510(k) Report

510(k) · Cleared

Recalled · Class II · 2015

Advanced Perfusion System 1 (100/120 Volt),Advanced Perfusion System 1 (220/240 Volt, Roller Pumps (6, 4 Inch

K022947510(k) TraditionalCleared Mar 20, 2003196-day reviewClass II21 CFR 870.4220Recalls 150

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Terumo Cardiovascular Systems Corp. · DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass

Cleared on Mar 20, 2003 as substantially equivalent to K871131 (3M Health Care, Sarns, 1987) and 4 other predicates after a 196-day review. Its family tree reaches back 2 generations to 1992. 3 of the 32 devices in its family, counting itself, have a Class I recall. 26 later devices descend from it. It has been recalled 150 times.

Decided
Mar 20, 2003
Review
196 days
Class
Class II
Product code
DTQ
Recalls
150
Predicates
6

Its longest line of descent runs back 2 generations to K915183 (1992). 5 ancestors in all, the oldest K871131 (1987). 26 later devices descend from it, 6 citing it directly. 3 of the 32 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 1996: 1 other ancestor1995: 1 other ancestor1987: 1 other ancestor2006: 1 descendant cleared2014: 3 descendants cleared, 1 with a Class I recall12008: 3 descendants cleared2007: 2 descendants cleared2009: 1 descendant cleared2015: 3 descendants cleared2023: 5 descendants cleared2013: 2 descendants cleared, 1 with a Class I recall12011: 2 descendants cleared, 1 with a Class I recall12010: 1 descendant cleared2017: 2 descendants cleared2016: 1 descendant clearedK915183 · Sarns 8000 Modular Perfusion System · 3M Health Care, Sarns · 1992 · recalledK953901 · Sarns 800 Roller Pump · 3M Health Care, Sarns · 19961990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 3 other ancestors 1 generation forward: 6 descendants2 generations forward: 6 descendants3 generations forward: 6 descendants, 2 with a Class I recall24 generations forward: 5 descendants, 1 with a Class I recall15 generations forward: 2 descendants6 generations forward: 1 descendant K953901 · Sarns 800 Roller Pump · 3M Health Care, Sarns · 1996K915183 · Sarns 8000 Modular Perfusion System · 3M Health Care, Sarns · 1992 · recalled0+5
  1. Root · 2 generations backK915183 Sarns 8000 Modular Perfusion System3M Health Care, Sarns · 1992 · Recalled
  2. 1 generation backK953901 Sarns 800 Roller Pump3M Health Care, Sarns · 1996
  3. This deviceK022947 Advanced Perfusion System 1 (100/120 Volt),Advanced Perfusion System…Terumo Cardiovascular Systems Corp. · 2003
  4. Descendant · cites it directly · 8 of its ownK052555 Medtronic Performer TM CPB, Model R51, PCPBRand S.R.L. · 2006 · Recalled
  5. Descendant · cites it directly · 6 of its ownK131041 Advanced Perfusion System 1Terumo Cardiovascular Systems Corp. · 2014
  6. Descendant · cites it directly · 6 of its ownK070286 Bio-Console, Model 560Medtronic · 2007
  7. Descendant · 2 generations on · 5 of its ownK151349 Advanced Perfusion System 1Terumo Cardiovascular Systems Corporation · 2015
  8. Descendant · 2 generations on · 5 of its ownK090006 Lifebridge B2T, Model 7001Lifebridge Medizintechnik AG · 2009

