510(k) Report

510(k) · Cleared

Cleared

TGS Universal Headrest with Mounting Arm

K193502510(k) TraditionalCleared Feb 14, 202058-day reviewClass II21 CFR 882.4560

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Stryker Ent · PGW Ear, Nose, And Throat Stereotaxic Instrument

Cleared on Feb 14, 2020 as substantially equivalent to K133573 (Fiagon GmbH, 2014) and 1 other predicate after a 58-day review. Its family tree reaches back 5 generations to 1994. 3 of the 15 devices in its family, counting itself, have a Class I recall.

Decided
Feb 14, 2020
Review
58 days
Class
Class II
Product code
PGW
Recalls
None
Predicates
2

Its longest line of descent runs back 5 generations to K911783 (1994). 14 ancestors in all, the oldest K871046 (1987). No later device cites it as a predicate yet. 3 of the 15 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year cleared 2000: 1 other ancestor1998: 2 other ancestors, 2 with a Class I recall21996: 2 other ancestors, 1 with a Class I recall11995: 3 other ancestors1987: 1 other ancestorK911783 · Allegro Viewing Wand · I.S.G. Technologies, Inc. · 1994K960714 · Isg Viewing Wand · I.S.G. Technologies, Inc. · 1996K964229 · Regulus Navigator · Compass Intl., Inc. · 1997K133573 · Fiagon Navigation System · Fiagon GmbH · 2014K161491 · Scopis Hybrid Navigation System EM · Scopis GmbH · 20171990200020102020
Other ancestors, by generation 2 generations back: 3 other ancestors, 2 with a Class I recall23 generations back: 4 other ancestors, 1 with a Class I recall14 generations back: 2 other ancestors K161491 · Scopis Hybrid Navigation System EM · Scopis GmbH · 2017K133573 · Fiagon Navigation System · Fiagon GmbH · 2014K964229 · Regulus Navigator · Compass Intl., Inc. · 1997K960714 · Isg Viewing Wand · I.S.G. Technologies, Inc. · 1996K911783 · Allegro Viewing Wand · I.S.G. Technologies, Inc. · 1994−50
  1. Root · 5 generations backK911783 Allegro Viewing WandI.S.G. Technologies, Inc. · 1994
  2. 4 generations backK960714 Isg Viewing WandI.S.G. Technologies, Inc. · 1996
  3. 3 generations backK964229 Regulus NavigatorCompass Intl., Inc. · 1997
  4. 2 generations backK133573 Fiagon Navigation SystemFiagon GmbH · 2014
  5. 1 generation backK161491 Scopis Hybrid Navigation System EMScopis GmbH · 2017
  6. This deviceK193502 TGS Universal Headrest with Mounting ArmStryker Ent · 2020

Class I recalls in this family

View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K911783Allegro Viewing WandI.S.G. Technologies, Inc.1994—
−4K960714Isg Viewing WandI.S.G. Technologies, Inc.1996—
−3K964229Regulus NavigatorCompass Intl., Inc.1997—
−2K133573Fiagon Navigation SystemFiagon GmbH2014—
−1K161491Scopis Hybrid Navigation System EMScopis GmbH2017—
0 (this device)K193502TGS Universal Headrest with Mounting ArmStryker Ent2020—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198710
199530
199621
199822
200010
FIG. 1 — Lineage of K193502, 1994–2020Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new PGW device

  1. #1 K261540 Cube 4D Navigation System, VirtuEye Proscore 12.9 · 6 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K133573Fiagon Navigation System
Summary page 4
…Class II Regulation Number: 21 CFR 882.4560 Predicate Devices: Navigation Headrest Office, K133573 Field Generator Mounting Arm, K161491 Device Description: The TGS Universal Hea…
Fiagon GmbH2014—
*K161491Scopis Hybrid Navigation System EM
Summary page 4
….4560 Predicate Devices: Navigation Headrest Office, K133573 Field Generator Mounting Arm, K161491 Device Description: The TGS Universal Headrest with Mounting Arm is a reusable device intended to be used with the Stryker ENT N…
Scopis GmbH2017—

Also cleared against K133573: 7 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (3 Class I)

No FDA recall names this device.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K983529Instatrak 3000Visualization Technology, Inc. · 1998 · 2 generations back · ◆ in the 1998 ancestors bar of FIG. 11 Class I recall
  • K960330InstatrakVisualization Technology, Inc. · 1996 · 2 generations back · ◆ in the 1996 ancestors bar of FIG. 11 Class I recall
  • K981998Pediatric Instatrak SystemVisualization Technology, Inc. · 1998 · 3 generations back · ◆ in the 1998 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 26 years and passes through 5 companies; a new generation was cleared every 2.8 years on median.
  • 5 of the 14 ancestors have a recall of any class; 3 have a Class I recall.
  • 7 other devices also cite its predicate K133573.

How does its category compare?

All PGW devices

PGW Ear, Nose, And Throat Stereotaxic Instrument: FDA decisions, review times, recalls and adverse event reports over 20 years. 2020, this device’s year, is highlighted.

FDA decisions per year · PGW

1in 2025

2014: 42015: 12016: 92017: 42018: 22019: 12020: 82021: 12022: 22023: 22025: 12026 (partial year): 3
2007high 9 (2016) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · PGW

100in 2020

2014: 116 days median2016: 115 days median2017: 125 days median2020: 100 days median2026 (partial year): 65 days median
2007high 125 (2017) · low 100 (2020)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · PGW

1in 2025

2019: 32021: 12023: 12025: 1
2007high 3 (2019) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · PGW

220in 2025

2014: 72015: 142016: 332017: 422018: 962019: 2472020: 1612021: 1902022: 1662023: 2002024: 2742025: 220
2007high 274 (2024) · low 0 (2007)2025

MAUDE reports naming product code PGW, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20100—00
20110—00
20120—00
20130—00
2014411607
20151—014
20169115033
20174125042
20182—096
20191—3247
202081000161
20211—1190
20222—0166
20232—1200
20240—0274
20251—1220
20263650—

What is it indicated for?

The Stryker ENT Navigation System, of which the TGS Universal Headrest with Mounting Arm is an accessory, is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure in the field of ENT surgery, such as paranasal sinuses, mastoid anatomy, can be identified relative to a CT or MR based model of the anatomy. Example procedures include, but are not limited to the…

Read the full statement (109 words)

The Stryker ENT Navigation System, of which the TGS Universal Headrest with Mounting Arm is an accessory, is

indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a

rigid anatomical structure in the field of ENT surgery, such as paranasal sinuses, mastoid anatomy, can be identified

relative to a CT or MR based model of the anatomy.

Example procedures include, but are not limited to the following ENT procedures:

* Transsphenoidal access procedures; '

* Intranasal procedures;

+ Sinus procedures, such as maxillary antrostomies, ethmoidectomies; sphenoidotomies/sphenoid explorations,

turbinate resections, and frontal sinusotomies;

+ ENT related anterior skull base procedures.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K193502: TGS Universal Headrest with Mounting Arm.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K193502

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026