510(k) Report

510(k) · Cleared

Cleared

3M Attest Super Rapid Steam Biological Indicator, 3M Attest Super Rapid Steam Challenge Pack, 3M Attest Auto-reader, 3M Attest Mini Auto-reader

K213809510(k) TraditionalCleared Apr 6, 2022121-day reviewClass II21 CFR 880.2800

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

3M Company · FRC Indicator, Biological Sterilization Process

Cleared on Apr 6, 2022 as substantially equivalent to K192550 (3M Company, 2020) after a 121-day review. Its family tree reaches back 6 generations to 1995. 1 later device descends from it.

Decided
Apr 6, 2022
Review
121 days
Class
Class II
Product code
FRC
Recalls
None
Predicates
1

Its longest line of descent runs back 6 generations to K926364 (1995). 11 ancestors in all, the oldest K900771 (1990). 1 later device descends from it, 1 citing it directly.

  1. Root · 6 generations backK926364 Attest Brand Rapid Readout Biological Ind. #129X3M Health Care, Ltd. · 1995
  2. 5 generations backK090569 Modification to: Attest Rapid Readout Biological Indicator, Mode 12923M Company · 2009
  3. 4 generations backK121484 3M Attest (TM) Super Rapid Readout Biological Indicator, 3M Attest…3M Company · 2012
  4. 3 generations backK121593 3M Attest (TM) Super Radid Readout Steam Challenge Pack 3M Attest…3M Company · 2013
  5. 2 generations backK173519 3M Attest Super Rapid Readout Steam Challenge Pack, 3M Attest Super…3M Company · 2018 · Recalled
  6. 1 generation backK192550 3MTM AttestTM Super Rapid Steam Biological Indicator 1592, 3MTM…3M Company · 2020
  7. This deviceK213809 3M Attest Super Rapid Steam Biological Indicator, 3M Attest Super…3M Company · 2022
  8. Descendant · cites it directly · 0 of its ownK242538 3M™ Attest™ Super Rapid Readout Biological Indicator 1493; 3M™…3M Company · 2024

Most influential descendants

Showing 1 of 1 · See all
  • K2425383M™ Attest™ Super Rapid Readout Biological Indicator 1493; 3M™ Attest™ Super Rapid Steam…3M Company · 2024 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K926364Attest Brand Rapid Readout Biological Ind. #129X3M Health Care, Ltd.1995—
−5K090569Modification to: Attest Rapid Readout Biological Indicator, Mode 12923M Company2009—
−4K1214843M Attest (TM) Super Rapid Readout Biological Indicator, 3M Attest (TM)…3M Company2012—
−3K1215933M Attest (TM) Super Radid Readout Steam Challenge Pack 3M Attest (TM) Super…3M Company2013—
−2K1735193M Attest Super Rapid Readout Steam Challenge Pack, 3M Attest Super Rapid 5…3M Company20181
−1K1925503MTM AttestTM Super Rapid Steam Biological Indicator 1592, 3MTM AttestTM Super…3M Company2020—
0 (this device)K2138093M Attest Super Rapid Steam Biological Indicator, 3M Attest Super Rapid Steam…3M Company2022—
+1K2425383M™ Attest™ Super Rapid Readout Biological Indicator 1493; 3M™ Attest™ Super…3M Company2024—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199010
199410
200210
201110
201810

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202410
FIG. 1 — Lineage of K213809, 1995–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FRC? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRC device

  1. #1 K250044 Celerity 5 HP Biological Indicator (LCB052)…score 14.5 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K1925503MTM AttestTM Super Rapid Steam Biological Indicator 1592, 3MTM AttestTM Super Rapid…
Summary page 6
…™ Attest™ Super Rapid Steam Biological Indicator 1592, 3M™ Attest™ Super Rapid Steam Challenge Pack 51582, 3M™ Attest™ Auto-reader 490, and 3M™ Attest™ Auto-reader 490H, K192550 Reference Device: 3M™ Attest™ Mini Auto-reader 490M, K200092 Indications for Use – 1592 Biological Indicator Use the 3M™ Attes…
3M Company2020—
†K2000923MTM AttestTM Mini Auto-reader 490M3M2020—

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K2425383M™ Attest™ Super Rapid Readout Biological Indicator 1493; 3M™ Attest™ Super Rapid Steam…3M Company20240

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 27 years and passes through 2 companies; a new generation was cleared every 3.5 years on median.
  • 3 of the 11 ancestors have a recall of any class.
  • Of its descendants, K242538 (3M Company, 2024) has gone furthest, with 0 descendants of its own.

How does its category compare?

All FRC devices

FRC Indicator, Biological Sterilization Process: FDA decisions, review times, recalls and adverse event reports over 20 years. 2022, this device’s year, is highlighted.

