510(k) Report

510(k) · Cleared

Recalled · Class II · 2012

Sarns 9000 Heart-Lung Console

K871131510(k) TraditionalCleared May 21, 198762-day reviewClass II21 CFR 870.4220Recalls 14

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

3M Health Care, Sarns · DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass

Cleared on May 21, 1987 after a 62-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 3 of the 33 devices in its family, counting itself, have a Class I recall. 32 later devices descend from it. It has been recalled 14 times.

Decided
May 21, 1987
Review
62 days
Class
Class II
Product code
DTQ
Recalls
14
Predicates
0

32 later devices descend from it, 4 citing it directly. 3 of the 33 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 1996: 2 descendants cleared1998: 1 descendant cleared2003: 1 descendant cleared2006: 1 descendant cleared2014: 3 descendants cleared, 1 with a Class I recall12008: 3 descendants cleared2007: 2 descendants cleared2011: 3 descendants cleared, 1 with a Class I recall12009: 1 descendant cleared2015: 3 descendants cleared2023: 5 descendants cleared2013: 3 descendants cleared, 1 with a Class I recall12010: 1 descendant cleared2017: 2 descendants cleared2016: 1 descendant cleared1990200020102020
  1. This deviceK871131 Sarns 9000 Heart-Lung Console3M Health Care, Sarns · 1987
  2. Descendant · cites it directly · 29 of its ownK953904 Sarns 9000 Universal Roller Pump3M Health Care, Ltd. · 1996 · Recalled
  3. Descendant · cites it directly · 26 of its ownK022947 Advanced Perfusion System 1 (100/120 Volt),Advanced Perfusion System…Terumo Cardiovascular Systems Corp. · 2003 · Recalled
  4. Descendant · 2 generations on · 8 of its ownK112587 Advanced Perfusion System 1 (APS1) Tube Clamp Assembly, Roller Pump…Terumo Cardiovascular Systems Corp. · 2011
  5. Descendant · 2 generations on · 8 of its ownK052555 Medtronic Performer TM CPB, Model R51, PCPBRand S.R.L. · 2006 · Recalled
  6. Descendant · 2 generations on · 6 of its ownK131041 Advanced Perfusion System 1Terumo Cardiovascular Systems Corp. · 2014

Showing 5 of 32 descendants. See all descendants

Descendants cleared per year

1996: 21998: 12003: 12006: 12007: 22008: 32009: 12010: 12011: 32013: 32014: 32015: 32016: 12017: 22023: 5
1996high 5 (2023) · low 0 (1997)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 32 · See all
  • K953904Sarns 9000 Universal Roller Pump3M Health Care, Ltd. · 1996 · cites it directly29descendants
  • K022947Advanced Perfusion System 1 (100/120 Volt),Advanced Perfusion System 1 (220/240 Volt…Terumo Cardiovascular Systems Corp. · 2003 · cites it directly26descendants
  • K112587Advanced Perfusion System 1 (APS1) Tube Clamp Assembly, Roller Pump 6 Inch Classic Color,Terumo Cardiovascular Systems Corp. · 2011 · 2 generations on8descendants
  • K052555Medtronic Performer TM CPB, Model R51, PCPBRand S.R.L. · 2006 · 2 generations on8descendants
  • K131041Advanced Perfusion System 1Terumo Cardiovascular Systems Corp. · 2014 · 2 generations on6descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K871131Sarns 9000 Heart-Lung Console3M Health Care, Sarns1987—
+1K953904Sarns 9000 Universal Roller Pump3M Health Care, Ltd.19963
+1K022947Advanced Perfusion System 1 (100/120 Volt),Advanced Perfusion System 1…Terumo Cardiovascular Systems Corp.2003150
+2K112587Advanced Perfusion System 1 (APS1) Tube Clamp Assembly, Roller Pump 6 Inch…Terumo Cardiovascular Systems Corp.2011—
+2K052555Medtronic Performer TM CPB, Model R51, PCPBRand S.R.L.20061
+2K131041Advanced Perfusion System 1Terumo Cardiovascular Systems Corp.2014—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
199620
199810
200310
200610
200720
200830
200910
201010
201131
201331
201431
201530
201610
201720
202350
FIG. 1 — Lineage of K871131, 1987–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DTQ device

  1. #1 K221373 Essenz HLMscore 8.9 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it? (32)

See all 32

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K953904Sarns 9000 Universal Roller Pump3M Health Care, Ltd.199629
K022947Advanced Perfusion System 1 (100/120 Volt),Advanced Perfusion System 1 (220/240 Volt…Terumo Cardiovascular Systems Corp.200326
K973237CSS-Cardioplegia Safety SystemMedtronic Bio-Medicus, Inc.19980
K960916Sams 8000 Cardioplegia Monitor3M Health Care, Sarns19960

