510(k) Report

510(k) · Cleared

Recalled · Class II · 2025

Ashvins

K200132510(k) TraditionalCleared Feb 18, 202028-day reviewClass II21 CFR 892.2050Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Medicalcommunications GmbH · LLZ System, Image Processing, Radiological

Cleared on Feb 18, 2020 as substantially equivalent to K142750 (Visus Technology Transfer GmbH, 2015) and 1 other predicate after a 28-day review. Its family tree reaches back 10 generations to 1997. It has been recalled once.

Decided
Feb 18, 2020
Review
28 days
Class
Class II
Product code
LLZ
Recalls
1
Predicates
2

Its longest line of descent runs back 10 generations to K971684 (1997). 44 ancestors in all, the oldest K870039 (1987). No later device cites it as a predicate yet.

  1. Root · 10 generations backK971684 Magnetom Project 047 System (Numaris 3.5 Software)Siemens Medical Solutions USA, Inc. · 1997 · Recalled
  2. 6 generations between
  3. 3 generations backK101666 Syngo. Plaza VA20ASiemens AG Medical Solutions · 2010 · Recalled
  4. 2 generations backK132532 Syngo.Plaza VB10ASiemens AG Healthcare · 2013 · Recalled
  5. 1 generation backK142750 JiveXVisus Technology Transfer GmbH · 2015
  6. This deviceK200132 AshvinsMedicalcommunications GmbH · 2020
Show all 10 generations
  1. Root · 10 generations backK971684 Magnetom Project 047 System (Numaris 3.5 Software)Siemens Medical Solutions USA, Inc. · 1997 · Recalled
  2. 9 generations backK032428 Magnetom Avanto MR SystemSiemens Medical Solutions USA, Inc. · 2003 · Recalled
  3. 8 generations backK041112 Magnetom Espree SystemSiemens Medical Solutions USA, Inc. · 2004 · Recalled
  4. 7 generations backK052423 Magnetom Systems with Syngo Expert - I OptionSiemens Medi Cal Solutions, Inc. · 2006
  5. 6 generations backK061964 Syngo Multimodality WorkplaceSiemens AG Medical Solutions · 2006
  6. 5 generations backK072728 Syngo Multimodality WorkplaceSiemens AG Medical Solutions · 2008
  7. 4 generations backK092519 Syngo.XSiemens AG Medical Solutions · 2009 · Recalled
  8. 3 generations backK101666 Syngo. Plaza VA20ASiemens AG Medical Solutions · 2010 · Recalled
  9. 2 generations backK132532 Syngo.Plaza VB10ASiemens AG Healthcare · 2013 · Recalled
  10. 1 generation backK142750 JiveXVisus Technology Transfer GmbH · 2015
  11. This deviceK200132 AshvinsMedicalcommunications GmbH · 2020
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−10K971684Magnetom Project 047 System (Numaris 3.5 Software)Siemens Medical Solutions USA, Inc.19974
−9K032428Magnetom Avanto MR SystemSiemens Medical Solutions USA, Inc.20034
−8K041112Magnetom Espree SystemSiemens Medical Solutions USA, Inc.20044
−7K052423Magnetom Systems with Syngo Expert - I OptionSiemens Medi Cal Solutions, Inc.2006—
−6K061964Syngo Multimodality WorkplaceSiemens AG Medical Solutions2006—
−5K072728Syngo Multimodality WorkplaceSiemens AG Medical Solutions2008—
−4K092519Syngo.XSiemens AG Medical Solutions20094
−3K101666Syngo. Plaza VA20ASiemens AG Medical Solutions20103
−2K132532Syngo.Plaza VB10ASiemens AG Healthcare201311
−1K142750JiveXVisus Technology Transfer GmbH2015—
0 (this device)K200132AshvinsMedicalcommunications GmbH2020—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198710
199010
199110
199510
199630
199720
199810
199910
200030
200130
200210
200530
200610
200740
200830
200930
201010
201210
FIG. 1 — Lineage of K200132, 1997–2020Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LLZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LLZ device

