510(k) Report

510(k) · Cleared

Cleared

Codman Certas Plus Electronic Tool Kit

K181902510(k) TraditionalCleared Oct 25, 2018101-day reviewClass II21 CFR 882.5550

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Integra Lifesciences Corp. · JXG Shunt, Central Nervous System And Components

Cleared on Oct 25, 2018 as substantially equivalent to K143111 (Medos International SARL, 2015) after a 101-day review. Its family tree reaches back 9 generations to 1995.

Decided
Oct 25, 2018
Review
101 days
Class
Class II
Product code
JXG
Recalls
None
Predicates
1

Its longest line of descent runs back 9 generations to K944222 (1995). 16 ancestors in all, the oldest K861377 (1986). No later device cites it as a predicate yet.

  1. Root · 9 generations backK944222 Codman-Medos Valve Shunt SystemJohnson & Johnson Professionals, Inc. · 1995
  2. 5 generations between
  3. 3 generations backK060681 Medtronic PS Medical Strata Type ValueMedtronic Neurosurgery · 2006 · Recalled
  4. 2 generations backK112156 Codman Certas Programmable Valve System, Codman Certas Therapy…Medos International SARL · 2011 · Recalled
  5. 1 generation backK143111 Codman Certas Plus Programmable Valve; Codman Certas Toll KitMedos International SARL · 2015
  6. This deviceK181902 Codman Certas Plus Electronic Tool KitIntegra Lifesciences Corp. · 2018
Show all 9 generations
  1. Root · 9 generations backK944222 Codman-Medos Valve Shunt SystemJohnson & Johnson Professionals, Inc. · 1995
  2. 8 generations backK973774 Codman Hakim Micro Precision ValveJohnson & Johnson Professionals, Inc. · 1997
  3. 7 generations backK980778 Hakim Micro Programmable Valve SystemJohnson & Johnson Professionals, Inc. · 1998
  4. 6 generations backK012052 Medtronic PS Medical Strata Valve, Models 42855, 42865, 92865, 92855Medtronic PS Medical · 2002
  5. 5 generations backK033850 Medtronic PS Medical Strata NSC Valve and Shunt Assemblies with and…Medtronic Neurosurgery · 2004 · Recalled
  6. 4 generations backK042465 PS Medical Strata II ValveMedtronic Neurosurgery · 2004 · Recalled
  7. 3 generations backK060681 Medtronic PS Medical Strata Type ValueMedtronic Neurosurgery · 2006 · Recalled
  8. 2 generations backK112156 Codman Certas Programmable Valve System, Codman Certas Therapy…Medos International SARL · 2011 · Recalled
  9. 1 generation backK143111 Codman Certas Plus Programmable Valve; Codman Certas Toll KitMedos International SARL · 2015
  10. This deviceK181902 Codman Certas Plus Electronic Tool KitIntegra Lifesciences Corp. · 2018
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−9K944222Codman-Medos Valve Shunt SystemJohnson & Johnson Professionals, Inc.1995—
−8K973774Codman Hakim Micro Precision ValveJohnson & Johnson Professionals, Inc.1997—
−7K980778Hakim Micro Programmable Valve SystemJohnson & Johnson Professionals, Inc.1998—
−6K012052Medtronic PS Medical Strata Valve, Models 42855, 42865, 92865, 92855Medtronic PS Medical2002—
−5K033850Medtronic PS Medical Strata NSC Valve and Shunt Assemblies with and Without…Medtronic Neurosurgery20042
−4K042465PS Medical Strata II ValveMedtronic Neurosurgery20044
−3K060681Medtronic PS Medical Strata Type ValueMedtronic Neurosurgery20062
−2K112156Codman Certas Programmable Valve System, Codman Certas Therapy Management SystemMedos International SARL201116
−1K143111Codman Certas Plus Programmable Valve; Codman Certas Toll KitMedos International SARL2015—
0 (this device)K181902Codman Certas Plus Electronic Tool KitIntegra Lifesciences Corp.2018—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198610
199010
199510
199810
200010
200410
200910
FIG. 1 — Lineage of K181902, 1995–2018Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JXG? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JXG device

  1. #1 K243531 Codman Libertís Shunt Catheter with Bactiseal…score 11.9 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K143111Codman Certas Plus Programmable Valve; Codman Certas Toll Kit
Summary page 4
…Predicate The predicate device for this submission is: Device 510(k) Number Date Cleared Title K143111 February 19, 2015 Codman Certas Tool Kit IV. Device The Certas Plus Electronic Tool Kit is used t…
Medos International SARL2015—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 23 years and passes through 5 companies; a new generation was cleared every 2.4 years on median.
  • 6 of the 16 ancestors have a recall of any class.

How does its category compare?

All JXG devices

JXG Shunt, Central Nervous System And Components: FDA decisions, review times, recalls and adverse event reports over 20 years. 2018, this device’s year, is highlighted.

FDA decisions per year · JXG

8in 2025

2007: 32008: 62009: 52010: 32011: 72012: 32013: 32015: 62016: 62017: 72018: 62019: 42020: 22021: 12022: 42023: 52024: 32025: 82026 (partial year): 2
2007high 8 (2025) · low 0 (2014)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JXG

255in 2025

2007: 184 days median2008: 60 days median2009: 106 days median2010: 30 days median2011: 136 days median2012: 78 days median2013: 108 days median2015: 276 days median2016: 143 days median2017: 90 days median2018: 102 days median2019: 101 days median2022: 101 days median2023: 134 days median2024: 90 days median2025: 255 days median
2007high 276 (2015) · low 30 (2010)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JXG

3in 2024

2008: 12009: 302011: 52012: 12013: 142014: 62017: 12020: 82023: 22024: 32026 (partial year): 1
2007high 30 (2009) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JXG

1,198in 2025

2007: 6492008: 6302009: 6292010: 8122011: 7872012: 9772013: 1,0042014: 1,2512015: 9842016: 1,0912017: 1,3352018: 1,4842019: 1,2952020: 1,2062021: 1,4672022: 1,3292023: 1,2212024: 1,2142025: 1,198
2007high 1,484 (2018) · low 629 (2009)2025

MAUDE reports naming product code JXG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200731840649
20086601630
2009510630629
20103300812
201171365787
20123781977
20133108141,004
20140—61,251
201562760984
2016614301,091
201779011,335
2018610201,484
2019410101,295
20202—81,206
20211—01,467
2022410101,329
2023513421,221
202439031,214
2025825501,198
20262—1—

What is it indicated for?

The Codman Certas Plus Electronic Tool Kit allows the non-invasive reading or adjustment of the valve setting for the Codman Certas and Certas Plus Programmable Valves.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K181902: Codman Certas Plus Electronic Tool Kit.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K181902

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026