Has anything in its family been recalled? (3 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 18, 2012Class IIZ-2052-2012Nonconforming Material/ComponentTerumo Cardiovascular Systems has received several reports of malfunctions of the Air Bubble Detection System used with its heart-lung machines over the last three years. The malfunctions were expressed in various failure modes (over…
Jun 18, 2012Class IIZ-2051-2012Nonconforming Material/ComponentTerumo Cardiovascular Systems has received several reports of malfunctions of the Air Bubble Detection System used with its heart-lung machines over the last three years. The malfunctions were expressed in various failure modes (over…
Jun 18, 2012Class IIZ-2050-2012Nonconforming Material/ComponentTerumo Cardiovascular Systems has received several reports of malfunctions of the Air Bubble Detection System used with its heart-lung machines over the last three years. The malfunctions were expressed in various failure modes (over…
May 3, 2011Not classifiedZ-2614-2011OtherThe firm has recieved complaints regarding 35 pressure monitoring kits in which the membrane in the kit separated form the pressure dome housing. The firm determined that the defect was caused by a change in the pressure dome's…
May 3, 2011Not classifiedZ-2620-2011OtherThe firm has recieved complaints regarding 35 pressure monitoring kits in which the membrane in the kit separated form the pressure dome housing. The firm determined that the defect was caused by a change in the pressure dome's…
May 3, 2011Not classifiedZ-2613-2011OtherThe firm has recieved complaints regarding 35 pressure monitoring kits in which the membrane in the kit separated form the pressure dome housing. The firm determined that the defect was caused by a change in the pressure dome's…
May 3, 2011Not classifiedZ-2615-2011OtherThe firm has recieved complaints regarding 35 pressure monitoring kits in which the membrane in the kit separated form the pressure dome housing. The firm determined that the defect was caused by a change in the pressure dome's…
May 3, 2011Not classifiedZ-2621-2011OtherThe firm has recieved complaints regarding 35 pressure monitoring kits in which the membrane in the kit separated form the pressure dome housing. The firm determined that the defect was caused by a change in the pressure dome's…
May 3, 2011Not classifiedZ-2618-2011OtherThe firm has recieved complaints regarding 35 pressure monitoring kits in which the membrane in the kit separated form the pressure dome housing. The firm determined that the defect was caused by a change in the pressure dome's…
May 3, 2011Not classifiedZ-2616-2011OtherThe firm has recieved complaints regarding 35 pressure monitoring kits in which the membrane in the kit separated form the pressure dome housing. The firm determined that the defect was caused by a change in the pressure dome's…
May 3, 2011Not classifiedZ-2619-2011OtherThe firm has recieved complaints regarding 35 pressure monitoring kits in which the membrane in the kit separated form the pressure dome housing. The firm determined that the defect was caused by a change in the pressure dome's…
May 3, 2011Not classifiedZ-2617-2011OtherThe firm has recieved complaints regarding 35 pressure monitoring kits in which the membrane in the kit separated form the pressure dome housing. The firm determined that the defect was caused by a change in the pressure dome's…
May 23, 2008Not classifiedZ-0295-2009Nonconforming Material/ComponentThe level sensor may not properly couple to the reservoir, resulting in a sensor "not attached" message, an alert or alarm condition, or a failure to detect a low level condition.
May 23, 2008Not classifiedZ-0296-2009Nonconforming Material/ComponentThe level sensor may not properly couple to the reservoir, resulting in a sensor "not attached" message, an alert or alarm condition, or a failure to detect a low level condition.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K131964Bio-Console 560 Extracorporeal Blood Pumping ConsoleMedtronic, Inc. · 2013 · 4 generations on · ◆ in the 2013 descendants bar of FIG. 11 Class I recall
  • K102726Cardiohelp SystemMaquet Cardiopulmonary, AG · 2011 · 4 generations on · ◆ in the 2011 descendants bar of FIG. 11 Class I recall
  • K133598Cardiohelp SystemMaquet Cardiopulmonary, AG · 2014 · 5 generations on · ◆ in the 2014 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • Of its descendants, K953904 (3M Health Care, Ltd., 1996) has gone furthest, with 29 descendants of its own.

How does its category compare?

All DTQ devices

DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · DTQ

3in 2023

2007: 22008: 12009: 32010: 12011: 22014: 22015: 22017: 22018: 12019: 12021: 22023: 3
2007high 3 (2009) · low 0 (2012)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DTQ

292in 2023

2009: 86 days median2023: 292 days median
2007high 292 (2023) · low 86 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DTQ

1in 2024

2007: 92008: 172009: 12010: 172011: 222012: 422013: 42014: 12018: 12019: 1, 1 Class I2022: 12023: 7, 1 Class I2024: 1
2007high 42 (2012) · low 0 (2015)2026

Class I (2 since 2007) Class II/III not classified

Adverse event reports per year · DTQ

1,014in 2025

2007: 4202008: 5482009: 5892010: 8112011: 1,0812012: 1,6262013: 9062014: 1,2902015: 1,1842016: 7142017: 5272018: 6762019: 8572020: 6272021: 7292022: 7172023: 8182024: 8632025: 1,014
2007high 1,626 (2012) · low 420 (2007)2025

MAUDE reports naming product code DTQ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—9420
20081—17548
20093861589
20101—17811
20112—221,081
20120—421,626
20130—4906
20142—11,290
20152—01,184
20160—0714
20172—0527
20181—1676
20191—1 (1)857
20200—0627
20212—0729
20220—1717
202332927 (1)818
20240—1863
20250—01,014
20260—0—
Cite this page

510(k) Report. “K871131: Sarns 9000 Heart-Lung Console.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K871131

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026