510(k) · Cleared
ClearedEnbio S
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Enbio Group AG · FLE Sterilizer, Steam
Cleared on Feb 22, 2022 as substantially equivalent to K210279 (Enbio Group AG, 2021) after a 63-day review. Its family tree reaches back 2 generations to 2019. 1 later device descends from it.
- Decided
- Feb 22, 2022
- Review
- 63 days
- Class
- Class II
- Product code
- FLE
- Recalls
- None
- Predicates
- 1
Its longest line of descent runs back 2 generations to K190062 (2019). 1 later device descends from it, 1 citing it directly.
Most influential descendants
Showing 1 of 1 · See all- K260254Enbio PROEnbio Group AG · 2026 · cites it directly0descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −2 | K190062 | STATCLAVE G4 Chamber Autoclave | Scican, Ltd. | 2019 | 1 |
| −1 | K210279 | Enbio S | Enbio Group AG | 2021 | — |
| 0 (this device) | K213991 | Enbio S | Enbio Group AG | 2022 | — |
| +1 | K260254 | Enbio PRO | Enbio Group AG | 2026 | — |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2026 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in FLE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new FLE device
- #1 K250168 Life Ultra 25L Tabletop Mono Chamber Steam…score 12.7 · 8 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K210279 | Enbio SSummary page 4…uct Code FLE Regulation Number 21 CFR 880.6880 Review Panel General Hospital 3. Predicate Device Enbio S (K210279) 4. Device Description The Enbio S is an air-removal (pre-vacuum) table-top steam sterilizer intended for use by a… | 2021 |
| † | K190062 | STATCLAVE G4 Chamber Autoclave | 2019 |
What descends from it? (1)
1 later device cites it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K260254 | Enbio PRO | 2026 | 0 |
Has anything in its family been recalled? (0 Class I)
No FDA recall names this device.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 3 years and passes through 2 companies; a new generation was cleared every 2.3 years on median.
- 1 of the 2 ancestors has a recall of any class.
- Of its descendants, K260254 (Enbio Group AG, 2026) has gone furthest, with 0 descendants of its own.
- 1 other device also cites its predicate K210279.
How does its category compare?
All FLE devicesFLE Sterilizer, Steam: FDA decisions, review times, recalls and adverse event reports over 20 years. 2022, this device’s year, is highlighted.
FDA decisions per year · FLE
4in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · FLE
184in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · FLE
1in 2021
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · FLE
71in 2025
MAUDE reports naming product code FLE, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 5 | 67 | 1 | 19 |
| 2008 | 1 | — | 1 | 21 |
| 2009 | 8 | 109 | 0 | 19 |
| 2010 | 2 | — | 0 | 20 |
| 2011 | 5 | 141 | 1 | 63 |
| 2012 | 5 | 111 | 1 | 69 |
| 2013 | 5 | 241 | 2 | 112 |
| 2014 | 2 | — | 0 | 113 |
| 2015 | 3 | 302 | 1 | 68 |
| 2016 | 1 | — | 0 | 76 |
| 2017 | 3 | 233 | 0 | 63 |
| 2018 | 8 | 146 | 1 | 90 |
| 2019 | 5 | 192 | 2 | 91 |
| 2020 | 2 | — | 0 | 87 |
| 2021 | 5 | 147 | 1 | 81 |
| 2022 | 10 | 133 | 0 | 70 |
| 2023 | 7 | 88 | 0 | 66 |
| 2024 | 2 | — | 0 | 64 |
| 2025 | 4 | 184 | 0 | 71 |
| 2026 | 3 | 85 | 1 | — |
What is it indicated for?
The Enbio S is an air-removal (pre-vacuum) table-top steam sterilizer intended for use by a health care provider to sterilize medical products by means of pressurized steam. It is suitable for the sterilization of dental and medical instruments that are validated to be sterilized by steam. The Enbio S has not been designed to sterilize liquid loads, bio-medical waste or materials not compatible with steam sterilization. The processing of such loads may…
Read the full statement (222 words)
The Enbio S is an air-removal (pre-vacuum) table-top steam sterilizer intended for use by a health care provider to sterilize medical products by means of pressurized steam. It is suitable for the sterilization of dental and medical instruments that are validated to be sterilized by steam. The Enbio S has not been designed to sterilize liquid loads, bio-medical waste or materials not compatible with steam sterilization. The processing of such loads may result in incomplete sterilization and/or damage to the autoclave.
Please refer to the table below for program name, load description, sterilization temperature, exposure time, drying time and maximum load.
Program Name: 134°C — solid objects, small porous objects, simple objects recessed, narrow-clearance items, dental handpieces, and textiles; wrapped and unwrapped. Sterilization temperature: 134°C (273°F); sterilization time: 4 minutes; drying time: 3 minutes; maximum load: 0.5 Kg/1.1 lbs.
Program Name: 121°C — solid objects, small porous objects, simple objects recessed, narrow-clearance items, dental handpieces, textiles, and plastics; wrapped and unwrapped. Sterilization temperature: 121°C (250°F); sterilization time: 30 minutes; drying time: 5 minutes; maximum load: 0.5 Kg/1.1 lbs.
Program Name: 134°C FAST* — solid objects, non-porous objects, simple instruments (such as scissors, handles, pliers, chisels, probes, etc.), and dental handpieces; unwrapped. Sterilization temperature: 134°C (273°F); sterilization time: 4 minutes; drying time: N/A; maximum load: 0.5 Kg/1.1 lbs. *Immediate Use Steam Sterilization cycle
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K213991 · printed Nov 6, 2026
Cite this page
510(k) Report. “K213991: Enbio S.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K213991
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026