Rotaflow Centrifugal Pump System

K991864510(k) TraditionalCleared Aug 31, 2000457-day reviewClass II21 CFR 870.4360Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Jostra Medizintechnik AG · KFM Pump, Blood, Cardiopulmonary Bypass, Non-Roller Type

Cleared on Aug 31, 2000 after a 457-day review. The FDA summary names its predicate without a number. 4 of the 31 devices in its family, counting itself, have a Class I recall. 30 later devices descend from it. It has been recalled 2 times.

Decided
Aug 31, 2000
Review
457 days
Class
Class II
Product code
KFM
Recalls
2
Predicates
0

30 later devices descend from it, 9 citing it directly. 4 of the 31 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2002: 2 descendants cleared2024: 3 descendants cleared2020: 3 descendants cleared, 1 with a Class I recall12006: 3 descendants cleared2011: 3 descendants cleared, 1 with a Class I recall12003: 3 descendants cleared2017: 1 descendant cleared2008: 2 descendants cleared2013: 1 descendant cleared2014: 2 descendants cleared, 2 with a Class I recall22025: 1 descendant cleared2023: 1 descendant cleared2022: 1 descendant cleared2009: 3 descendants cleared2026: 1 descendant cleared200020052010201520202025
  1. This deviceK991864 Rotaflow Centrifugal Pump SystemJostra Medizintechnik AG · 2000
  2. Descendant · cites it directly · 10 of its ownK061072 Rotaflow Centrifugal Pump with Safeline Coating, Model BSQ-RF-32Maquet Cardiopulmonary, AG · 2006
  3. Descendant · cites it directly · 4 of its ownK011838 Centrifugal PumpStockert Instrumente GmbH · 2002 · Recalled
  4. Descendant · cites it directly · 3 of its ownK191407 Novalung SystemFresenius Medical Care Renal Therapies Group, LLC · 2020 · Recalled
  5. Descendant · 2 generations on · 3 of its ownK061743 Venous Hardshell Cardiotomy Reservoir W/Safeline Coating, Model…Maquet Cardiopulmonary, AG · 2006
  6. Descendant · cites it directly · 2 of its ownK200109 OXY-1 SystemAbiomed, Inc. · 2020 · Recalled (Class I)

Showing 5 of 30 descendants. See all descendants

Descendants cleared per year

2002: 22003: 32006: 32008: 22009: 32011: 32013: 12014: 22017: 12020: 32022: 12023: 12024: 32025: 12026: 1
2002high 3 (2003) · low 0 (2004)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 30 · See all
  • K061072Rotaflow Centrifugal Pump with Safeline Coating, Model BSQ-RF-32Maquet Cardiopulmonary, AG · 2006 · cites it directly10descendants
  • K011838Centrifugal PumpStockert Instrumente GmbH · 2002 · cites it directly4descendants
  • K191407Novalung SystemFresenius Medical Care Renal Therapies Group, LLC · 2020 · cites it directly3descendants
  • K061743Venous Hardshell Cardiotomy Reservoir W/Safeline Coating, Model BSQ-VHK 2000;Vacuum…Maquet Cardiopulmonary, AG · 2006 · 2 generations on3descendants
  • K200109OXY-1 SystemAbiomed, Inc. · 2020 · cites it directly2descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K991864Rotaflow Centrifugal Pump SystemJostra Medizintechnik AG2000—
+1K061072Rotaflow Centrifugal Pump with Safeline Coating, Model BSQ-RF-32Maquet Cardiopulmonary, AG2006—
+1K011838Centrifugal PumpStockert Instrumente GmbH20024
+1K191407Novalung SystemFresenius Medical Care Renal Therapies Group, LLC20202
+2K061743Venous Hardshell Cardiotomy Reservoir W/Safeline Coating, Model BSQ-VHK…Maquet Cardiopulmonary, AG2006—
+1K200109OXY-1 SystemAbiomed, Inc.2020Class I

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200220
200330
200630
200820
200930
201131
201310
201422
201710
202031
202210
202310
202430
202510
202610
FIG. 1 — Lineage of K991864, 2000–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