Showing 5 of 26 descendants. See all descendants

Descendants cleared per year

2006: 12007: 22008: 32009: 12010: 12011: 22013: 22014: 32015: 32016: 12017: 22023: 5
2006high 5 (2023) · low 0 (2012)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 26 · See all
  • K052555Medtronic Performer TM CPB, Model R51, PCPBRand S.R.L. · 2006 · cites it directly8descendants
  • K131041Advanced Perfusion System 1Terumo Cardiovascular Systems Corp. · 2014 · cites it directly6descendants
  • K070286Bio-Console, Model 560Medtronic · 2007 · cites it directly6descendants
  • K151349Advanced Perfusion System 1Terumo Cardiovascular Systems Corporation · 2015 · 2 generations on5descendants
  • K090006Lifebridge B2T, Model 7001Lifebridge Medizintechnik AG · 2009 · 2 generations on5descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K915183Sarns 8000 Modular Perfusion System3M Health Care, Sarns199245
−1K953901Sarns 800 Roller Pump3M Health Care, Sarns1996—
0 (this device)K022947Advanced Perfusion System 1 (100/120 Volt),Advanced Perfusion System 1…Terumo Cardiovascular Systems Corp.2003—
+1K052555Medtronic Performer TM CPB, Model R51, PCPBRand S.R.L.20061
+1K131041Advanced Perfusion System 1Terumo Cardiovascular Systems Corp.2014—
+1K070286Bio-Console, Model 560Medtronic2007—
+2K151349Advanced Perfusion System 1Terumo Cardiovascular Systems Corporation2015—
+2K090006Lifebridge B2T, Model 7001Lifebridge Medizintechnik AG2009—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198710
199510
199610

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200610
200720
200830
200910
201010
201121
201321
201431
201530
201610
201720
202350
FIG. 1 — Lineage of K022947, 1992–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DTQ device

  1. #1 K221373 Essenz HLMscore 8.9 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (6)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary; numbers the summary cites that cannot be its predicate are marked – and kept off the family tree.

TypeNumberDeviceApplicantDecidedRecalls
*K871131Sarns 9000 Heart-Lung Console
Summary page 1
…Device Classification Name: Cardiopulmonary bypass heart-lung machine (21 CFR 870.4220) Predicate Devices: SarnsTM9000Heart-Lung Console (K871131) SarnsTM8000 Modular Perfusion System (K915183) SarnsTM8000 Roller Pump (K953901) SarnsTM9000 Unive…
3M Health Care, Sarns198714
*K915183Sarns 8000 Modular Perfusion System
Summary page 1
…pass heart-lung machine (21 CFR 870.4220) Predicate Devices: SarnsTM9000Heart-Lung Console (K871131) SarnsTM8000 Modular Perfusion System (K915183) SarnsTM8000 Roller Pump (K953901) SarnsTM9000 Universal Roller Pump (K953904) Flow Sensor Centrifu…
3M Health Care, Sarns199245
*K953901Sarns 800 Roller Pump
Summary page 1
…Predicate Devices: SarnsTM9000Heart-Lung Console (K871131) SarnsTM8000 Modular Perfusion System (K915183) SarnsTM8000 Roller Pump (K953901) SarnsTM9000 Universal Roller Pump (K953904) Flow Sensor Centrifugal Control Module (K950739) Delph…
3M Health Care, Sarns1996—
*K953904Sarns 9000 Universal Roller Pump
Summary page 1
…-Lung Console (K871131) SarnsTM8000 Modular Perfusion System (K915183) SarnsTM8000 Roller Pump (K953901) SarnsTM9000 Universal Roller Pump (K953904) Flow Sensor Centrifugal Control Module (K950739) Delphin II/Stand-Alone Centrifugal Module (K935978)…
3M Health Care, Ltd.19963
*K9507393M Sarns Flow Sensor, Centrifugal Control Module 115, 220 V, 900 Perfusion System
Summary page 1
…sion System (K915183) SarnsTM8000 Roller Pump (K953901) SarnsTM9000 Universal Roller Pump (K953904) Flow Sensor Centrifugal Control Module (K950739) Delphin II/Stand-Alone Centrifugal Module (K935978) Indication for Use The TerumoB Advanced Perfu…
3M Health Care, Sarns1995—
Kept off the family treeK935978Delphin II/Stand-Alone Centrifugal Module
Not in FDA’s 510(k), De Novo or PMA lists as we hold them (a misreading, a typo, or a number FDA does not list), so it is kept off the family tree
Summary page 1
…01) SarnsTM9000 Universal Roller Pump (K953904) Flow Sensor Centrifugal Control Module (K950739) Delphin II/Stand-Alone Centrifugal Module (K935978) Indication for Use The TerumoB Advanced Perfusion System 1 is indicated for up to 6 hours of use in…

Also cleared against K871131: 3 other devices →

What descends from it? (26)

