510(k) Report

510(k) · Cleared

Recalled · Class II · 2026

Centricity Universal Viewer

K182419510(k) TraditionalCleared Dec 14, 2018100-day reviewClass II21 CFR 892.2050Recalls 8

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

GE Healthcare · LLZ System, Image Processing, Radiological

Cleared on Dec 14, 2018 as substantially equivalent to K150420 (GE Healthcare, 2015) after a 100-day review. Its family tree reaches back 8 generations to 1994. 2 later devices descend from it. It has been recalled 8 times.

Decided
Dec 14, 2018
Review
100 days
Class
Class II
Product code
LLZ
Recalls
8
Predicates
1

Its longest line of descent runs back 8 generations to K942120 (1994). 15 ancestors in all. 2 later devices descend from it, 1 citing it directly.

  1. Root · 8 generations backK942120 Advantage Windows Diagnostic X-Ray WorkstationGE Medical Systems · 1994
  2. 7 generations backK013630 Isite RadiologyStentor, Inc. · 2001
  3. 6 generations backK031311 Integradweb by Dynamic Imaging, Inc.Dynamic Imaging, Inc. · 2003
  4. 5 generations backK042313 Integradweb MPR/MipDynamic Imaging, Inc. · 2004
  5. 4 generations backK082318 Centricity Pacs IWGe Healthcare Dynamic Imaging Solutions · 2008 · Recalled
  6. 3 generations backK121387 Centricity Pacs-IWGe Healthcare Integrated IT Solutions · 2012 · Recalled
  7. 2 generations backK123174 Centricity Pacs-IW with Universal ViewerGE Healthcare · 2012 · Recalled
  8. 1 generation backK150420 Centricity Universal ViewerGE Healthcare · 2015 · Recalled
  9. This deviceK182419 Centricity Universal ViewerGE Healthcare · 2018
  10. Descendant · cites it directly · 1 of its ownK211312 Universal ViewerGE Healthcare · 2021 · Recalled
  11. Descendant · 2 generations on · 0 of its ownK253639 ViewGE Healthcare · 2026

Most influential descendants

Showing 2 of 2 · See all
  • K211312Universal ViewerGE Healthcare · 2021 · cites it directly1descendant
  • K253639ViewGE Healthcare · 2026 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−8K942120Advantage Windows Diagnostic X-Ray WorkstationGE Medical Systems1994—
−7K013630Isite RadiologyStentor, Inc.2001—
−6K031311Integradweb by Dynamic Imaging, Inc.Dynamic Imaging, Inc.2003—
−5K042313Integradweb MPR/MipDynamic Imaging, Inc.2004—
−4K082318Centricity Pacs IWGe Healthcare Dynamic Imaging Solutions20087
−3K121387Centricity Pacs-IWGe Healthcare Integrated IT Solutions20121
−2K123174Centricity Pacs-IW with Universal ViewerGE Healthcare201210
−1K150420Centricity Universal ViewerGE Healthcare201515
0 (this device)K182419Centricity Universal ViewerGE Healthcare2018—
+1K211312Universal ViewerGE Healthcare20212
+2K253639ViewGE Healthcare2026—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199610
200010
200320
200410
200510
200710

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202110
202610
FIG. 1 — Lineage of K182419, 1994–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LLZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LLZ device

  1. #1 K230955 syngo Application Softwarescore 21.5 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K150420Centricity Universal Viewer
Summary page 5
…Page 5 === GE Healthcare K182419, Page 2 of 13 III. PREDICATE DEVICE Primary - K150420 (May 2015)- Centricity Universal Viewer by GE Healthcare Reference - K063628 (December 2006) -Centricity Radiology RA600, Cardiol…
GE Healthcare201515
†K063628Centricity Radiology RA600, Cardiology CA1000 and Digital HardcopyGe Healthcare Integrated IT Solutions20063

What descends from it? (2)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K211312Universal ViewerGE Healthcare20211

