510(k) Report

510(k) · Cleared

Cleared

CareGUARD Sterilization Wrap

K213907510(k) TraditionalCleared Mar 14, 202290-day reviewClass II21 CFR 880.6850

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Care Essentials Pty, Ltd. · FRG Wrap, Sterilization

Cleared on Mar 14, 2022 as substantially equivalent to K082554 (Kimberly-Clark Corp., 2009) after a 90-day review. Its family tree reaches back 2 generations to 1988.

Decided
Mar 14, 2022
Review
90 days
Class
Class II
Product code
FRG
Recalls
None
Predicates
1

Its longest line of descent runs back 2 generations to K881471 (1988). No later device cites it as a predicate yet.

  1. Root · 2 generations backK881471 Kimguard (Regular & Heavy Duty) Sterile WrapKimberly-Clark Corp. · 1988 · Recalled
  2. 1 generation backK082554 Kimguard Sterilization Wrap, Models KC100, KC200, KC300, KC400…Kimberly-Clark Corp. · 2009
  3. This deviceK213907 CareGUARD Sterilization WrapCare Essentials Pty, Ltd. · 2022
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K881471Kimguard (Regular & Heavy Duty) Sterile WrapKimberly-Clark Corp.19881
−1K082554Kimguard Sterilization Wrap, Models KC100, KC200, KC300, KC400, KC500 and KC600Kimberly-Clark Corp.2009—
0 (this device)K213907CareGUARD Sterilization WrapCare Essentials Pty, Ltd.2022—
FIG. 1 — Lineage of K213907, 1988–2022Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FRG? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRG device

  1. #1 K234050 HALYARD® ONE-STEP® Sterilization Wrap…score 15.8 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K082554Kimguard Sterilization Wrap, Models KC100, KC200, KC300, KC400, KC500 and KC600
Summary page 4
…Regulation #: 21 CFR 880.6850 Predicate Device: Kimberly-Clark* Corporation KIMGUARD* Sterilization Wrap – K082554 Page 1 of 9 Device Descri…
Kimberly-Clark Corp.2009—

Also cleared against K082554: 3 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 34 years and passes through 2 companies; a new generation was cleared every 20.7 years on median.
  • 1 of the 2 ancestors has a recall of any class.
  • 3 other devices also cite its predicate K082554.

How does its category compare?

All FRG devices

FRG Wrap, Sterilization: FDA decisions, review times, recalls and adverse event reports over 20 years. 2022, this device’s year, is highlighted.

FDA decisions per year · FRG

9in 2025

2007: 32008: 22009: 42010: 42011: 62012: 102013: 62014: 82015: 122016: 92017: 92018: 52019: 92020: 52021: 42022: 92023: 92024: 112025: 92026 (partial year): 4
2007high 12 (2015) · low 2 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRG

151in 2025

2007: 182 days median2009: 171 days median2010: 213 days median2011: 110 days median2012: 133 days median2013: 146 days median2014: 177 days median2015: 196 days median2016: 131 days median2017: 260 days median2018: 150 days median2019: 258 days median2020: 122 days median2021: 170 days median2022: 217 days median2023: 230 days median2024: 182 days median2025: 151 days median2026 (partial year): 247 days median
2007high 260 (2017) · low 110 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRG

2in 2022

2013: 22017: 292020: 142022: 2
2007high 29 (2017) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FRG

18in 2025

2007: 12008: 22009: 22011: 122012: 42013: 52014: 132015: 92016: 202017: 72018: 222019: 102020: 212021: 92022: 42023: 132024: 212025: 18
2007high 22 (2018) · low 0 (2010)2025

MAUDE reports naming product code FRG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007318201
20082—02
2009417102
2010421300
20116110012
20121013304
2013614625
20148177013
20151219609
20169131020
20179260297
20185150022
20199258010
202051221421
2021417009
2022921724
20239230013
202411182021
20259151018
202642470—

What is it indicated for?

CareGUARD* Sterilization Wrap is a single-use, non-woven sterilization wrap intended to be sterilized by a health care provider using pre-vacuum steam at 273°F / 134°C for 4 minutes. The CareGUARD* Sterilization Wrap was validated for a dry time of 30 minutes. The CareGUARD* Sterilization wrap is recommended to be used with trays weighing up to 13.7 lbs. The wrap is intended to allow sterilization of the enclosed medical device(s) and maintain sterility…

Read the full statement (184 words)

CareGUARD* Sterilization Wrap is a single-use, non-woven sterilization wrap intended to be sterilized by a health care provider using pre-vacuum steam at 273°F / 134°C for 4 minutes. The CareGUARD* Sterilization Wrap was validated for a dry time of 30 minutes. The CareGUARD* Sterilization wrap is recommended to be used with trays weighing up to 13.7 lbs.

The wrap is intended to allow sterilization of the enclosed medical device(s) and maintain sterility of the enclosed device(s) until used.

Test results validated that CareGUARD™* Sterilization wraps allowed sterilization of the enclosed devices by pre-vacuum cycles.

Refer to Table 1 for recommendations for use per Pre-vacuum steam Sterility Maintenance Study.

Table 1: Basis weight recommendations for pre-vacuum steam

Basis Weight Maximum Wrapped Package Content Weight

50 GSM

60 GSM 6.2 kg/ 13.7 lbs

75 GSM

The following load was used in the Pre-vacuum steam Sterility Maintenance Study:

* 50GSM: 21 in. x 10 in. x 4 % in. (54 cm x 25 cm x 12 cm) tray containing 13.7 lbs (6.2 kg) of metal mass

The metal mass included surgical instruments used for orthopedic joint replacement

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K213907: CareGUARD Sterilization Wrap.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K213907

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026