510(k) Report

510(k) · Cleared

Recalled · Class II · 2023

ProxiDiagnost N90

K173433510(k) TraditionalCleared Feb 5, 201895-day reviewClass II21 CFR 892.1650Recalls 7

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Philips Medical Systems Dmc GmbH · OWB Interventional Fluoroscopic X-Ray System

Cleared on Feb 5, 2018 as substantially equivalent to K031535 (Philips Medical Systems North America Co., 2003) after a 95-day review. Its family tree reaches back 3 generations to 1992. 2 later devices descend from it. It has been recalled 7 times.

Decided
Feb 5, 2018
Review
95 days
Class
Class II
Product code
OWB
Recalls
7
Predicates
1

Its longest line of descent runs back 3 generations to K924593 (1992). 2 later devices descend from it, 1 citing it directly.

  1. Root · 3 generations backK924593 Diagnost 76 Family*R/F Systems - Release 4.4Philips Medical Systems, Inc. · 1992 · Recalled
  2. 2 generations backK970640 Philips Easy DiagnostPhilips Medical Systems North America, Inc. · 1997 · Recalled
  3. 1 generation backK031535 Philips Easydiagnost ElevaPhilips Medical Systems North America Co. · 2003 · Recalled
  4. This deviceK173433 ProxiDiagnost N90Philips Medical Systems Dmc GmbH · 2018
  5. Descendant · cites it directly · 1 of its ownK212837 ProxiDiagnost N90Philips Medical Systems Dmc GmbH · 2021 · Recalled
  6. Descendant · 2 generations on · 0 of its ownK233945 ProxiDiagnost N90 / Precision CRF (706110, 706400)Philips Medical Systems Dmc GmbH · 2024 · Recalled

Most influential descendants

Showing 2 of 2 · See all
  • K212837ProxiDiagnost N90Philips Medical Systems Dmc GmbH · 2021 · cites it directly1descendant
  • K233945ProxiDiagnost N90 / Precision CRF (706110, 706400)Philips Medical Systems Dmc GmbH · 2024 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K924593Diagnost 76 Family*R/F Systems - Release 4.4Philips Medical Systems, Inc.19921
−2K970640Philips Easy DiagnostPhilips Medical Systems North America, Inc.19975
−1K031535Philips Easydiagnost ElevaPhilips Medical Systems North America Co.200315
0 (this device)K173433ProxiDiagnost N90Philips Medical Systems Dmc GmbH2018—
+1K212837ProxiDiagnost N90Philips Medical Systems Dmc GmbH20215
+2K233945ProxiDiagnost N90 / Precision CRF (706110, 706400)Philips Medical Systems Dmc GmbH20242

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202110
202410
FIG. 1 — Lineage of K173433, 1992–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in OWB? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new OWB device

  1. #1 K251602 Alphenix, INFX-8000V/B, INFX-8000V/S, V9.6…score 18.3 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K031535Philips Easydiagnost Eleva
Summary page 4
…hilips EasyDiagnost Eleva Manufacturer: Philips Medical System DMC 510(k) Clearance: K031535 – June 17, 2003 Classification Name: Image-intensified fluoroscopic x-ray…
Philips Medical Systems North America Co.200315
†K163210Philips CombiDiagnost R90Philips Medical Systems Dmc GmbH20178
†K153318Philips Eleva Workspot with SkyFlowPhilips Medical Systems Dmc GmbH2015—
†K170113pixium 4343RCEPhilips Medical Systems Dmc GmbH2017—
†K171461SkyPlate Detector for Philips Radiography/Fluoroscopy SystemsPhilips Medical Systems Dmc GmbH2017—

Also cleared against K031535: 1 other device →

What descends from it? (2)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K212837ProxiDiagnost N90Philips Medical Systems Dmc GmbH20211

