510(k) Report

510(k) · Cleared

Cleared

VERIFY Assert Self-Contained Biological Indicator

K181442510(k) SpecialCleared Jun 13, 201812-day reviewClass II21 CFR 880.2805

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

STERIS Corporation · OWP Biological Sterilization Process Indicator With Recombinant-Dna Plasmid

Cleared on Jun 13, 2018 as substantially equivalent to K162701 (STERIS Corporation, 2017) after a 12-day review. Its family tree reaches back 2 generations to 2013.

Decided
Jun 13, 2018
Review
12 days
Class
Class II
Product code
OWP
Recalls
None
Predicates
1

Its longest line of descent runs back 2 generations to DEN110006 (2013). No later device cites it as a predicate yet.

  1. Root · 2 generations backDEN110006 Verify Cronos Self-Contained Biological IndicatorSTERIS Corporation · 2013
  2. 1 generation backK162701 VERIFY Assert Self-Contained Biological IndicatorSTERIS Corporation · 2017
  3. This deviceK181442 VERIFY Assert Self-Contained Biological IndicatorSTERIS Corporation · 2018
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2DEN110006Verify Cronos Self-Contained Biological IndicatorSTERIS Corporation2013—
−1K162701VERIFY Assert Self-Contained Biological IndicatorSTERIS Corporation2017—
0 (this device)K181442VERIFY Assert Self-Contained Biological IndicatorSTERIS Corporation2018—
FIG. 1 — Lineage of K181442, 2013–2018Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in OWP? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K162701VERIFY Assert Self-Contained Biological Indicator
Summary page 5
…tor with Recombinant-DNA Plasmid (21 CFR 880.2805, OWP) 2. Predicate Device VERIFY Assert Self-Contained Biological Indicator, K162701 3. Description of Device The product is intended to monitor the critical parameters of steam sterilization cycles…
STERIS Corporation2017—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 5 years and passes through 1 company; a new generation was cleared every 3.7 years on median.
  • None of the 2 ancestors has a recall.

How does its category compare?

All OWP devices

OWP Biological Sterilization Process Indicator With Recombinant-Dna Plasmid: FDA decisions, review times, recalls and adverse event reports over 20 years. 2018, this device’s year, is highlighted.

FDA decisions per year · OWP

1in 2018

2013: 12017: 32018: 1
2007high 3 (2017) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OWP

177in 2017

2017: 177 days median
2007high 177 (2017)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · OWP

No recall names OWP since 2007.

Adverse event reports per year · OWP

2in 2022

2017: 12018: 22022: 2
2007high 2 (2018) · low 0 (2007)2025

MAUDE reports naming product code OWP, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20100—00
20110—00
20120—00
20131—00
20140—00
20150—00
20160—00
2017317701
20181—02
20190—00
20200—00
20210—00
20220—02
20230—00
20240—00
20250—00
20260—0—

What is it indicated for?

The VERIFY Assert Self-Contained Biological Indicator (SCBI) is for routine monitoring, qualification testing and product testing of the following steam sterilization processes.

Cycle Type Temperature Time

Dynamic Air Removal 270°F (132°C) 4 minutes

Dynamic Air Removal 275°F (135°C) 3 minutes

Gravity 250°F (121°C) 30 minutes

Gravity 270°F (132°C) 15 minutes

When used in conjunction with the Reader for the VERIFY Assert Self-Contained Biological Indicator, the VERIFY Assert Self-Contained Indicator provides a fluorescent result within 40 minutes

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K181442: VERIFY Assert Self-Contained Biological Indicator.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K181442

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026