Koios DS for Breast

K190442510(k) TraditionalCleared Jul 3, 2019128-day reviewClass II21 CFR 892.2060

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Koios Medical, Inc. · POK Computer-Assisted Diagnostic Software For Lesions Suspicious For Cancer

Cleared on Jul 3, 2019 as substantially equivalent to DEN170022 (Quantitative Insights, Inc., 2017) and 1 other predicate after a 128-day review. Its family tree reaches back 7 generations to 1991. 12 later devices descend from it.

Decided
Jul 3, 2019
Review
128 days
Class
Class II
Product code
POK
Recalls
None
Predicates
2

Its longest line of descent runs back 7 generations to K913770 (1991). 14 ancestors in all. 12 later devices descend from it, 2 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2017: 1 other ancestor2004: 2 other ancestors2002: 2 other ancestors2003: 1 other ancestor1994: 1 other ancestor2020: 1 descendant cleared2021: 1 descendant cleared2024: 4 descendants cleared2026: 3 descendants cleared2025: 3 descendants clearedK913770 · Advantage V.S. CT/MR Rad Diagnostic Workstation · General Electric Co. · 1991K960613 · GE Advantage Windows Review Workstation · GE Medical Systems · 1996K020483 · Advantage Workstation 4.1 · GE Medical Systems · 2002 · recalledK033400 · Seno Advantage · Ge Medical Systems W-400 · 2003 · recalledK043216 · Cadstream Version 4.0 · Confirma, Inc. · 2005K050846 · B-Cad System, Version 1.0 · The Medipattern Corporation · 2005K161959 · ClearView cCAD · Clearview Diagnostics, Inc. · 2016 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor3 generations back: 3 other ancestors4 generations back: 2 other ancestors7 generations back: 1 other ancestor 1 generation forward: 2 descendants2 generations forward: 6 descendants3 generations forward: 4 descendants K161959 · ClearView cCAD · Clearview Diagnostics, Inc. · 2016 · recalledK050846 · B-Cad System, Version 1.0 · The Medipattern Corporation · 2005K043216 · Cadstream Version 4.0 · Confirma, Inc. · 2005K033400 · Seno Advantage · Ge Medical Systems W-400 · 2003 · recalledK020483 · Advantage Workstation 4.1 · GE Medical Systems · 2002 · recalledK960613 · GE Advantage Windows Review Workstation · GE Medical Systems · 1996K913770 · Advantage V.S. CT/MR Rad Diagnostic Workstation · General Electric Co. · 1991−50
  1. Root · 7 generations backK913770 Advantage V.S. CT/MR Rad Diagnostic WorkstationGeneral Electric Co. · 1991
  2. 6 generations backK960613 GE Advantage Windows Review WorkstationGE Medical Systems · 1996
  3. 5 generations backK020483 Advantage Workstation 4.1GE Medical Systems · 2002 · Recalled
  4. 4 generations backK033400 Seno AdvantageGe Medical Systems W-400 · 2003 · Recalled
  5. 3 generations backK043216 Cadstream Version 4.0Confirma, Inc. · 2005
  6. 2 generations backK050846 B-Cad System, Version 1.0The Medipattern Corporation · 2005
  7. 1 generation backK161959 ClearView cCADClearview Diagnostics, Inc. · 2016 · Recalled
  8. This deviceK190442 Koios DS for BreastKoios Medical, Inc. · 2019
  9. Descendant · cites it directly · 6 of its ownK201555 EchoGo ProUltromics, Ltd. · 2020
  10. Descendant · cites it directly · 4 of its ownK212616 Koios DSKoios Medical, Inc. · 2021
  11. Descendant · 2 generations on · 2 of its ownK242342 Fetal EchoScanBrightheart · 2024
  12. Descendant · 3 generations on · 1 of its ownK251071 Fetal EchoScan (v1.1)Brightheart · 2025
  13. Descendant · 2 generations on · 1 of its ownK243614 Sonio SuspectSonio · 2025

Showing 5 of 12 descendants. See all descendants

Descendants cleared per year

2020: 12021: 12024: 42025: 32026: 3
2020high 4 (2024) · low 0 (2022)2026

Most influential descendants

Showing 5 of 12 · See all
  • K201555EchoGo ProUltromics, Ltd. · 2020 · cites it directly6descendants
  • K212616Koios DSKoios Medical, Inc. · 2021 · cites it directly4descendants
  • K242342Fetal EchoScanBrightheart · 2024 · 2 generations on2descendants
  • K251071Fetal EchoScan (v1.1)Brightheart · 2025 · 3 generations on1descendant
  • K243614Sonio SuspectSonio · 2025 · 2 generations on1descendant
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K913770Advantage V.S. CT/MR Rad Diagnostic WorkstationGeneral Electric Co.1991—
−6K960613GE Advantage Windows Review WorkstationGE Medical Systems1996—
−5K020483Advantage Workstation 4.1GE Medical Systems20022
−4K033400Seno AdvantageGe Medical Systems W-40020036
−3K043216Cadstream Version 4.0Confirma, Inc.2005—
−2K050846B-Cad System, Version 1.0The Medipattern Corporation2005—
−1K161959ClearView cCADClearview Diagnostics, Inc.20161
0 (this device)K190442Koios DS for BreastKoios Medical, Inc.2019—
+1K201555EchoGo ProUltromics, Ltd.2020—
+1K212616Koios DSKoios Medical, Inc.2021—
+2K242342Fetal EchoScanBrightheart2024—
+3K251071Fetal EchoScan (v1.1)Brightheart2025—
+2K243614Sonio SuspectSonio2025—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199410
200220
200310
200420
201710

