510(k) · Cleared
ClearedKimguard Sterilization Wrap, Models KC100, KC200, KC300, KC400, KC500 and KC600
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Kimberly-Clark Corp. · FRG Wrap, Sterilization
Cleared on May 1, 2009 as substantially equivalent to K881471 (Kimberly-Clark Corp., 1988) after a 240-day review. Its family tree reaches back 1 generation to 1988. 10 later devices descend from it.
- Decided
- May 1, 2009
- Review
- 240 days
- Class
- Class II
- Product code
- FRG
- Recalls
- None
- Predicates
- 1
Its longest line of descent runs back 1 generation to K881471 (1988). 10 later devices descend from it, 4 citing it directly.
- Root · 1 generation backK881471 Kimguard (Regular & Heavy Duty) Sterile WrapKimberly-Clark Corp. · 1988 · Recalled
- This deviceK082554 Kimguard Sterilization Wrap, Models KC100, KC200, KC300, KC400…Kimberly-Clark Corp. · 2009
- Descendant · cites it directly · 6 of its ownK160755 Reliance Solo Sterilization Wrap, Reliance Tandem Sterilization WrapAhlstrom Nonwovens, LLC · 2016
- Descendant · 2 generations on · 2 of its ownK171533 Reliance Solo Sterilization Wrap, Reliance Tandem Sterilization WrapAhlstrom Nonwovens, LLC · 2017
- Descendant · 2 generations on · 1 of its ownK182656 JAMBRO Single Core A Sterilization wrapJiangsu Zhande Medical Supplies Co., Ltd. · 2019
- Descendant · cites it directly · 1 of its ownK143053 KIMGUARD ONE-STEP Sterilization WrapKimberly-Clark Health Care · 2015
- Descendant · 3 generations on · 0 of its ownK240438 CDB Hospital ® Sterilization Wrap (70 gsm); CDB Hospital ®…Comercializadora Brever · 2024
Showing 5 of 10 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 10 · See all- K160755Reliance Solo Sterilization Wrap, Reliance Tandem Sterilization WrapAhlstrom Nonwovens, LLC · 2016 · cites it directly6descendants
- K171533Reliance Solo Sterilization Wrap, Reliance Tandem Sterilization WrapAhlstrom Nonwovens, LLC · 2017 · 2 generations on2descendants
- K182656JAMBRO Single Core A Sterilization wrapJiangsu Zhande Medical Supplies Co., Ltd. · 2019 · 2 generations on1descendant
- K143053KIMGUARD ONE-STEP Sterilization WrapKimberly-Clark Health Care · 2015 · cites it directly1descendant
- K240438CDB Hospital ® Sterilization Wrap (70 gsm); CDB Hospital ® Sterilization Wrap (44 gsm)Comercializadora Brever · 2024 · 3 generations on0descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −1 | K881471 | Kimguard (Regular & Heavy Duty) Sterile Wrap | Kimberly-Clark Corp. | 1988 | 1 |
| 0 (this device) | K082554 | Kimguard Sterilization Wrap, Models KC100, KC200, KC300, KC400, KC500 and KC600 | Kimberly-Clark Corp. | 2009 | — |
| +1 | K160755 | Reliance Solo Sterilization Wrap, Reliance Tandem Sterilization Wrap | Ahlstrom Nonwovens, LLC | 2016 | — |
| +2 | K171533 | Reliance Solo Sterilization Wrap, Reliance Tandem Sterilization Wrap | Ahlstrom Nonwovens, LLC | 2017 | — |
| +2 | K182656 | JAMBRO Single Core A Sterilization wrap | Jiangsu Zhande Medical Supplies Co., Ltd. | 2019 | — |
| +1 | K143053 | KIMGUARD ONE-STEP Sterilization Wrap | Kimberly-Clark Health Care | 2015 | — |
| +3 | K240438 | CDB Hospital ® Sterilization Wrap (70 gsm); CDB Hospital ® Sterilization Wrap… | Comercializadora Brever | 2024 | — |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2015 | 1 | 0 |
| 2016 | 1 | 0 |
| 2017 | 1 | 0 |
| 2019 | 1 | 0 |
| 2022 | 2 | 0 |
| 2023 | 2 | 0 |
| 2024 | 2 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in FRG? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new FRG device
- #1 K234050 HALYARD® ONE-STEP® Sterilization Wrap…score 15.8 · 7 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K881471 | Kimguard (Regular & Heavy Duty) Sterile WrapSummary page 2…predicate KIMGUARD* Regular and Heavy Duty Sterile Wraps Page 1 of 3 - 510(k) Summary (K881471) Substantial The KIMGUARD* Sterilization Wrap is substantially equivalent to the Equivalence: predicates in intended use, d… | 1988 |
What descends from it? (10)
4 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
No FDA recall names this device.
