510(k) Report

510(k) · Cleared

Recalled · Class II · 2020

SOMATOM Edge Plus

K173607510(k) TraditionalCleared Mar 21, 2018120-day reviewClass II21 CFR 892.1750Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Siemens Medical Solutions USA, Inc. · JAK System, X-Ray, Tomography, Computed

Cleared on Mar 21, 2018 as substantially equivalent to K152036 (Siemens Medical Solutions USA, Inc., 2015) after a 120-day review. Its family tree reaches back 8 generations to 1998. 20 later devices descend from it. It has been recalled 3 times.

Decided
Mar 21, 2018
Review
120 days
Class
Class II
Product code
JAK
Recalls
3
Predicates
1

Its longest line of descent runs back 8 generations to K982349 (1998). 11 ancestors in all. 20 later devices descend from it, 1 citing it directly.

  1. Root · 8 generations backK982349 Somatom Plus 4WITH Volume Zoom CT ScannersSiemens Medical Solutions USA, Inc. · 1998
  2. 7 generations backK013522 Somatom P30 CT SystemsSiemens Medical Systems, Inc. · 2001 · Recalled
  3. 6 generations backK040665 Somatom Sensation 64 and Sensation CardiacSiemens Medical Solutions USA, Inc. · 2004 · Recalled
  4. 5 generations backK081022 Somatom Definition, Model as/as+Siemens Medical Solutions USA, Inc. · 2008 · Recalled
  5. 4 generations backK103127 Somatom Definition asSiemens Medical Solutions USA, Inc. · 2011 · Recalled
  6. 3 generations backK130901 Somatom Definition as OpenSiemens Medical Systems, Inc. · 2014
  7. 2 generations backK143401 SOMATOM Definition EdgeSiemens Medical Solutions USA, Inc. · 2015 · Recalled
  8. 1 generation backK152036 SOMATOM Definition Edge, SOMATOM Definition AS/AS+Siemens Medical Solutions USA, Inc. · 2015 · Recalled
  9. This deviceK173607 SOMATOM Edge PlusSiemens Medical Solutions USA, Inc. · 2018
  10. Descendant · cites it directly · 19 of its ownK190578 SOMATOM Force, SOMATOM Definition Flash, SOMATOM Drive, SOMATOM…Siemens Medical Solutions USA, Inc. · 2019 · Recalled
  11. Descendant · 2 generations on · 16 of its ownK192061 SOMATOM Go Platform – SOMATOM go.Now, go.Up, go.All, go.Top, go.Sim…Siemens Medical Solutions USA, Inc. · 2019 · Recalled
  12. Descendant · 2 generations on · 15 of its ownK191891 SOMATOM X.citeSiemens Medical Solutions USA, Inc. · 2019
  13. Descendant · 3 generations on · 14 of its ownK200524 SOMATOM X.cite, Scan&GO, SOMATOM go.Platform - go.Up, go.Now…Siemens Medical Solutions USA, Inc. · 2020 · Recalled
  14. Descendant · 4 generations on · 12 of its ownK211373 SOMATOM go.Platform Scanners - SOMATOM go.Now, go.Up, go.All…Siemens Medical Solutions USA, Inc. · 2021 · Recalled

Showing 5 of 20 descendants. See all descendants

Descendants cleared per year

2019: 32020: 22021: 22022: 12023: 32024: 32025: 42026: 2
2019high 4 (2025) · low 1 (2022)2026

Most influential descendants

Showing 5 of 20 · See all
  • K190578SOMATOM Force, SOMATOM Definition Flash, SOMATOM Drive, SOMATOM Definition Edge, SOMATOM…Siemens Medical Solutions USA, Inc. · 2019 · cites it directly19descendants
  • K192061SOMATOM Go Platform – SOMATOM go.Now, go.Up, go.All, go.Top, go.Sim, go.Open Pro, Scan &…Siemens Medical Solutions USA, Inc. · 2019 · 2 generations on16descendants
  • K191891SOMATOM X.citeSiemens Medical Solutions USA, Inc. · 2019 · 2 generations on15descendants
  • K200524SOMATOM X.cite, Scan&GO, SOMATOM go.Platform - go.Up, go.Now, go.Top, go.All, go. Sim…Siemens Medical Solutions USA, Inc. · 2020 · 3 generations on14descendants
  • K211373SOMATOM go.Platform Scanners - SOMATOM go.Now, go.Up, go.All, go.Top, SOMATOM X.…Siemens Medical Solutions USA, Inc. · 2021 · 4 generations on12descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−8K982349Somatom Plus 4WITH Volume Zoom CT ScannersSiemens Medical Solutions USA, Inc.1998—
−7K013522Somatom P30 CT SystemsSiemens Medical Systems, Inc.20019
−6K040665Somatom Sensation 64 and Sensation CardiacSiemens Medical Solutions USA, Inc.20045
−5K081022Somatom Definition, Model as/as+Siemens Medical Solutions USA, Inc.20083
−4K103127Somatom Definition asSiemens Medical Solutions USA, Inc.20116
−3K130901Somatom Definition as OpenSiemens Medical Systems, Inc.2014—
−2K143401SOMATOM Definition EdgeSiemens Medical Solutions USA, Inc.20153
−1K152036SOMATOM Definition Edge, SOMATOM Definition AS/AS+Siemens Medical Solutions USA, Inc.20154
0 (this device)K173607SOMATOM Edge PlusSiemens Medical Solutions USA, Inc.2018—
+1K190578SOMATOM Force, SOMATOM Definition Flash, SOMATOM Drive, SOMATOM Definition…Siemens Medical Solutions USA, Inc.201913
+2K192061SOMATOM Go Platform – SOMATOM go.Now, go.Up, go.All, go.Top, go.Sim, go.Open…Siemens Medical Solutions USA, Inc.20192
+2K191891SOMATOM X.citeSiemens Medical Solutions USA, Inc.2019—
+3K200524SOMATOM X.cite, Scan&GO, SOMATOM go.Platform - go.Up, go.Now, go.Top, go.All…Siemens Medical Solutions USA, Inc.20207
+4K211373SOMATOM go.Platform Scanners - SOMATOM go.Now, go.Up, go.All, go.Top, SOMATOM…Siemens Medical Solutions USA, Inc.20213

