510(k) Report

510(k) · Cleared

Cleared

PRO-LITE Sterilization Tray

K183301510(k) TraditionalCleared Jan 3, 201937-day reviewClass II21 CFR 880.6850

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

STERIS Corporation · KCT Sterilization Wrap Containers, Trays, Cassettes & Other Accessories

Cleared on Jan 3, 2019 as substantially equivalent to K172755 (STERIS Corporation, 2018) after a 37-day review. Its family tree reaches back 10 generations to 1984.

Decided
Jan 3, 2019
Review
37 days
Class
Class II
Product code
KCT
Recalls
None
Predicates
1

Its longest line of descent runs back 10 generations to K844002 (1984). 16 ancestors in all. No later device cites it as a predicate yet.

  1. Root · 10 generations backK844002 HU-Friedy Ims Cassette SystemHu-Friedy Mfg. Co., Inc. · 1984
  2. 6 generations between
  3. 3 generations backK140494 V-Pro Sterilization Tray/ Sterilization MatSTERIS Corporation · 2014
  4. 2 generations backK160912 Sterilization TraySTERIS Corporation · 2016
  5. 1 generation backK172755 PRO-LITE Sterilization TraysSTERIS Corporation · 2018
  6. This deviceK183301 PRO-LITE Sterilization TraySTERIS Corporation · 2019
Show all 10 generations
  1. Root · 10 generations backK844002 HU-Friedy Ims Cassette SystemHu-Friedy Mfg. Co., Inc. · 1984
  2. 9 generations backK980065 C/T Med-Systems Cassette SystemC/T Med System, Ltd. · 1998
  3. 8 generations backK990761 Tuttnauer Mini Container Cassette, Tuttnauer Standard Case CassetteTuttnauer U.S.A Co, Ltd. · 1999
  4. 7 generations backK012105 Polyvac Surgical Instrument Delivery SystemSymmetry Medical, Inc. · 2002
  5. 6 generations backK023658 Sterilization Cases and TraysCarr Metal Products, Inc. · 2003
  6. 5 generations backK052992 Conmed Linvatec Instrument Sterilization TrayConmed Linvatec · 2006
  7. 4 generations backK070769 VHP 136 TraySTERIS Corporation · 2007
  8. 3 generations backK140494 V-Pro Sterilization Tray/ Sterilization MatSTERIS Corporation · 2014
  9. 2 generations backK160912 Sterilization TraySTERIS Corporation · 2016
  10. 1 generation backK172755 PRO-LITE Sterilization TraysSTERIS Corporation · 2018
  11. This deviceK183301 PRO-LITE Sterilization TraySTERIS Corporation · 2019
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−10K844002HU-Friedy Ims Cassette SystemHu-Friedy Mfg. Co., Inc.1984—
−9K980065C/T Med-Systems Cassette SystemC/T Med System, Ltd.1998—
−8K990761Tuttnauer Mini Container Cassette, Tuttnauer Standard Case CassetteTuttnauer U.S.A Co, Ltd.1999—
−7K012105Polyvac Surgical Instrument Delivery SystemSymmetry Medical, Inc.2002—
−6K023658Sterilization Cases and TraysCarr Metal Products, Inc.2003—
−5K052992Conmed Linvatec Instrument Sterilization TrayConmed Linvatec2006—
−4K070769VHP 136 TraySTERIS Corporation2007—
−3K140494V-Pro Sterilization Tray/ Sterilization MatSTERIS Corporation2014—
−2K160912Sterilization TraySTERIS Corporation2016—
−1K172755PRO-LITE Sterilization TraysSTERIS Corporation2018—
0 (this device)K183301PRO-LITE Sterilization TraySTERIS Corporation2019—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199410
199710
200110
200210
200410
201110
FIG. 1 — Lineage of K183301, 1984–2019Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KCT? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KCT device

  1. #1 K214041 Aesculap AICON Containerscore 18.1 · 6 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K172755PRO-LITE Sterilization Trays
Summary page 7
…21 CFR 880.6850 Class II Product Code KCT 2. Predicate Device PRO-LITE Sterilization Tray, K172755 3. Description of Device The PRO-LITE Sterilization Trays contain medical devices for sterilization in the V-PRO L…
STERIS Corporation2018—

Also cleared against K172755: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 35 years and passes through 7 companies; a new generation was cleared every 2.2 years on median.
  • None of the 16 ancestors has a recall.
  • 1 other device also cites its predicate K172755.

