510(k) Report

510(k) · Cleared

Recalled · Class I · 2023

Quantum Pump Console

K173834510(k) TraditionalCleared May 25, 2018158-day reviewClass II21 CFR 870.4220Recalls 2 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Spectrum Medical, Ltd. · DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass

Cleared on May 25, 2018 as substantially equivalent to K060053 (Sorin Group Deutschland GmbH, 2006) after a 158-day review. Its family tree reaches back 1 generation to 2006. 1 of the 2 devices in its family, counting itself, has a Class I recall. It has been recalled 2 times, including a Class I recall.

Decided
May 25, 2018
Review
158 days
Class
Class II
Product code
DTQ
Recalls
2 (1 Class I)
Predicates
1

Its longest line of descent runs back 1 generation to K060053 (2006). No later device cites it as a predicate yet. 1 of the 2 devices in this family, counting this one, has a Class I recall.

K060053 · Stockert S5 System · Sorin Group Deutschland GmbH · 2006 · recalled2005201020152020
  1. Root · 1 generation backK060053 Stockert S5 SystemSorin Group Deutschland GmbH · 2006 · Recalled
  2. This deviceK173834 Quantum Pump ConsoleSpectrum Medical, Ltd. · 2018
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K060053Stockert S5 SystemSorin Group Deutschland GmbH200620
0 (this device)K173834Quantum Pump ConsoleSpectrum Medical, Ltd.2018—
FIG. 1 — Lineage of K173834, 2006–2018Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DTQ device

  1. #1 K221373 Essenz HLMscore 8.9 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K060053Stockert S5 System
Summary page 4
…Division of Cardiovascular Devices (DCD) Circulatory Support Devices Branch (CSDB) III. PREDICATE DEVICE Stöckert S5 System (K060053) currently marketed by LivaNova IV. DEVICE DESCRIPTION The Quantum Pump Console is a configurable heart-lung machine consisting…
Sorin Group Deutschland GmbH200620

Also cleared against K060053: 2 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 13, 2023Class IIZ-0224-2025Component design/selectionThe ratchet and pawl had an out of tolerance geometry which exhibited as the left bobbin would not lock.
May 29, 2019Class IZ-2238-2019Software designPotential for Quantum Pump Console, part of the Quantum Perfusion Systems, to unexpectedly shut down while in use.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has a recall.
  • 2 other devices also cite its predicate K060053.

How does its category compare?

All DTQ devices

DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass: FDA decisions, review times, recalls and adverse event reports over 20 years. 2018, this device’s year, is highlighted.

FDA decisions per year · DTQ

3in 2023

2007: 22008: 12009: 32010: 12011: 22014: 22015: 22017: 22018: 12019: 12021: 22023: 3
2007high 3 (2009) · low 0 (2012)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DTQ

292in 2023

2009: 86 days median2023: 292 days median
2007high 292 (2023) · low 86 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DTQ

1in 2024

2007: 92008: 172009: 12010: 172011: 222012: 422013: 42014: 12018: 12019: 1, 1 Class I2022: 12023: 7, 1 Class I2024: 1
2007high 42 (2012) · low 0 (2015)2026

Class I (2 since 2007) Class II/III not classified

Adverse event reports per year · DTQ

1,014in 2025

2007: 4202008: 5482009: 5892010: 8112011: 1,0812012: 1,6262013: 9062014: 1,2902015: 1,1842016: 7142017: 5272018: 6762019: 8572020: 6272021: 7292022: 7172023: 8182024: 8632025: 1,014
2007high 1,626 (2012) · low 420 (2007)2025

MAUDE reports naming product code DTQ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—9420
20081—17548
20093861589
20101—17811
20112—221,081
20120—421,626
20130—4906
20142—11,290
20152—01,184
20160—0714
20172—0527
20181—1676
20191—1 (1)857
20200—0627
20212—0729
20220—1717
202332927 (1)818
20240—1863
20250—01,014
20260—0—

What is it indicated for?

The Quantum Pump Console is indicated for use for up to 6 hours in cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K173834: Quantum Pump Console.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K173834

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026