See all 26

6 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K052555Medtronic Performer TM CPB, Model R51, PCPBRand S.R.L.20068
K131041Advanced Perfusion System 1Terumo Cardiovascular Systems Corp.20146
K070286Bio-Console, Model 560Medtronic20076
K082344Stockert S5 System, Stockert Electrical Venous Occluder (Evo)Sorin Group Deutschland GmbH20082
K070213Performer CBP SystemRand S.R.L.20071
K080832Stockert S5 System, Stockert Gas BlenderSorin Group Deutschland GmbH20080

Has anything in its family been recalled? (3 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 1, 2015Class IIZ-1687-2015Device DesignTerumo CVS is implementing field correction activities to address the identified causes of the Terumo System 1 experiencing loss of system power. These activities will be implemented in a phased approach beginning late 2015: -…
May 1, 2015Class IIZ-1688-2015Device DesignTerumo CVS is implementing field correction activities to address the identified causes of the Terumo System 1 experiencing loss of system power. These activities will be implemented in a phased approach beginning late 2015: -…
Jun 17, 2014Class IIZ-2019-2014Labeling Change ControlAn evaluation of the APS1 Operator Manual found that existing instructions lack clarity and accuracy related to specific items.
Jun 17, 2014Class IIZ-2018-2014Labeling Change ControlAn evaluation of the APS1 Operator Manual found that existing instructions lack clarity and accuracy related to specific items.
Jul 26, 2013Class IIZ-1974-2013Device DesignTerumo Cardiovascular Systems (TCVS) has become aware that an Advanced Perfusion System 1 (APS1) Centrifugal Pump 'Service Pump' message may occur due to user induced motor movement. In this instance, the APS1 will identify the movement…
Jul 26, 2013Class IIZ-1969-2013Software designThe operators manual does not fully describe the sequence of expected events that will occur when a specific operating condition exists. Specifically, if the user is adjusting the System 1 centrifugal pump speed via the Central Control…
Jul 26, 2013Class IIZ-1978-2013Software designTCVS is taking action to update the Operators Manual with the specific instructions to respond to this condition and will issue a labeling addendum for current users. In the event of a false high pressure alarm condition occurs that…
Jul 26, 2013Class IIZ-1973-2013Device DesignTerumo Cardiovascular Systems (TCVS) has become aware that an Advanced Perfusion System 1 (APS1) Centrifugal Pump 'Service Pump' message may occur due to user induced motor movement. In this instance, the APS1 will identify the movement…
Jul 26, 2013Class IIZ-1968-2013Software designThe operators manual does not fully describe the sequence of expected events that will occur when a specific operating condition exists. Specifically, if the user is adjusting the System 1 centrifugal pump speed via the Central Control…
Jul 26, 2013Class IIZ-1977-2013Software designTCVS is taking action to update the Operators Manual with the specific instructions to respond to this condition and will issue a labeling addendum for current users. In the event of a false high pressure alarm condition occurs that…
Nov 14, 2012Class IIZ-0435-2013Device DesignTerumo Cardiovascular System (TCVS) has received reports of a situation where users experienced a total loss of functionality for some System 1 units. The reports indicate that the units went blank and shut down with no sign of power and…
Nov 14, 2012Class IIZ-0436-2013Device DesignTerumo Cardiovascular System (TCVS) has received reports of a situation where users experienced a total loss of functionality for some System 1 units. The reports indicate that the units went blank and shut down with no sign of power and…
Sep 20, 2012Class IIZ-2449-2012Process controlTerumo Cardiovascular Systems (CVS) received one report of a flow module for the Terumo Advanced Perfusion System 1 not displaying blood flow information on the Central Control Monitor or the Centrifugal Control Unit. The tricolor LED on…
Sep 10, 2012Class IIZ-2443-2012Release of Material/Component prior to receiving test resultsTerumo Cardiovascular Systems (CVS) received two reports of the false back flow alarms for the Terumo¿ Advanced Perfusion System 1 when there was no tubing in the flow sensor. Terumo CVS' investigation found that the voltage level on a…
Jul 26, 2012Class IIZ-2177-2012Component design/selectionTerumo Cardiovascular Systems sent a Safety Advisory (AA-2010-007-C) on December 2, 2010 (Recall# Z-0882/0883-2011 RES 57416) regarding malfunctions of the Occluder and Air Bubble Detection (ABD) Modules to a confined number of Terumo…
Jul 26, 2012Class IIZ-2178-2012Component design/selectionTerumo Cardiovascular Systems sent a Safety Advisory (AA-2010-007-C) on December 2, 2010 (Recall# Z-0882/0883-2011 RES 57416) regarding malfunctions of the Occluder and Air Bubble Detection (ABD) Modules to a confined number of Terumo…
Jul 26, 2012Class IIZ-2233-2012Device DesignTerumo Cardiovascular Systems has received reports of malfunctions involving the loss of data transfer capabilities using the data transfer capabilities using modules on the Terumo Advanced Perfusion System 1. Some reports stated that the…