  1. #1 K230955 syngo Application Softwarescore 21.5 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K142750JiveX
Summary page 5
…Device Class: II Product Code: LLZ Product Code: NFJ 21 CFR 892.2050 21 CFR 892.2050 K142750 K122938 MedicalCommunications GmbH MCDOC-1419004256-26…
Visus Technology Transfer GmbH2015—
*K122938Forum Forum Archive Forum Archive & Viewer Forum Assist Match
Summary page 5
…Product Code: NFJ 21 CFR 892.2050 21 CFR 892.2050 K142750 K122938 MedicalCommunications GmbH MCDOC-1419004256-2672 Max-Jarecki-Str. 8, D-69115 Heidelberg Ver…
Carl Zeiss Meditec, AG20121

Also cleared against K142750: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 9, 2025Class IIZ-2673-2025Software designPotential that the measured value may be smaller than the actual area.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 23 years and passes through 6 companies; a new generation was cleared every 1.5 years on median.
  • 16 of the 44 ancestors have a recall of any class.
  • 1 other device also cites its predicate K142750.

How does its category compare?

All LLZ devices

LLZ System, Image Processing, Radiological: FDA decisions, review times, recalls and adverse event reports over 20 years. 2020, this device’s year, is highlighted.

FDA decisions per year · LLZ

58in 2025

2007: 762008: 942009: 702010: 512011: 1062012: 822013: 952014: 782015: 642016: 822017: 782018: 1042019: 742020: 812021: 832022: 652023: 872024: 622025: 582026 (partial year): 45
2007high 106 (2011) · low 51 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LLZ

173in 2025

2007: 36 days median2008: 52 days median2009: 62 days median2010: 98 days median2011: 79 days median2012: 87 days median2013: 84 days median2014: 89 days median2015: 90 days median2016: 90 days median2017: 107 days median2018: 105 days median2019: 107 days median2020: 115 days median2021: 116 days median2022: 137 days median2023: 157 days median2024: 131 days median2025: 173 days median2026 (partial year): 110 days median
2007high 173 (2025) · low 36 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LLZ

11in 2025

2007: 202008: 322009: 212010: 172011: 302012: 122013: 642014: 62015: 162016: 122017: 52018: 72019: 22020: 62021: 32022: 42023: 82024: 62025: 112026 (partial year): 14
2007high 64 (2013) · low 2 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LLZ

116in 2025

2007: 1522008: 1542009: 1452010: 1452011: 4542012: 2662013: 1,1302014: 15,6312015: 2112016: 2902017: 1042018: 722019: 882020: 822021: 702022: 282023: 262024: 422025: 116
2007high 15,631 (2014) · low 26 (2023)2025

MAUDE reports naming product code LLZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007763620152
2008945232154
2009706221145
2010519817145
20111067930454
2012828712266
20139584641,130
20147889615,631
2015649016211
2016829012290
2017781075104
2018104105772
201974107288
202081115682
202183116370
202265137428
202387157826
202462131642
20255817311116
20264511014—

What is it indicated for?

Ashvins is a modular software solution and is used for viewing, reporting, processing, storing, printing, and archiving, as well as for the exchange and distribution or web-based distribution of digital, multimodal medical images, including mammographic images and bio-signals, as well as findings and demographic information. Ashvins also collects, manages, and distributes patient-related and device-related information. Ashvins offers advanced image…

Read the full statement (137 words)

Ashvins is a modular software solution and is used for viewing, reporting, processing, storing, printing, and archiving, as well as for the exchange and distribution or web-based distribution of digital, multimodal medical images, including mammographic images and bio-signals, as well as findings and demographic information. Ashvins also collects, manages, and distributes patient-related and device-related information.

Ashvins offers advanced image processing functions such as multiplanar reconstruction, 3D reconstruction and measurement functions.

For primary image diagnosis in Mammography only uncompressed or non-lossy compressed images and only preprocessed DICOM “For Presentation” images must be used. Existing regulatory or legal requirements regarding hardware, such as diagnostic monitors, that are applicable for diagnosing in respective imaging specialties must be observed.

Ashvins is designed to be used by trained professionals, including but not limited to physicians, radiologists, ophthalmologists, nurses, medical technicians, and assistants.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K200132: Ashvins.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K200132

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026