No ancestor or descendant has a Class I recall.
What does this lineage say?
How does its category compare?
All FRG devicesFRG Wrap, Sterilization: FDA decisions, review times, recalls and adverse event reports over 20 years. 2009, this device’s year, is highlighted.
FDA decisions per year · FRG
9in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · FRG
151in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · FRG
2in 2022
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · FRG
18in 2025
MAUDE reports naming product code FRG, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 3 | 182 | 0 | 1 |
| 2008 | 2 | — | 0 | 2 |
| 2009 | 4 | 171 | 0 | 2 |
| 2010 | 4 | 213 | 0 | 0 |
| 2011 | 6 | 110 | 0 | 12 |
| 2012 | 10 | 133 | 0 | 4 |
| 2013 | 6 | 146 | 2 | 5 |
| 2014 | 8 | 177 | 0 | 13 |
| 2015 | 12 | 196 | 0 | 9 |
| 2016 | 9 | 131 | 0 | 20 |
| 2017 | 9 | 260 | 29 | 7 |
| 2018 | 5 | 150 | 0 | 22 |
| 2019 | 9 | 258 | 0 | 10 |
| 2020 | 5 | 122 | 14 | 21 |
| 2021 | 4 | 170 | 0 | 9 |
| 2022 | 9 | 217 | 2 | 4 |
| 2023 | 9 | 230 | 0 | 13 |
| 2024 | 11 | 182 | 0 | 21 |
| 2025 | 9 | 151 | 0 | 18 |
| 2026 | 4 | 247 | 0 | — |
What is it indicated for?
KIMGUARD* Sterilization Wrap is intended to be used to enclose another medical device that is to be sterilized by a health care provider by pre-vacuum steam at 270-F/1 32-C for 4 minutes or by 100% ethylene oxide (EtO) with a concentration of 725-735 mg/L at 131°F/ 55-C and 40% - 80% relative humidity for 60 minutes. The wrap is intended to allow sterilization of the enclosed medical device(s) and also to maintain sterility of the enclosed device(s)…
Read the full statement (153 words)
KIMGUARD* Sterilization Wrap is intended to be used to enclose another medical device that is to be sterilized by a health care provider by pre-vacuum steam at 270-F/1 32-C for 4 minutes or by 100% ethylene oxide (EtO) with a concentration of 725-735 mg/L at 131°F/ 55-C and 40% - 80% relative humidity for 60 minutes. The wrap is intended to allow sterilization of the enclosed medical device(s) and also to maintain sterility of the enclosed device(s) until opened. The wrap was validated for aeration times for EtO sterilization of 8 hours at 55 QC or 12 hours at 43.3 QC. The wrap was validated for dry times for pre-vacuum steam sterilization of 20 minutes for Models 100, 200, and 300 and for dry times of 30 minutes for Models 400, 500., and 600.
KIMGUARD* Sterilization Wrap is not indicated for use for gravity steam sterilization.
See Wrap Model Recommendations on Page 2.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K082554 · printed Nov 6, 2026
Cite this page
510(k) Report. “K082554: Kimguard Sterilization Wrap, Models KC100, KC200, KC300, KC400, KC500 and KC600.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K082554
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026