How does its category compare?

All KCT devices

KCT Sterilization Wrap Containers, Trays, Cassettes & Other Accessories: FDA decisions, review times, recalls and adverse event reports over 20 years. 2019, this device’s year, is highlighted.

FDA decisions per year · KCT

11in 2025

2007: 32008: 62009: 62010: 122011: 62012: 102013: 82014: 112015: 152016: 122017: 102018: 152019: 182020: 142021: 182022: 112023: 102024: 82025: 112026 (partial year): 9
2007high 18 (2019) · low 3 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KCT

205in 2025

2007: 199 days median2008: 182 days median2009: 89 days median2010: 202 days median2011: 117 days median2012: 236 days median2013: 131 days median2014: 169 days median2015: 151 days median2016: 182 days median2017: 211 days median2018: 192 days median2019: 187 days median2020: 134 days median2021: 185 days median2022: 137 days median2023: 94 days median2024: 106 days median2025: 205 days median2026 (partial year): 122 days median
2007high 236 (2012) · low 89 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KCT

17in 2022

2010: 12011: 22013: 12014: 12016: 32017: 12021: 12022: 17
2007high 17 (2022) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · KCT

50in 2025

2007: 22008: 82009: 52010: 32011: 32012: 22013: 62014: 52015: 62016: 162017: 32018: 222019: 152020: 662021: 262022: 82023: 302024: 182025: 50
2007high 66 (2020) · low 2 (2007)2025

MAUDE reports naming product code KCT, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007319902
2008618208
200968905
20101220213
2011611723
20121023602
2013813116
20141116915
20151515106
201612182316
20171021113
201815192022
201918187015
202014134066
202118185126
202211137178
20231094030
20248106018
202511205050
202691220—

What is it indicated for?

The PRO-LITE™ Sterilization Trays are used to contain medical devices for sterilization in the following cycles: Lumen, Non Lumen, Flexible, Fast Non Lumen and Fast Cycles of the V-PRO® Low Temperature Sterilization Systems Prior to placing in the Sterilizer, the trays must either be: • wrapped with a legally marketed sterilization wrap for use in the Sterilizers listed above or • placed inside a legally marketed pouch for enclosing trays in the…

Read the full statement (879 words)

The PRO-LITE™ Sterilization Trays are used to contain medical devices for sterilization in the following cycles:

Lumen, Non Lumen, Flexible, Fast Non Lumen and Fast Cycles of the V-PRO® Low Temperature Sterilization Systems

Prior to placing in the Sterilizer, the trays must either be:

• wrapped with a legally marketed sterilization wrap for use in the Sterilizers listed above

or

• placed inside a legally marketed pouch for enclosing trays in the Sterilizers listed above

The PRO-LITE Sterilization Trays are not intended to maintain sterility; they are intended to be used in conjunction with a validated, FDA-cleared sterilization wrap or pouch in order to maintain sterility of the enclosed medical instruments.

Intended Sterilization Cycles and Intended Tray Load for Tray Models: VP0045, VP0046, VP0047, VP0048, VP0049, VP0050, VP0050, VP0051, VP0052

V-PRO 60 & s2 Lumen Cycle

• Instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors

• Non-lumened devices including non-lumened rigid and semi-rigid endoscopes

• Medical devices, including single, dual and triple channeled rigid and semi-rigid endoscopes, with the following configurations:

• single or dual lumen devices

• ≥ 0.77 mm internal diameter (ID) and ≤ 410 mm in length

• triple lumen devices

• ≥ 1.2 mm ID and ≤ 275 mm in length

• ≥ 1.8 mm ID and ≤ 310 mm in length

or

• ≥ 2.8 mm ID and ≤ 317 mm in length

V-PRO 60 & s2 Non Lumen Cycle

Non-lumened devices including non-lumened rigid, semi-rigid and flexible endoscopes and non-lumened devices with diffusion-restricted spaces such as the hinged portion of forceps and scissors.