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202010
202110
202440
202530
202630
FIG. 1 — Lineage of K190442, 1991–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in POK? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new POK device

  1. #1 K242342 Fetal EchoScanscore 16.7 · 9 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*DEN170022QuantX
Summary page 4
…erformance, this software medical device is substantially equivalent to the devices listed below: Model: QuantX Manufacturer: Quantitative Insights, Inc. 510(k) Number: DEN170022 510(k) Summary Koios Medical, Inc. 5-1…
Quantitative Insights, Inc.2017—
*K161959ClearView cCAD
Summary page 5
…ical, Inc. 5-1 Model: ClearView cCAD, Version 1.0 Manufacturer: ClearView Diagnostics, Inc. 510(k) Number: K161959 4. Device Description Koios Decision Support (DS) for Breast is a software application designed to assist trained interpreting p…
Clearview Diagnostics, Inc.20161

Also cleared against DEN170022: 3 other devices →

What descends from it? (12)

See all 12

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K201555EchoGo ProUltromics, Ltd.20206
K212616Koios DSKoios Medical, Inc.20214

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 28 years and passes through 7 companies; a new generation was cleared every 2.5 years on median.
  • 3 of the 14 ancestors have a recall of any class.
  • Of its descendants, K201555 (Ultromics, Ltd., 2020) has gone furthest, with 6 descendants of its own.
  • 3 other devices also cite its predicate DEN170022.

How does its category compare?

All POK devices

POK Computer-Assisted Diagnostic Software For Lesions Suspicious For Cancer: FDA decisions, review times, recalls and adverse event reports over 20 years. 2019, this device’s year, is highlighted.

FDA decisions per year · POK

5in 2025

2017: 12019: 12020: 22021: 22022: 12023: 12024: 62025: 52026 (partial year): 5
2007high 6 (2024) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · POK

119in 2025

2024: 161 days median2025: 119 days median2026 (partial year): 155 days median
2007high 161 (2024) · low 119 (2025)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · POK

No recall names POK since 2007.

Adverse event reports per year · POK

1in 2021

2021: 1
2007high 1 (2021) · low 0 (2007)2025

MAUDE reports naming product code POK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20100—00
20110—00
20120—00
20130—00
20140—00
20150—00
20160—00
20171—00
20180—00
20191—00
20202—00
20212—01
20221—00
20231—00
2024616100
2025511900
202651550—

What is it indicated for?

Koios Decision Support (DS) for Breast is a software application designed to assist trained interpreting physicians in analyzing the breast ultrasound images of patients with soft tissue breast lesions who are being referred for further diagnostic ultrasound examination. Koios DS for Breast is a machine learning-based decision support system, indicated as an adjunct to diagnostic ultrasound for breast cancer. Koios DS for Breast automatically classifies…

Read the full statement (298 words)

Koios Decision Support (DS) for Breast is a software application designed to assist trained interpreting physicians in analyzing the breast ultrasound images of patients with soft tissue breast lesions who are being referred for further diagnostic ultrasound examination. Koios DS for Breast is a machine learning-based decision support system, indicated as an adjunct to diagnostic ultrasound for breast cancer. Koios DS for Breast automatically classifies user-selected region(s) of interest (ROIs) containing a breast lesion into four BI-RADS-aligned categories (Benign, Probably Benign, Suspicious, Probably Malignant), and displays a continuous graphical confidence level indicator of where the lesion falls across all categories. Koios DS for Breast also automatically classifies lesion shape and orientation according to BI-RADS descriptors.

The software requires a user to select up to two ROIs, from up to two orthogonal views, that represent a single lesion to be selected and processed. When utilized by an interpreting physician who has completed the prescribed training, this device provides information that may be useful in rendering an accurate diagnosis. Patient management decisions should not be made solely on the results of the Koios DS for Breast analysis. This device is intended to help trained interpreting physicians improve their overall accuracy as well as reduce inter- and intra-operator variability.

Koios DS for Breast may also be used as an image viewer of multi-modality digital images, including ultrasound and mammography. The software includes tools that allow users to adjust, measure and document images, and output into a structured report.

Limitations: Koios DS for Breast is not to be used on sites of post-surgical excision, or images with doppler, elastography, or other overlays present in them. Koios DS for Breast is not intended for the primary interpretation of digital mammography images. Koios DS for Breast is not intended for use on mobile devices.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K190442: Koios DS for Breast.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K190442

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026