Zenition 70

K183040510(k) AbbreviatedCleared Dec 6, 201835-day reviewClass II21 CFR 892.1650Recalls 5

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Philips Medical Systems Nederland B.V. · OWB Interventional Fluoroscopic X-Ray System

Cleared on Dec 6, 2018 as substantially equivalent to K142708 (Philips Medical Systems Nethrlands BV, 2014) after a 35-day review. Its family tree reaches back 5 generations to 2004. 6 later devices descend from it. It has been recalled 5 times.

Decided
Dec 6, 2018
Review
35 days
Class
Class II
Product code
OWB
Recalls
5
Predicates
1

Its longest line of descent runs back 5 generations to K042646 (2004). 6 later devices descend from it, 2 citing it directly.

Descendants, by year cleared 2022: 1 descendant cleared2021: 1 descendant cleared2024: 2 descendants cleared2025: 1 descendant cleared2026: 1 descendant clearedK042646 · Dynact · Siemens Medical Systems, Inc. · 2004K061685 · Pulsera · Philips Medical Systems North America Co. · 2006K090590 · Veradius, Model 718-130 · Philips Medical Systems North America Co. · 2009K133819 · Veradius · Philips Medical Systems Nederland B.V. · 2014K142708 · Veradius Unity · Philips Medical Systems Nethrlands BV · 2014 · recalled200020052010201520202025
  1. Root · 5 generations backK042646 DynactSiemens Medical Systems, Inc. · 2004
  2. 4 generations backK061685 PulseraPhilips Medical Systems North America Co. · 2006
  3. 3 generations backK090590 Veradius, Model 718-130Philips Medical Systems North America Co. · 2009
  4. 2 generations backK133819 VeradiusPhilips Medical Systems Nederland B.V. · 2014
  5. 1 generation backK142708 Veradius UnityPhilips Medical Systems Nethrlands BV · 2014 · Recalled
  6. This deviceK183040 Zenition 70Philips Medical Systems Nederland B.V. · 2018
  7. Descendant · cites it directly · 3 of its ownK212813 Zenition 70Philips Medical Systems Nederland B.V. · 2021 · Recalled
  8. Descendant · 2 generations on · 1 of its ownK232420 Zenition 30Philips Medical Systems Nederland B.V. · 2024
  9. Descendant · cites it directly · 1 of its ownK222080 Garion Image-Intensified Fluoroscopic X-Ray System, MobileSg Healthcare Co, Ltd. · 2022
  10. Descendant · 3 generations on · 0 of its ownK254205 Zenition 30Philips Medical Systems Nederland B.V. · 2026
  11. Descendant · 2 generations on · 0 of its ownK243411 Diagnostic X-ray SystemNanjing Perlove Medical Equipment Co., Ltd. · 2025

Showing 5 of 6 descendants. See all descendants

Descendants cleared per year

2021: 12022: 12024: 22025: 12026: 1
2021high 2 (2024) · low 0 (2023)2026

Most influential descendants

Showing 5 of 6 · See all
  • K212813Zenition 70Philips Medical Systems Nederland B.V. · 2021 · cites it directly3descendants
  • K232420Zenition 30Philips Medical Systems Nederland B.V. · 2024 · 2 generations on1descendant
  • K222080Garion Image-Intensified Fluoroscopic X-Ray System, MobileSg Healthcare Co, Ltd. · 2022 · cites it directly1descendant
  • K254205Zenition 30Philips Medical Systems Nederland B.V. · 2026 · 3 generations on0descendants
  • K243411Diagnostic X-ray SystemNanjing Perlove Medical Equipment Co., Ltd. · 2025 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K042646DynactSiemens Medical Systems, Inc.2004—
−4K061685PulseraPhilips Medical Systems North America Co.2006—
−3K090590Veradius, Model 718-130Philips Medical Systems North America Co.2009—
−2K133819VeradiusPhilips Medical Systems Nederland B.V.2014—
−1K142708Veradius UnityPhilips Medical Systems Nethrlands BV20144
0 (this device)K183040Zenition 70Philips Medical Systems Nederland B.V.2018—
+1K212813Zenition 70Philips Medical Systems Nederland B.V.20214
+2K232420Zenition 30Philips Medical Systems Nederland B.V.2024—
+1K222080Garion Image-Intensified Fluoroscopic X-Ray System, MobileSg Healthcare Co, Ltd.2022—
+3K254205Zenition 30Philips Medical Systems Nederland B.V.2026—
+2K243411Diagnostic X-ray SystemNanjing Perlove Medical Equipment Co., Ltd.2025—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202110
202210
202420
202510
202610
FIG. 1 — Lineage of K183040, 2004–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in OWB? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new OWB device