FDA decisions per year · FRC

6in 2025

2007: 32008: 12009: 62010: 42011: 62012: 92013: 42014: 132015: 32016: 22017: 102018: 182019: 52020: 102021: 52022: 52023: 62024: 82025: 62026 (partial year): 1
2007high 18 (2018) · low 1 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRC

29in 2025

2007: 83 days median2009: 183 days median2010: 165 days median2011: 118 days median2012: 178 days median2013: 260 days median2014: 174 days median2015: 130 days median2017: 130 days median2018: 97 days median2019: 88 days median2020: 150 days median2021: 107 days median2022: 193 days median2023: 190 days median2024: 105 days median2025: 29 days median
2007high 260 (2013) · low 29 (2025)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRC

8so far in 2026, the most since 2007

2009: 12012: 22013: 22022: 42023: 12026 (partial year): 8
2007high 8 (2026) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FRC

9in 2025

2007: 1002008: 602009: 452010: 572011: 372012: 462013: 392014: 512015: 1822016: 4822017: 1822018: 222019: 302020: 162021: 102022: 92023: 202024: 222025: 9
2007high 482 (2016) · low 9 (2022)2025

MAUDE reports naming product code FRC, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20073830100
20081—060
20096183145
20104165057
20116118037
20129178246
20134260239
201413174051
201531300182
20162—0482
2017101300182
20181897022
2019588030
202010150016
20215107010
2022519349
20236190120
20248105022
202562909
20261—8—

What is it indicated for?

3M™ Attest™ Super Rapid Steam Biological Indicator 1592: Use the 3M™ Attest™ Super Rapid Steam Biological Indicator 1592 in conjunction with the 3M™ Attest™ Auto-reader 490 having software version 4.0.0 or greater, the 3SM™ Attest™ Auto-reader 490H having software version 4.0.0 or greater, or the 3M™ Attest™ Mini Auto-reader 490M to qualify or monitor the following steam sterilization cycles: Cycle Type | Exposure Temperature | Exposure Time Gravity…

Read the full statement (408 words)

3M™ Attest™ Super Rapid Steam Biological Indicator 1592:

Use the 3M™ Attest™ Super Rapid Steam Biological Indicator 1592 in conjunction with the 3M™ Attest™ Auto-reader 490 having software version 4.0.0 or greater, the 3SM™ Attest™ Auto-reader 490H having software version 4.0.0 or greater, or the 3M™ Attest™ Mini Auto-reader 490M to qualify or monitor the following steam sterilization cycles:

Cycle Type | Exposure Temperature | Exposure Time

Gravity Displacement | 250°F (121°C) | 30 minutes

Dynamic-air-removal (pre-vacuum and SFPP) | 250°F (121°C) | 15, 20, 30, or 35 minutes

Dynamic-air-removal (pre-vacuum and SFPP) | 270°F (132°C) | 3,3.5,4,5.5, or 6 minutes

Dynamic-air-removal (pre-vacuum and SFPP) | 273°F (134°C) | 3 or 4 minutes

Dynamic-air-removal (pre-vacuum and SFPP) | 275°F (135°C) | 3, 3.5, or 10 minutes

3M™ Attest™ Super Rapid Steam Challenge Pack 51582:

Use the 3M™ Attest™ Super Rapid Steam Challenge Pack 51582 in conjunction with the 3M™ Attest™ Auto-reader 490 having software version 4.0.0 or greater, the 3M™ Attest™ Auto-reader 490H having software version 4.0.0 or greater, or the 3M™ Attest™ Mini Auto-reader 490M to qualify or monitor the following steam sterilization cycles:

Cycle Type | Exposure Temperature | Exposure Time

Gravity Displacement | 250°F (121°C) | 30 minutes

Dynamic-air-removal (pre-vacuum and SFPP) | 250°F (121°C) | 30 minutes

Dynamic-air-removal (pre-vacuum and SFPP) | 270°F (132°C) | 4 minutes

Dynamic-air-removal (pre-vacuum and SFPP) | 273°F (134°C) | 4 minutes

Dynamic-air-removal (pre-vacuum and SFPP) | 275°F (135°C) | 3 minutes

3M™ Attest™ Auto-reader 490:

The 3M™ Attest™ Auto reader 490 is designed to incubate and automatically read 3M™ Attest™ Rapid Readout Biological Indicators 1295 and 3M™ Attest™ Super Rapid Readout Biological Indicators, catalog numbers 1491, 1492V, and 1592 at 60°C for a final fluorescent result at 24 minutes.

3M™ Attest™ Auto-reader 490H:

The 3M™ Attest™ Auto-reader 490H is designed to incubate and automatically read 3M™ Attest™ Rapid Readout Biological Indicators 1295, 3M™ Attest™ Super Rapid Readout Biological Indicators, catalog numbers 1491 and 1492V, and 3M™ Attest Super Rapid Steam Biological Indicators 1592 at 60°C for a final fluorescent result at 24 minutes.

3M™ Attest™ Mini Auto-reader 490M:

The 3M™ Attest™ Mini Auto-reader 490M is designed to incubate and automatically read 3M™ Attest™ Rapid Readout Biological Indicators 1295 and 3M™ Attest™ Super Rapid Readout Biological Indicators, catalog numbers 1491, 1492V and 1592, at 60°C for a final fluorescent result at 24 minutes.

The indications for use for Attest™ Biological Indicators are provided in the product’s Instructions for Use.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K213809: 3M Attest Super Rapid Steam Biological Indicator, 3M Attest Super Rapid Steam Challenge Pack, 3M Attest Auto-reader, 3M Attest Mini Auto-reader.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K213809

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026