

All NHV devices

NHV System, Water Purification, General Medical Use: FDA decisions, review times, recalls and adverse event reports over 20 years. 2018, this device’s year, is highlighted.

FDA decisions per year · NHV

1in 2023

2014: 12016: 42017: 12018: 12019: 22023: 12026 (partial year): 1
2007high 4 (2016) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · NHV

203in 2016

2016: 203 days median
2007high 203 (2016)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · NHV

No recall names NHV since 2007.

Adverse event reports per year · NHV

1in 2025

2017: 12018: 22019: 12022: 42025: 1
2007high 4 (2022) · low 0 (2007)2025

MAUDE reports naming product code NHV, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20100—00
20110—00
20120—00
20130—00
20141—00
20150—00
2016420300
20171—01
20181—02
20192—01
20200—00
20210—00
20220—04
20231—00
20240—00
20250—01
20261—0—

What is it indicated for?

The ECOLAB POU Water Filters are intended to operate on EPA quality drinking water sources as a microbial retention filter. ECOLAB POU Water Filters are suitable for the control of bacteria equal to or greater in size than Brevundimons diminuta. The POU Filters are suitable for general point of use infection control for procedures such as superficial wound cleansing, cleaning of equipment, washing of surgeon’s hands and bathing where the reduction of such microorganisms in the water is desired.

The POU Filters are not intended for use in the production of USP sterile water for use in infusion, injection or production of fluids for use in dialysis treatments.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K173164: Ecolab POU Water Filter.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K173164

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026