V-PRO 60 & s2 Flexible Cycle

Load 1: One flexible surgical endoscope or bronchoscope with a light cord (if not integral to endoscope) and mat without any additional load. The flexible endoscope may be a:

• single or dual lumen device with lumens that are ≥ 1 mm ID and ≤ 990 mm in length

Load 2: Non-lumened devices including non-lumened rigid, semi-rigid, and flexible endoscopes and non-lumened devices with diffusion-restricted areas such as the hinged portion of forceps or scissors. Medical devices, including rigid and semi-rigid endoscopes, with the following dimensions:

• ≥ 2 mm ID and ≤ 400 mm in length

• ≥ 0.76 mm ID and ≤ 233 mm in length

• ≥ 1.0 mm ID and ≤ 254 mm in length

Intended Sterilization Cycles and Intended Tray Load for Tray Models: VP0045, VP0046, VP0047, VP0048, VP0049

V-PRO s2 Fast Cycle

• Non-lumened devices including non-lumened rigid, semi-rigid and flexible endoscopes, and non-lumened devices with diffusion-restricted areas such as the hinged portion of forceps or scissors.

• Medical devices (including single, dual and triple channeled rigid and semi-rigid endoscopes) with the following configurations:

• Single or dual channeled devices with stainless steel lumens that are ≥ 0.77 mm ID and ≤ 410 mm in length

• Triple channeled devices with stainless steel lumens that are either:

• ≥ 1.2 mm ID and ≤ 275 mm in length

• ≥ 1.8 mm ID and ≤ 310 mm in length

or

• ≥ 2.8 mm ID and ≤ 317 mm in length

Intended Sterilization Cycles and Intended Tray Load for Tray Models: VP0045, VP0046, VP0047, VP0048, VP0049, VP0050, VP0050, VP0051, VP0052, VP0053

V-PRO 1, 1 Plus, maX & maX 2 Lumen Cycle

• Instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors

• Non-lumened devices including non-lumened rigid and semi-rigid endoscopes

• Medical devices, including single, dual and triple channeled rigid and semi-rigid endoscopes, with the following configurations:

• single or dual lumen devices

• ≥ 0.77 mm internal diameter (ID) and ≤ 410 mm in length

• triple lumen devices

• ≥ 1.2 mm ID and ≤ 275 mm in length

• ≥ 1.8 mm ID and ≤ 310 mm in length

or

• ≥ 2.8 mm ID and ≤ 317 mm in length

V-PRO 1, 1 Plus, maX & maX 2 Non Lumen Cycle

Non-lumened devices including non-lumened rigid, semi-rigid and flexible endoscopes and non-lumened devices with stainless steel or titanium diffusion-restricted spaces such as the hinged portion of forceps and scissors

V-PRO 1, 1 Plus, maX & maX 2 Flexible Cycle

Load 1: Single or dual lumen surgical flexible endoscopes (such as those used in ENT, Urology and Surgical Care) and bronchoscopes with a light cord (if not integral to endoscope) and mat with no additional load.

The flexible endoscopes may contain either:

• a single lumen that is ≥ 1 mm ID and ≤ 1050 mm in length

• or two lumens with:

• one lumen that is ≥ 1 mm ID and < 990 mm in length

• and the other lumen that is ≥ 1 mm ID and ≤ 850 mm in length

Load 2: Non-lumened devices including non-lumened rigid, semi-rigid, and flexible endoscopes and non-lumened devices with diffusion-restricted areas such as the hinged portion of forceps or scissors.

V-PRO maX 2 Fast Non Lumen Cycle

Non-lumened devices including non-lumened rigid, semi-rigid and flexible endoscopes and non-lumened devices with stainless steel or titanium diffusion-restricted spaces such as the hinged portion of forceps and scissors.

Instrument organizers are optional accessories intended to stabilize cylindrical medical instruments within the Sterilization Trays.

Sterilization mats are optional accessories intended to cushion and stabilize instruments within the Sterilization Trays.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K183301: PRO-LITE Sterilization Tray.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K183301

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026