510(k) Report

510(k) · Cleared

Cleared

087 Balloon Guide Catheter System

K190749510(k) TraditionalCleared Jul 15, 2019112-day reviewClass II21 CFR 870.1250

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Q'Apel · DQY Catheter, Percutaneous

Cleared on Jul 15, 2019 as substantially equivalent to K153729 (Concentric Medical, 2016) after a 112-day review. Its family tree reaches back 7 generations to 2001.

Decided
Jul 15, 2019
Review
112 days
Class
Class II
Product code
DQY
Recalls
None
Predicates
1

Its longest line of descent runs back 7 generations to K010954 (2001). No later device cites it as a predicate yet.

  1. Root · 7 generations backK010954 Concentric Balloon Guide Catheter, Model 90031Concentric Medical, Inc. · 2001 · Recalled
  2. 6 generations backK021899 Concentric Balloon Guide Catheter, Model 90072Concentric Medical, Inc. · 2002 · Recalled
  3. 5 generations backK102657 Concentric Balloon Guide CatheterConcentric Medical, Inc. · 2010
  4. 4 generations backK112404 Concentric Balloon Guide CatheterConcentric Medical, Inc. · 2012 · Recalled
  5. 3 generations backK122581 Modified Balloon Guide Catheter Model 90253, 90254Concentric Medical, Inc. · 2012 · Recalled
  6. 2 generations backK131492 Modified Flowgate Balloon Guide CatheterConcentric Medical, Inc. · 2013
  7. 1 generation backK153729 8F FlowGate2 Balloon Guide CatheterConcentric Medical · 2016 · Recalled
  8. This deviceK190749 087 Balloon Guide Catheter SystemQ'Apel · 2019
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K010954Concentric Balloon Guide Catheter, Model 90031Concentric Medical, Inc.20011
−6K021899Concentric Balloon Guide Catheter, Model 90072Concentric Medical, Inc.20021
−5K102657Concentric Balloon Guide CatheterConcentric Medical, Inc.2010—
−4K112404Concentric Balloon Guide CatheterConcentric Medical, Inc.20121
−3K122581Modified Balloon Guide Catheter Model 90253, 90254Concentric Medical, Inc.20121
−2K131492Modified Flowgate Balloon Guide CatheterConcentric Medical, Inc.2013—
−1K1537298F FlowGate2 Balloon Guide CatheterConcentric Medical20162
0 (this device)K190749087 Balloon Guide Catheter SystemQ'Apel2019—
FIG. 1 — Lineage of K190749, 2001–2019Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DQY? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DQY device

  1. #1 K234115 APRO 55 Catheterscore 17.5 · 6 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K1537298F FlowGate2 Balloon Guide Catheter
Summary page 4
…Percutaneous Catheter Product Code(s): DQY, MJN Regulation Number(s): 870.1250 Predicate Devices: Concentric Medical, FlowGate 2™ (K153729) Indication for Use: The 087 Balloon Guide Catheter System is indicated for use in facilitating the insertion and guidance of an…
Concentric Medical20162

Also cleared against K153729: 6 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 18 years and passes through 3 companies; a new generation was cleared every 1.3 years on median.
  • 5 of the 7 ancestors have a recall of any class.
  • 6 other devices also cite its predicate K153729.

How does its category compare?

All DQY devices

DQY Catheter, Percutaneous: FDA decisions, review times, recalls and adverse event reports over 20 years. 2019, this device’s year, is highlighted.

FDA decisions per year · DQY

21in 2025

2007: 162008: 312009: 372010: 222011: 362012: 302013: 322014: 192015: 272016: 222017: 262018: 252019: 302020: 202021: 182022: 182023: 182024: 162025: 212026 (partial year): 11
2007high 37 (2009) · low 16 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DQY

115in 2025

2007: 85 days median2008: 66 days median2009: 73 days median2010: 74 days median2011: 71 days median2012: 87 days median2013: 92 days median2014: 127 days median2015: 118 days median2016: 117 days median2017: 113 days median2018: 106 days median2019: 111 days median2020: 120 days median2021: 81 days median2022: 74 days median2023: 114 days median2024: 87 days median2025: 115 days median2026 (partial year): 30 days median
2007high 127 (2014) · low 66 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DQY

1in 2025

2007: 62008: 122009: 172010: 322011: 62012: 22013: 36, 1 Class I2014: 32015: 32016: 22, 7 Class I2017: 82018: 22019: 12020: 22021: 22023: 172024: 12025: 1, 1 Class I2026 (partial year): 2, 1 Class I
2007high 36 (2013) · low 0 (2022)2026

Class I (10 since 2007) Class II/III not classified

Adverse event reports per year · DQY

3,056in 2025

2007: 3132008: 4962009: 8192010: 1,2932011: 1,2112012: 1,1912013: 1,3522014: 1,4972015: 2,3372016: 2,2522017: 2,4892018: 2,8552019: 2,9022020: 2,8892021: 3,0062022: 2,6432023: 3,0492024: 3,1902025: 3,056
2007high 3,190 (2024) · low 313 (2007)2025

MAUDE reports naming product code DQY, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200716856313
2008316612496
2009377317819
20102274321,293
2011367161,211
2012308721,191
2013329236 (1)1,352
20141912731,497
20152711832,337
20162211722 (7)2,252
20172611382,489
20182510622,855
20193011112,902
20202012022,889
2021188123,006
2022187402,643
202318114173,049
2024168713,190
2025211151 (1)3,056
202611302 (1)—

What is it indicated for?

The 087 Balloon Guide Catheter System is indicated for use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel in the peripheral and neuro vasculature. The balloon provides temporary vascular occlusion during such procedures.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K190749: 087 Balloon Guide Catheter System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K190749

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026