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 30, 2026Class IIZ-1569-2026Software designThere is a potential cybersecurity vulnerability affecting certain versions of Centricity Universal Viewer. User login credentials may be exposed on the local client workstation, which could allow an unauthorized individual to potentially…
Feb 19, 2025Class IIZ-1483-2025Software designThere is a potential security vulnerability in Centricity Universal Viewer (UV), Centricity PACS-IW (PACS-IW), Centricity Radiology RA600 (RA600), & Centricity Cardiology CA1000 (CA1000) where the service login credentials are able to be…
Sep 8, 2023Class IIZ-0142-2024Software designWhen using Universal Viewer or Centricity Universal Viewer with Workflow Manager (WFM), if a third-party reporting application is used to launch exams, it could display information on a different patient than WFM and the Viewer.
Jul 15, 2022Class IIZ-1566-2022Process designInaccurate Distance and Area measurements with use of Global Stack viewport.
Aug 12, 2021Class IIZ-0012-2022OtherA software anomaly exists in the Centricity Universal Viewer study management feature in which study changes are not propagated to either the Centricity Enterprise Archive (EA) or another Vendor Neutral Archive (VNA).
Mar 19, 2021Class IIZ-1533-2021Under Investigation by firmWhen the Cross Enterprise Display option is enabled on the Centricity Universal Viewer and Workflow Manager option is used as the PACS worklist launching the viewer, the patient jacket may potentially include a different patient s…
Dec 16, 2019Class IIZ-0819-2020Software designCentricity Universal Viewer measurements saved into a DICOM Grayscale Presentation State are incorrect in subsequent views for exams containing series with different pixel sizes and may lead to a potential misdiagnosis.
May 16, 2019Class IIZ-1910-2019Device DesignThere is the possibility of viewing studies directly from the Enterprise Archive or VNA with incorrect patient images because the updated series or study is not present in the archive.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 24 years and passes through 6 companies; a new generation was cleared every 3.6 years on median.
  • 6 of the 15 ancestors have a recall of any class.
  • Of its descendants, K211312 (GE Healthcare, 2021) has gone furthest, with 1 descendant of its own.

How does its category compare?

All LLZ devices

LLZ System, Image Processing, Radiological: FDA decisions, review times, recalls and adverse event reports over 20 years. 2018, this device’s year, is highlighted.

FDA decisions per year · LLZ

58in 2025

2007: 762008: 942009: 702010: 512011: 1062012: 822013: 952014: 782015: 642016: 822017: 782018: 1042019: 742020: 812021: 832022: 652023: 872024: 622025: 582026 (partial year): 45
2007high 106 (2011) · low 51 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LLZ

173in 2025

2007: 36 days median2008: 52 days median2009: 62 days median2010: 98 days median2011: 79 days median2012: 87 days median2013: 84 days median2014: 89 days median2015: 90 days median2016: 90 days median2017: 107 days median2018: 105 days median2019: 107 days median2020: 115 days median2021: 116 days median2022: 137 days median2023: 157 days median2024: 131 days median2025: 173 days median2026 (partial year): 110 days median
2007high 173 (2025) · low 36 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LLZ

11in 2025

2007: 202008: 322009: 212010: 172011: 302012: 122013: 642014: 62015: 162016: 122017: 52018: 72019: 22020: 62021: 32022: 42023: 82024: 62025: 112026 (partial year): 14
2007high 64 (2013) · low 2 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LLZ

116in 2025

2007: 1522008: 1542009: 1452010: 1452011: 4542012: 2662013: 1,1302014: 15,6312015: 2112016: 2902017: 1042018: 722019: 882020: 822021: 702022: 282023: 262024: 422025: 116
2007high 15,631 (2014) · low 26 (2023)2025

MAUDE reports naming product code LLZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007763620152
2008945232154
2009706221145
2010519817145
20111067930454
2012828712266
20139584641,130
20147889615,631
2015649016211
2016829012290
2017781075104
2018104105772
201974107288
202081115682
202183116370
202265137428
202387157826
202462131642
20255817311116
20264511014—

What is it indicated for?

Centricity Universal Viewer is a device that displays medical images and data from various imaging sources, and from other healthcare information sources. Medical images and data can be displayed, communicated, stored, and processed. Typical users of this system are authorized healthcare professionals. Centricity Universal Viewer is intended to assist in the viewing, analysis, diagnostic interpretation, and sharing of images and other information.…

Read the full statement (122 words)

Centricity Universal Viewer is a device that displays medical images and data from various imaging sources, and from other healthcare information sources. Medical images and data can be displayed, communicated, stored, and processed. Typical users of this system are authorized healthcare professionals.

Centricity Universal Viewer is intended to assist in the viewing, analysis, diagnostic interpretation, and sharing of images and other information.

Mammography images may only be interpreted using a monitor compliant with requirements of local regulations and must meet other technical specifications reviewed and accepted by the local regulatory agencies.

Contraindications:

Centricity Universal Viewer is contraindicated for the use of lossy compressed mammographic images. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K182419: Centricity Universal Viewer.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K182419

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026