510(k) Report

510(k) · Cleared

Recalled · Class II · 2021

FilmArray Blood Culture Identification (BCID) Panel

K181493510(k) SpecialCleared Jul 5, 201829-day reviewClass II21 CFR 866.3365Recalls 6

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Biofire Diagnostics, LLC · PEN Gram-Negative Bacteria And Associated Resistance Markers

Cleared on Jul 5, 2018 after a 29-day review. The FDA summary names no predicate by number. 2 later devices descend from it. It has been recalled 6 times.

Decided
Jul 5, 2018
Review
29 days
Class
Class II
Product code
PEN
Recalls
6
Predicates
0

2 later devices descend from it, 1 citing it directly.

Descendants, by year cleared 2020: 1 descendant cleared2024: 1 descendant cleared201620182020202220242026
  1. This deviceK181493 FilmArray Blood Culture Identification (BCID) PanelBiofire Diagnostics, LLC · 2018
  2. Descendant · cites it directly · 1 of its ownK193519 BioFire Blood Culture Identification 2 (BCID2) PanelBiofire Diagnostics, LLC · 2020 · Recalled
  3. Descendant · 2 generations on · 0 of its ownK243759 BIOFIRE Blood Culture Identification 2 (BCID2) Panel (RFIT-ASY-0147…Biofire Diagnostics, LLC · 2024

Most influential descendants

Showing 2 of 2 · See all
  • K193519BioFire Blood Culture Identification 2 (BCID2) PanelBiofire Diagnostics, LLC · 2020 · cites it directly1descendant
  • K243759BIOFIRE Blood Culture Identification 2 (BCID2) Panel (RFIT-ASY-0147 (30 pack)…Biofire Diagnostics, LLC · 2024 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K181493FilmArray Blood Culture Identification (BCID) PanelBiofire Diagnostics, LLC2018—
+1K193519BioFire Blood Culture Identification 2 (BCID2) PanelBiofire Diagnostics, LLC20207
+2K243759BIOFIRE Blood Culture Identification 2 (BCID2) Panel (RFIT-ASY-0147 (30 pack)…Biofire Diagnostics, LLC2024—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202010
202410
FIG. 1 — Lineage of K181493, 2018–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in PEN? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new PEN device

  1. #1 K262168 QIAstat-Dx BCID GN Plus AMR Panelscore 12.1 · 1 generation · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

The FDA summary names no predicate by number.

What descends from it? (2)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K193519BioFire Blood Culture Identification 2 (BCID2) PanelBiofire Diagnostics, LLC20201

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jul 20, 2021Class IIZ-2245-2021Process controlDue to product complaints received for false negative results (dropouts) with blood culture identification panel potential caused by a pouch manufacturing error.
Jan 26, 2021Class IIZ-1157-2021OtherThe firm has identified an increased risk of false positive Pseudomonas aeruginosa when BCID/BCID2 Panels are used with specific BD BACTEC" blood culture vials.
Jan 22, 2020Class IIZ-2296-2020Under Investigation by firmThere is an increased risk of false positive Proteus results when using the Film Array Blood Culture Identification Panel (BCID) in vitro diagnostic test of blood cultures used in conjunction with specific media types. False positive…
Jan 22, 2020Class IIZ-2295-2020Under Investigation by firmThere is an increased risk of false positive Proteus results when using the Film Array Blood Culture Identification Panel (BCID) in vitro diagnostic test of blood cultures used in conjunction with specific media types. False positive…
Mar 7, 2019Class IIZ-1133-2019Material/Component ContaminationIncreased risk of false positive Enterobacteriaceae and Escherichia coli results on affected blood culture identification panels when using certain blood culture bottles.
Sep 5, 2018Class IIZ-0296-2019Material/Component ContaminationIncreased risk of false positive results for Proteus when the product is used with a specific blood culture bottles.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K193519 (Biofire Diagnostics, LLC, 2020) has gone furthest, with 1 descendant of its own.

How does its category compare?

All PEN devices

PEN Gram-Negative Bacteria And Associated Resistance Markers: FDA decisions, review times, recalls and adverse event reports over 20 years. 2018, this device’s year, is highlighted.

FDA decisions per year · PEN

1in 2025

2013: 12014: 12015: 12016: 12018: 12019: 22020: 12022: 12024: 12025: 12026 (partial year): 1
2007high 2 (2019) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · PEN

Too few PEN 510(k)s in any year since 2007 to give a median.

Recalls per year · PEN

1in 2023

2020: 12021: 12023: 1
2007high 1 (2020) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · PEN

13in 2025

2018: 12019: 52020: 42021: 152022: 52023: 12024: 82025: 13
2007high 15 (2021) · low 0 (2007)2025

MAUDE reports naming product code PEN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20100—00
20110—00
20120—00
20131—00
20141—00
20151—00
20161—00
20170—00
20181—01
20192—05
20201—14
20210—115
20221—05
20230—11
20241—08
20251—013
20261—0—
Cite this page

510(k) Report. “K181493: FilmArray Blood Culture Identification (BCID) Panel.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K181493

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026