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200610
201210
201510

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201930
202020
202120
202210
202330
202430
202540
202620
FIG. 1 — Lineage of K173607, 1998–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JAK? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JAK device

  1. #1 K230223 iCAC Devicescore 23.5 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K152036SOMATOM Definition Edge, SOMATOM Definition AS/AS+
Summary page 5
…Siemens Medical Solutions USA, Inc. Product Code: 90 JAK Device Class: Class II 510(k) Number: K152036 Recall Information: This predicate device has not been the subject of any design r…
Siemens Medical Solutions USA, Inc.20154
†K161196SOMATOM DriveSiemens Medical Solutions USA, Inc.20164
†K162302Somatom Confidence ® RT ProSiemens Medical Solutions USA, Inc.20161

Also cleared against K152036: 1 other device →

What descends from it? (20)

See all 20

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K190578SOMATOM Force, SOMATOM Definition Flash, SOMATOM Drive, SOMATOM Definition Edge, SOMATOM…Siemens Medical Solutions USA, Inc.201919

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Nov 24, 2020Class IIZ-0771-2021Software designSystem does not trigger a cancel command followed by a reload of the scan with the new parameter(s). This issue may result in a delay in diagnosis and/or need for patient rescan
Jul 3, 2019Class IIZ-2273-2019Software designPotential for four safety-related software issues impacting systems utilizing syngo.CT software version VB10A which may result in a scan abort.
Feb 14, 2019Class IIZ-1246-2019Software designPotential risk of scans being aborted when using the optional Dual Spiral Dual Engery mode due to a software issue found in syngo CT VB10A for the Siemens Healthineers CT scanners with a single X-ray tube.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 20 years and passes through 2 companies; a new generation was cleared every 2.8 years on median.
  • 8 of the 11 ancestors have a recall of any class.
  • Of its descendants, K190578 (Siemens Medical Solutions USA, Inc., 2019) has gone furthest, with 19 descendants of its own.
  • 1 other device also cites its predicate K152036.

How does its category compare?

All JAK devices

JAK System, X-Ray, Tomography, Computed: FDA decisions, review times, recalls and adverse event reports over 20 years. 2018, this device’s year, is highlighted.

FDA decisions per year · JAK

27in 2025

2007: 142008: 262009: 152010: 72011: 162012: 282013: 172014: 302015: 302016: 252017: 292018: 242019: 282020: 302021: 202022: 302023: 272024: 262025: 272026 (partial year): 21
2007high 30 (2014) · low 7 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JAK

141in 2025

2007: 27 days median2008: 42 days median2009: 31 days median2010: 92 days median2011: 155 days median2012: 121 days median2013: 132 days median2014: 200 days median2015: 140 days median2016: 129 days median2017: 96 days median2018: 114 days median2019: 94 days median2020: 112 days median2021: 145 days median2022: 131 days median2023: 139 days median2024: 141 days median2025: 141 days median2026 (partial year): 167 days median
2007high 200 (2014) · low 27 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JAK

17in 2025

2007: 112008: 162009: 292010: 512011: 342012: 232013: 112014: 102015: 282016: 442017: 332018: 102019: 52020: 82021: 152022: 92023: 92024: 202025: 172026 (partial year): 23
2007high 51 (2010) · low 5 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JAK

219in 2025

2007: 762008: 1202009: 1052010: 1192011: 2052012: 1922013: 4142014: 3082015: 4542016: 3652017: 3382018: 5432019: 2332020: 1772021: 2612022: 1992023: 3102024: 2122025: 219
2007high 543 (2018) · low 76 (2007)2025

MAUDE reports naming product code JAK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200714271176
2008264216120
2009153129105
201079251119
20111615534205
20122812123192
20131713211414
20143020010308
20153014028454
20162512944365
2017299633338
20182411410543
201928945233
2020301128177
20212014515261
2022301319199
2023271399310
20242614120212
20252714117219
20262116723—

What is it indicated for?

This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis. The images delivered by the system can be used by trained staff as an aid in diagnosis, treatment preparation and radiation therapy planning. This CT system can be used for low dose lung cancer screening in…

Read the full statement (109 words)

This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data.

The images delivered by the system can be used by a trained physician as an aid in diagnosis.

The images delivered by the system can be used by trained staff as an aid in diagnosis, treatment preparation and radiation therapy planning.

This CT system can be used for low dose lung cancer screening in high risk populations.*

*As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011;365:395-409) and subsequent literature, for further information.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K173607: SOMATOM Edge Plus.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K173607

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026