Jul 26, 2012Class IIZ-2232-2012Device DesignTerumo Cardiovascular Systems has received reports of malfunctions involving the loss of data transfer capabilities using the data transfer capabilities using modules on the Terumo Advanced Perfusion System 1. Some reports stated that the…
Jul 26, 2012Class IIZ-2231-2012Device DesignTerumo Cardiovascular Systems has received reports of malfunctions involving the loss of data transfer capabilities using the data transfer capabilities using modules on the Terumo Advanced Perfusion System 1. Some reports stated that the…
Jul 26, 2012Class IIZ-2230-2012Device DesignTerumo Cardiovascular Systems has received reports of malfunctions involving the loss of data transfer capabilities using the data transfer capabilities using modules on the Terumo Advanced Perfusion System 1. Some reports stated that the…
Jun 25, 2012Class IIZ-1969-2012Equipment maintenanceTerumo Cardiovascular Systems (Terumo CVS) has received reports of various roller pump occlusion malfunctions that could result in the inability to use the pump and would then require replacement with back-up equipment. Occlusion…
Jun 25, 2012Class IIZ-1971-2012Device DesignTerumo Cardiovascular Systems (Terumo CVS) has received reports of various roller pump malfunctions of the Terumo Advanced Perfusion System 1 that could result in the inability to use the pump and would then require replacement with…
Jun 25, 2012Class IIZ-1970-2012Device DesignTerumo Cardiovascular Systems (Terumo CVS) has received reports of various roller pump malfunctions of the Terumo Advanced Perfusion System 1 that could result in the inability to use the pump and would then require replacement with…
Jun 25, 2012Class IIZ-1968-2012Equipment maintenanceTerumo Cardiovascular Systems (Terumo CVS) has received reports of various roller pump occlusion malfunctions that could result in the inability to use the pump and would then require replacement with back-up equipment. Occlusion…
Jun 22, 2012Class IIZ-1962-2012Error in labelingTerumo Cardiovascular Systems (Terumo CVS) received two complaints related to the Gas System for the Terumo Advanced Perfusion System 1. Both were reported instances of users experiencing difficulty adjusting the System 1 Gas System local…
Jun 22, 2012Class IIZ-2055-2012Equipment maintenanceTerumo Cardiovascular Systems (Terumo CVS) received eight complaints since September 2008 of drive motor failure of the Sarns Centrifugal System of the Terumo Advanced Perfusion System 1. A failure of the drive motor for the Terumo System…
Jun 22, 2012Class IIZ-1958-2012Device DesignTerumo CVS has received reports of the Central Control Monitor (CCM) for Terumo Advanced Perfusion System 1 losing partial or full functionality. Symptoms that have been associated with a malfunctioning CCM include: full or partial loss…
Jun 22, 2012Class IIZ-2056-2012Equipment maintenanceTerumo Cardiovascular Systems (Terumo CVS) received eight complaints since September 2008 of drive motor failure of the Sarns Centrifugal System of the Terumo Advanced Perfusion System 1. A failure of the drive motor for the Terumo System…
Jun 22, 2012Class IIZ-1957-2012Device DesignTerumo CVS has received reports of the Central Control Monitor (CCM) for Terumo Advanced Perfusion System 1 losing partial or full functionality. Symptoms that have been associated with a malfunctioning CCM include: full or partial loss…
Jun 22, 2012Class IIZ-1966-2012Device DesignTerumo Advanced Perfusion System 1 battery charge is being depleted faster than expected during periods of system shutdown. The System 1 may falsely indicate a longer battery life than actual battery time remaining if not subjected to…
Jun 22, 2012Class IIZ-1963-2012Error in labelingTerumo Cardiovascular Systems (Terumo CVS) received two complaints related to the Gas System for the Terumo Advanced Perfusion System 1. Both were reported instances of users experiencing difficulty adjusting the System 1 Gas System local…
Jun 22, 2012Class IIZ-1965-2012Device DesignTerumo Advanced Perfusion System 1 battery charge is being depleted faster than expected during periods of system shutdown. The System 1 may falsely indicate a longer battery life than actual battery time remaining if not subjected to…
Jun 22, 2012Class IIZ-1964-2012Error in labelingTerumo Cardiovascular Systems (Terumo CVS) received two complaints related to the Gas System for the Terumo Advanced Perfusion System 1. Both were reported instances of users experiencing difficulty adjusting the System 1 Gas System local…
Jun 21, 2012Class IIZ-2021-2012Process controlTerumo¿ Advanced Perfusion System 1 has displayed a Red X error message on multiple module icons indicating a loss or fluctuation of internal power on one side of the system's base. Terumo Cardiovascular Systems' initial investigation has…
Jun 21, 2012Class IIZ-2020-2012Process controlTerumo¿ Advanced Perfusion System 1 has displayed a Red X error message on multiple module icons indicating a loss or fluctuation of internal power on one side of the system's base. Terumo Cardiovascular Systems' initial investigation has…