4 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KFM device

  1. #1 K250326 Quantum Perfusion Centrifugal Blood Pump…score 11.4 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—Medtronic Bio-Medicus Bio-Console Model 550
Named without a submission number
—
*—Medtronic Bio-Medicus Bio-Pump Centrifugal Blood Pump BP 80
Named without a submission number
—
*—St. Jude Medical Pulse 2100 Centrifugal Pump Interface
Named without a submission number
—
*—Medtronic Bio-Probe DP-38
Named without a submission number
—

What descends from it? (30)

See all 30

9 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K061072Rotaflow Centrifugal Pump with Safeline Coating, Model BSQ-RF-32Maquet Cardiopulmonary, AG200610
K011838Centrifugal PumpStockert Instrumente GmbH20024
K191407Novalung SystemFresenius Medical Care Renal Therapies Group, LLC20203
K200109OXY-1 System Class IAbiomed, Inc.20202
K102726Cardiohelp System Class IMaquet Cardiopulmonary, AG20112
K170029FloPump 32mL, Sterile, FloPump 32mL, Non-SterileInternational Biophysics Corporation20171
K021694Cardiovention Powerbase Console, Model PBC-100Cardiovention, Inc.20031
K232788Inspira ART100Inspira Technologies Oxy B.H.N., Ltd.20240
K020571Stockert Centrifugal Pump ConsoleStockert Instrumente GmbH20020

Has anything in its family been recalled? (4 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 7, 2020Class IIIZ-2957-2020Labeling Change ControlInstruction For Use (IFU) references incorrect serial number of 90437000. It should state This document applies to ROTAFLOW Consoles as of serial number 94175452 affected by the implementation of the new IEC standard 60601 1 2, 4th edition
Dec 3, 2019Class IIZ-0678-2020Nonconforming Material/ComponentA loose ODU plug / Coaxial cable connection may result in fluctuating flow values on the Console.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K200109OXY-1 SystemAbiomed, Inc. · 2020 · Cites it directly · ◆ in the 2020 descendants bar of FIG. 11 Class I recall
  • K102726Cardiohelp SystemMaquet Cardiopulmonary, AG · 2011 · Cites it directly · ◆ in the 2011 descendants bar of FIG. 11 Class I recall
  • K133598Cardiohelp SystemMaquet Cardiopulmonary, AG · 2014 · 2 generations on · ◆ in the 2014 descendants bar of FIG. 11 Class I recall
  • K141432Neonatal Venous Hardshell Cardiotomy Reservior, Pediatric Venous Hardshell Cardiotomy Reservior, Neonatal…Maquet Cardiopulmonary, AG · 2014 · 5 generations on · ◆ in the 2014 descendants bar of FIG. 12 Class I recalls

What does this lineage say?

  • Of its descendants, K061072 (Maquet Cardiopulmonary, AG, 2006) has gone furthest, with 10 descendants of its own.

How does its category compare?

All KFM devices

KFM Pump, Blood, Cardiopulmonary Bypass, Non-Roller Type: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · KFM

2in 2025

2008: 42009: 52010: 32011: 72013: 22017: 12019: 32020: 32021: 32023: 32025: 2
2007high 7 (2011) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KFM

30in 2023

2008: 194 days median2009: 184 days median2010: 109 days median2011: 16 days median2019: 147 days median2020: 30 days median2021: 534 days median2023: 30 days median
2007high 534 (2021) · low 16 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KFM

2in 2024

2008: 22009: 42010: 22011: 12012: 52018: 12019: 3, 1 Class I2024: 2
2007high 5 (2012) · low 0 (2007)2026

Class I (1 since 2007) Class II/III not classified

Adverse event reports per year · KFM

249in 2025

2007: 602008: 622009: 852010: 812011: 572012: 822013: 1042014: 3472015: 2992016: 1562017: 1212018: 1762019: 3112020: 4372021: 4462022: 2942023: 3462024: 2962025: 249
2007high 446 (2021) · low 57 (2011)2025

MAUDE reports naming product code KFM, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—060
20084194262
20095184485
20103109281
2011716157
20120—582
20132—0104
20140—0347
20150—0299
20160—0156
20171—0121
20180—1176
201931473 (1)311
20203300437
202135340446
20220—0294
20233300346
20240—2296
20252—0249
20260—0—

What is it indicated for?

The RotaFlow Centrifugal Pump System is intended for use in an extracorporeal perfusion circuit to pump blood during short duration cardiopulmonary bypass procedures lasting 6 hours or less.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K991864: Rotaflow Centrifugal Pump System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K991864

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026