  1. #1 K251602 Alphenix, INFX-8000V/B, INFX-8000V/S, V9.6…score 18.3 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K142708Veradius Unity
Summary page 5
…Veradius Unity Predicate Manufacturer: Philips Medical Systems Nederland B.V. Device: 510(k) Clearance: K142708 Classification Name: Image-intensified fluoroscopic x-ray system Classification…
Philips Medical Systems Nethrlands BV20144

Also cleared against K142708: 2 other devices →

What descends from it? (6)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K212813Zenition 70Philips Medical Systems Nederland B.V.20213
K222080Garion Image-Intensified Fluoroscopic X-Ray System, MobileSg Healthcare Co, Ltd.20221

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 25, 2025Class IIZ-1581-2025Labeling designA wireless foot switch pedal may get stuck in the active position when the user releases the pedal, resulting in the emission of unintended radiation.
Oct 4, 2023Class IIZ-0226-2024Device DesignPhilips has become aware of an issue with the Philips Integris-Allura, Allura Xper and Azurion systems that have been delivered to Puerto Rico without a warning label and the Source-to-Skin Distance Spacer (Tube-spacer) is missing.
Jul 20, 2023Class IIZ-1583-2025Software designPhilips has become aware of the potential for unintended radiation exposure with Zenition 70 systems with Vascular Extension option and a Wireless Foot Switch.
Dec 2, 2021Class IIZ-0648-2022Under Investigation by firmFirm discovered that the wireless foot switch can suddenly stop responding due to a firmware issue. If the wireless foot switch fails, there could be a delay or interruption of procedure if the hand switch is not used or if the wired foot…
Apr 10, 2020Class IIZ-1933-2020Device DesignWhen strain relief is lost at the stand side of the Stand MVS Trolley Cable, over time one or more electrical contacts could break due to pulling, twisting, or dropping of the connector. If this happens, live parts may be exposed (230 V)…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 14 years and passes through 4 companies; a new generation was cleared every 2.5 years on median.
  • 1 of the 5 ancestors has a recall of any class.
  • Of its descendants, K212813 (Philips Medical Systems Nederland B.V., 2021) has gone furthest, with 3 descendants of its own.
  • 2 other devices also cite its predicate K142708.

How does its category compare?

All OWB devices

OWB Interventional Fluoroscopic X-Ray System: FDA decisions, review times, recalls and adverse event reports over 20 years. 2018, this device’s year, is highlighted.

FDA decisions per year · OWB

22in 2025

2007: 32008: 42009: 92010: 72011: 62012: 62013: 112014: 152015: 112016: 162017: 202018: 232019: 92020: 162021: 242022: 212023: 122024: 132025: 222026 (partial year): 15
2007high 24 (2021) · low 3 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OWB

148in 2025

2007: 52 days median2008: 135 days median2009: 99 days median2010: 66 days median2011: 79 days median2012: 37 days median2013: 119 days median2014: 158 days median2015: 66 days median2016: 103 days median2017: 76 days median2018: 54 days median2019: 174 days median2020: 53 days median2021: 60 days median2022: 90 days median2023: 152 days median2024: 126 days median2025: 148 days median2026 (partial year): 221 days median
2007high 221 (2026) · low 37 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · OWB

42so far in 2026, the most since 2007

2008: 12013: 102014: 92015: 352016: 132017: 132018: 62019: 72020: 82021: 312022: 242023: 222024: 112025: 302026 (partial year): 42
2007high 42 (2026) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · OWB

14,008in 2025

2007: 12011: 12012: 12013: 272014: 1032015: 2,1342016: 4,0742017: 1,6042018: 1,4492019: 1,5582020: 1,5812021: 1,9812022: 2,9322023: 7,6042024: 7,9422025: 14,008
2007high 14,008 (2025) · low 0 (2008)2025

MAUDE reports naming product code OWB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200735201
2008413510
200999900
201076600
201167901
201263701
2013111191027
2014151589103
20151166352,134
201616103134,074
20172076131,604
2018235461,449
2019917471,558
2020165381,581
20212460311,981
20222190242,932
202312152227,604
202413126117,942
2025221483014,008
20261522142—

What is it indicated for?

The Zenition 70 device is intended to be used and operated by: adequately trained, qualified and authorized health care professionals who have full understanding of the safety information and emergency procedures as well as the capabilities and functions of the device. The device is used for radiological guidance and visualization during diagnostic, interventional and surgical procedures on all patients, except neonates (birth to one month), within the…

Read the full statement (112 words)

The Zenition 70 device is intended to be used and operated by: adequately trained, qualified and authorized health care professionals who have full understanding of the safety information and emergency procedures as well as the capabilities and functions of the device.

The device is used for radiological guidance and visualization during diagnostic, interventional and surgical procedures on all patients, except neonates (birth to one month), within the limits of the device. The device is to be used in health care facilities both inside and outside the operating room, sterile as well as non-sterile environment in a variety of procedures.

Applications:

• Orthopedic

• Neuro

• Abdominal

• Vascular

• Thoracic

• Cardiac

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K183040: Zenition 70.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K183040

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026