510(k) Report

510(k) · Cleared

Cleared

OK Plus Indicator

K222137510(k) AbbreviatedCleared Dec 5, 2022139-day reviewClass II21 CFR 880.2800

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Propper Manufacturing Co., Inc. · JOJ Indicator, Physical/Chemical Sterilization Process

Cleared on Dec 5, 2022 as substantially equivalent to K191021 (Terragene S.A., 2019) after a 139-day review. Its family tree reaches back 1 generation to 2019.

Decided
Dec 5, 2022
Review
139 days
Class
Class II
Product code
JOJ
Recalls
None
Predicates
1

Its longest line of descent runs back 1 generation to K191021 (2019). No later device cites it as a predicate yet.

  1. Root · 1 generation backK191021 Terragene Bionova, Terragene Bionova, Terragene Integron, Terragene…Terragene S.A. · 2019 · Recalled
  2. This deviceK222137 OK Plus IndicatorPropper Manufacturing Co., Inc. · 2022
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K191021Terragene Bionova, Terragene Bionova, Terragene Integron, Terragene Cintape…Terragene S.A.20191
0 (this device)K222137OK Plus IndicatorPropper Manufacturing Co., Inc.2022—
FIG. 1 — Lineage of K222137, 2019–2022Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JOJ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JOJ device

  1. #1 K261992 3M™ Attest™ Steam Chemical Integrator (1243A…score 13.3 · 5 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K191021Terragene Bionova, Terragene Bionova, Terragene Integron, Terragene Cintape, Terragene…
Summary page 5
…on to Legally Marketed Predicate Device Comparison of the subject device (OK Plus indicator, Propper Manufacturing Co., Inc) to Predicate device (Chemdye CD29 indicator, k191021, Terragene S.A.). Subject device Predicate device Comparison Product…
Terragene S.A.20191

Also cleared against K191021: 5 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has a recall.
  • 5 other devices also cite its predicate K191021.

How does its category compare?

All JOJ devices

JOJ Indicator, Physical/Chemical Sterilization Process: FDA decisions, review times, recalls and adverse event reports over 20 years. 2022, this device’s year, is highlighted.

FDA decisions per year · JOJ

6in 2025

2007: 82008: 92009: 72010: 62011: 132012: 72013: 62014: 62015: 82016: 42017: 72018: 52019: 122020: 72021: 42022: 82023: 42024: 122025: 62026 (partial year): 3
2007high 13 (2011) · low 4 (2016)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JOJ

90in 2025

2007: 140 days median2008: 57 days median2009: 188 days median2010: 115 days median2011: 151 days median2012: 224 days median2013: 212 days median2014: 198 days median2015: 188 days median2016: 150 days median2017: 172 days median2018: 150 days median2019: 38 days median2020: 108 days median2021: 87 days median2022: 90 days median2023: 53 days median2024: 71 days median2025: 90 days median2026 (partial year): 30 days median
2007high 224 (2012) · low 38 (2019)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JOJ

1in 2023

2008: 22009: 22011: 22021: 12023: 1
2007high 2 (2008) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JOJ

12in 2025

2007: 12008: 12009: 92010: 32011: 12012: 82013: 62014: 112015: 72016: 112017: 352018: 122019: 82020: 62021: 292022: 232023: 392024: 252025: 12
2007high 39 (2023) · low 1 (2007)2025

MAUDE reports naming product code JOJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007814001
200895721
2009718829
2010611503
20111315121
2012722408
2013621206
20146198011
2015818807
20164150011
20177172035
20185150012
2019123808
2020710806
2021487129
2022890023
2023453139
20241271025
2025690012
20263300—

What is it indicated for?

The OK Plus indicator is a chemical indicator designed for monitoring the efficacy of steam sterilization process. The indicator responds to all critical sterilization parameters. The OK Plus indicator changes color from off-white to black to indicate that the conditions of the cycle have been met. The indicator is intended to be placed in each pack, pouch, tray or other containment device to function as an independent monitor of critical parameters for…

Read the full statement (168 words)

The OK Plus indicator is a chemical indicator designed for monitoring the efficacy of steam sterilization process. The indicator responds to all critical sterilization parameters. The OK Plus indicator changes color from off-white to black to indicate that the conditions of the cycle have been met. The indicator is intended to be placed in each pack, pouch, tray or other containment device to function as an independent monitor of critical parameters for the following sterilization cycles:

*Gravity: 121 C/250 F - 30 minutes

*Vacuum assisted (prevacuum): 132 C/270 F - 4 minutes

*Vacuum assisted (prevacuum): 132 C/270 F - 10 minutes

*Vacuum assisted (prevacuum): 134 C/273 F - 3.5 minutes

*Vacuum assisted (prevacuum): 134 C/273 F - 4 minutes

*Vacuum assisted (prevacuum): 135 C/275 F - 3 minutes

The OK Plus indicator has the following Stated Values determined in Resistometer:

* 121 C/250 F - 15 minutes

* 132 C/270 F - 4 minutes

* 134 C/273 F - 3 minutes

* 135 C/275 F - 2.9 minutes

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K222137: OK Plus Indicator.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K222137

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026