Jun 21, 2012Class IIZ-2010-2012Nonconforming Material/ComponentTerumo Cardiovascular Systems has received reports of several issues that manifest themselves as communications or power failures in the system network for Terumo Advanced Perfusion System 1. Issues that affect a single pump or module…
Jun 21, 2012Class IIZ-2009-2012Nonconforming Material/ComponentTerumo Cardiovascular Systems has received reports of several issues that manifest themselves as communications or power failures in the system network for Terumo Advanced Perfusion System 1. Issues that affect a single pump or module…
Jun 18, 2012Class IIZ-2042-2012Nonconforming Material/ComponentTerumo Cardiovascular Systems has received several reports of malfunctions of the Air Bubble Detection System used with its heart-lung machines over the last three years. The malfunctions were expressed in various failure modes…
Jun 18, 2012Class IIZ-2043-2012Nonconforming Material/ComponentTerumo Cardiovascular Systems has received several reports of malfunctions of the Air Bubble Detection System used with its heart-lung machines over the last three years. The malfunctions were expressed in various failure modes…
Mar 16, 2012Not classifiedZ-1320-2012Device DesignTerumo CVS has received reports of 27 instances in which the 6 inch Roller Pump jammed during cardioplegia delivery. There has been one report of the pump jamming during a case that concluded with an adverse patient outcome, although…
Mar 16, 2012Not classifiedZ-1321-2012Device DesignTerumo CVS has received reports of 27 instances in which the 6 inch Roller Pump jammed during cardioplegia delivery. There has been one report of the pump jamming during a case that concluded with an adverse patient outcome, although…
Jan 13, 2012Not classifiedZ-0839-2012Nonconforming Material/ComponentTerumo Cardiovascular Systems (Terumo CVS) discovered and rejected a single digital encoder cable assembly during incoming inspection because it had a loose wire. The assembly is used in the 4" and the 6" roller pumps for the Terumo…
Jan 13, 2012Not classifiedZ-0840-2012Nonconforming Material/ComponentTerumo Cardiovascular Systems (Terumo CVS) discovered and rejected a single digital encoder cable assembly during incoming inspection because it had a loose wire. The assembly is used in the 4" and the 6" roller pumps for the Terumo…
Nov 10, 2011Not classifiedZ-0301-2012Nonconforming Material/ComponentTerumo CVS discovered that our In house service center has incorrectly installed a felt washer into the main bearing of nine roller pumps that had been returned for repair or preventative maintenance unrelated to the felt washer. The felt…
Nov 10, 2011Not classifiedZ-0302-2012Nonconforming Material/ComponentTerumo CVS discovered that our In house service center has incorrectly installed a felt washer into the main bearing of nine roller pumps that had been returned for repair or preventative maintenance unrelated to the felt washer. The felt…
Nov 10, 2011Not classifiedZ-0300-2012Nonconforming Material/ComponentTerumo CVS discovered that our In house service center has incorrectly installed a felt washer into the main bearing of nine roller pumps that had been returned for repair or preventative maintenance unrelated to the felt washer. The felt…
Nov 8, 2011Not classifiedZ-0299-2012Component design/selection1 US customer was operating the Advanced Perfusion System 1 with an unapproved, modified cable, which was in product development, to connect the system an an occluder for the Sarns Modular Perfusion System 8000 . Five(5)of twelve(12)…
Nov 8, 2011Not classifiedZ-0298-2012Component design/selection1 US customer was operating the Advanced Perfusion System 1 with an unapproved, modified cable, which was in product development, to connect the system an an occluder for the Sarns Modular Perfusion System 8000 . Five(5)of twelve(12)…
May 3, 2011Not classifiedZ-2613-2011OtherThe firm has recieved complaints regarding 35 pressure monitoring kits in which the membrane in the kit separated form the pressure dome housing. The firm determined that the defect was caused by a change in the pressure dome's…
May 3, 2011Not classifiedZ-2620-2011OtherThe firm has recieved complaints regarding 35 pressure monitoring kits in which the membrane in the kit separated form the pressure dome housing. The firm determined that the defect was caused by a change in the pressure dome's…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K131964Bio-Console 560 Extracorporeal Blood Pumping ConsoleMedtronic, Inc. · 2013 · 3 generations on · ◆ in the 2013 descendants bar of FIG. 11 Class I recall
  • K102726Cardiohelp SystemMaquet Cardiopulmonary, AG · 2011 · 3 generations on · ◆ in the 2011 descendants bar of FIG. 11 Class I recall
  • K133598Cardiohelp SystemMaquet Cardiopulmonary, AG · 2014 · 4 generations on · ◆ in the 2014 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 11 years and passes through 2 companies; a new generation was cleared every 6.9 years on median.
  • 3 of the 5 ancestors have a recall of any class.
  • Of its descendants, K052555 (Rand S.R.L., 2006) has gone furthest, with 8 descendants of its own.
  • 3 other devices also cite its predicate K871131.