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 3, 2023Class IIZ-1587-2023Radiation Control for Health and Safety ActPhilips has identified a compliance issue where ProxiDiagnost N90 R.1.0, CombiDiagnost R90 R.1.0 and CombiDiagnost R90 R1.1 systems are missing a certification label required by the Food and Drug Administration (FDA).
Oct 21, 2022Class IIZ-1477-2023Component design/selectionPotential table stop due to a broken tabletop cable: On the ProxiDiagnost N90 table there is a cable located under the tabletop that is covered by a plastic sleeve which can crack and damage the cable. If the tabletop cable breaks, the…
Oct 21, 2021Class IIZ-1356-2023Device DesignIf the system experiences an overvoltage / power surge it can cause a short circuit condition within the Main- cabinet. If this issue occurs, the user may experience an electrical fire within the cabinet, see or smell smoke emitting from…
Sep 22, 2021Class IIZ-0240-2022Radiation Control for Health and Safety ActThe Aleva monitor requires certain labels to be applied, some of which are required by FDA Electronic Product Radiation Control performance standards. This includes the Radiation Warning Label [21CFR 1020.30 (j)] and the X-ray control…
Dec 8, 2020Class IIZ-0727-2021Radiation Control for Health and Safety ActAfter the user selects the Lock-in command, the kV and mA values are not locked immediately and can adjust if the collimator shutter position is changed directly after the Lock-in command.
Apr 23, 2020Class IIZ-2256-2020Process controlThe thermo switches in the three-phase transformer, which normally aid in powering down the System Power Distribution Unit, may be incorrectly installed and not working. If the transformer overheats due to a first failure and the thermo…
Nov 20, 2019Class IIZ-3023-2020Radiation Control for Health and Safety ActUnexpected increase in peak tube potential (kV) will lead to an increased patient radiation dose.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 26 years and passes through 4 companies; a new generation was cleared every 6.2 years on median.
  • 3 of the 3 ancestors have a recall of any class.
  • Of its descendants, K212837 (Philips Medical Systems Dmc GmbH, 2021) has gone furthest, with 1 descendant of its own.
  • 1 other device also cites its predicate K031535.

How does its category compare?

All OWB devices

OWB Interventional Fluoroscopic X-Ray System: FDA decisions, review times, recalls and adverse event reports over 20 years. 2018, this device’s year, is highlighted.

FDA decisions per year · OWB

22in 2025

2007: 32008: 42009: 92010: 72011: 62012: 62013: 112014: 152015: 112016: 162017: 202018: 232019: 92020: 162021: 242022: 212023: 122024: 132025: 222026 (partial year): 15
2007high 24 (2021) · low 3 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OWB

148in 2025

2007: 52 days median2008: 135 days median2009: 99 days median2010: 66 days median2011: 79 days median2012: 37 days median2013: 119 days median2014: 158 days median2015: 66 days median2016: 103 days median2017: 76 days median2018: 54 days median2019: 174 days median2020: 53 days median2021: 60 days median2022: 90 days median2023: 152 days median2024: 126 days median2025: 148 days median2026 (partial year): 221 days median
2007high 221 (2026) · low 37 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · OWB

42so far in 2026, the most since 2007

2008: 12013: 102014: 92015: 352016: 132017: 132018: 62019: 72020: 82021: 312022: 242023: 222024: 112025: 302026 (partial year): 42
2007high 42 (2026) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · OWB

14,008in 2025

2007: 12011: 12012: 12013: 272014: 1032015: 2,1342016: 4,0742017: 1,6042018: 1,4492019: 1,5582020: 1,5812021: 1,9812022: 2,9322023: 7,6042024: 7,9422025: 14,008
2007high 14,008 (2025) · low 0 (2008)2025

MAUDE reports naming product code OWB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200735201
2008413510
200999900
201076600
201167901
201263701
2013111191027
2014151589103
20151166352,134
201616103134,074
20172076131,604
2018235461,449
2019917471,558
2020165381,581
20212460311,981
20222190242,932
202312152227,604
202413126117,942
2025221483014,008
20261522142—

What is it indicated for?

ProxiDiagnost N90 is a multi-functional general R/F system. It is suitable for all routine radiography and fluoroscopy exams, including specialist areas like angiography or pediatric work, excluding mammography.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K173433: ProxiDiagnost N90.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K173433

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026