How does its category compare?

All DTQ devices

DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · DTQ

3in 2023

2007: 22008: 12009: 32010: 12011: 22014: 22015: 22017: 22018: 12019: 12021: 22023: 3
2007high 3 (2009) · low 0 (2012)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DTQ

292in 2023

2009: 86 days median2023: 292 days median
2007high 292 (2023) · low 86 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DTQ

1in 2024

2007: 92008: 172009: 12010: 172011: 222012: 422013: 42014: 12018: 12019: 1, 1 Class I2022: 12023: 7, 1 Class I2024: 1
2007high 42 (2012) · low 0 (2015)2026

Class I (2 since 2007) Class II/III not classified

Adverse event reports per year · DTQ

1,014in 2025

2007: 4202008: 5482009: 5892010: 8112011: 1,0812012: 1,6262013: 9062014: 1,2902015: 1,1842016: 7142017: 5272018: 6762019: 8572020: 6272021: 7292022: 7172023: 8182024: 8632025: 1,014
2007high 1,626 (2012) · low 420 (2007)2025

MAUDE reports naming product code DTQ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—9420
20081—17548
20093861589
20101—17811
20112—221,081
20120—421,626
20130—4906
20142—11,290
20152—01,184
20160—0714
20172—0527
20181—1676
20191—1 (1)857
20200—0627
20212—0729
20220—1717
202332927 (1)818
20240—1863
20250—01,014
20260—0—

What is it indicated for?

The TerumoB Advanced Perfusion System 1 is indicated for use for up to 6 hours in the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment.

The centrifugal pump is indicated for use in cardiopulmonary bypass procedures only.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K022947: Advanced Perfusion System 1 (100/120 Volt),Advanced Perfusion System 1 (220/240 Volt, Roller Pumps (6, 4 Inch.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K022947

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026