510(k) · Cleared
ClearedEVA15 insufflator
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Palliare, Ltd. · HIF Insufflator, Laparoscopic
Cleared on May 2, 2024 as substantially equivalent to K222901 (Palliare, Ltd., 2023) after a 227-day review. Its family tree reaches back 12 generations to 1997. 3 of the 35 devices in its family, counting itself, have a Class I recall.
- Decided
- May 2, 2024
- Review
- 227 days
- Class
- Class II
- Product code
- HIF
- Recalls
- None
- Predicates
- 1
Its longest line of descent runs back 12 generations to K964171 (1997). 34 ancestors in all, the oldest K903419 (1990). No later device cites it as a predicate yet. 3 of the 35 devices in this family, counting this one, have a Class I recall.
- Root · 12 generations backK964171 Vasoview Balloon Dissection SystemOrigin Medsystems, Inc. · 1997
- 8 generations between
- 3 generations backK193520 EVA 15 InsufflatorPalliare, Ltd. · 2020
- 2 generations backK202799 EVA15Palliare, Ltd. · 2020
- 1 generation backK222901 EVA15 insufflatorPalliare, Ltd. · 2023
- This deviceK232902 EVA15 insufflatorPalliare, Ltd. · 2024
Show all 12 generations
- Root · 12 generations backK964171 Vasoview Balloon Dissection SystemOrigin Medsystems, Inc. · 1997
- 11 generations backK981700 Vasoview Dissection/Vessel Harvesting SystemOrigin Medsystems, Inc. · 1998
- 10 generations backK030512 Guidant Vasoview Endoscopic Vessel Harvesting SystemGuidant Corp. · 2003 · Recalled
- 9 generations backK060723 40 L High Flow Insufflator F113W.O.M. World of Medicine AG · 2006
- 8 generations backK063367 45L Core Insufflator, Model F114W.O.M. World of Medicine AG · 2007 · Recalled
- 7 generations backK110294 High Flow Insufflation Unit Uhi-4Olympus Medical Systems Corporation · 2011 · Recalled (Class I)
- 6 generations backK122180 High Flow Insufflantion UnitOlympus Medical Systems Corporation · 2012 · Recalled (Class I)
- 5 generations backK143404 SurgiQuest AirSeal iFS SystemSurgiquest, Inc. · 2015 · Recalled
- 4 generations backK172516 SurgiQuest AIRSEAL iFS SystemConmed Corporation · 2018
- 3 generations backK193520 EVA 15 InsufflatorPalliare, Ltd. · 2020
- 2 generations backK202799 EVA15Palliare, Ltd. · 2020
- 1 generation backK222901 EVA15 insufflatorPalliare, Ltd. · 2023
- This deviceK232902 EVA15 insufflatorPalliare, Ltd. · 2024
Class I recalls in this family
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −12 | K964171 | Vasoview Balloon Dissection System | Origin Medsystems, Inc. | 1997 | — |
| −11 | K981700 | Vasoview Dissection/Vessel Harvesting System | Origin Medsystems, Inc. | 1998 | — |
| −10 | K030512 | Guidant Vasoview Endoscopic Vessel Harvesting System | Guidant Corp. | 2003 | 1 |
| −9 | K060723 | 40 L High Flow Insufflator F113 | W.O.M. World of Medicine AG | 2006 | — |
| −8 | K063367 | 45L Core Insufflator, Model F114 | W.O.M. World of Medicine AG | 2007 | 4 |
| −7 | K110294 | High Flow Insufflation Unit Uhi-4 | Olympus Medical Systems Corporation | 2011 | Class I |
| −6 | K122180 | High Flow Insufflantion Unit | Olympus Medical Systems Corporation | 2012 | Class I |
| −5 | K143404 | SurgiQuest AirSeal iFS System | Surgiquest, Inc. | 2015 | 11 |
| −4 | K172516 | SurgiQuest AIRSEAL iFS System | Conmed Corporation | 2018 | — |
| −3 | K193520 | EVA 15 Insufflator | Palliare, Ltd. | 2020 | — |
| −2 | K202799 | EVA15 | Palliare, Ltd. | 2020 | — |
| −1 | K222901 | EVA15 insufflator | Palliare, Ltd. | 2023 | — |
| 0 (this device) | K232902 | EVA15 insufflator | Palliare, Ltd. | 2024 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1990 | 1 | 0 |
| 1991 | 2 | 0 |
| 1994 | 1 | 0 |
| 1996 | 2 | 0 |
| 1999 | 2 | 0 |
| 2000 | 1 | 0 |
| 2001 | 2 | 0 |
| 2002 | 1 | 1 |
| 2003 | 3 | 0 |
| 2004 | 1 | 0 |
| 2006 | 1 | 0 |
| 2007 | 2 | 0 |
| 2008 | 1 | 0 |
| 2009 | 1 | 0 |
| 2011 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new HIF device
- #1 K250388 Endoflator +score 15.8 · 10 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K222901 | EVA15 insufflatorSummary page 4…21 CFR 884.1730 Regulation Name: Laparoscopic insufflator Product Code: HIF Predicate Device: K222901 – Palliare EVA15 Reference Device: K131402 – Buffalo Filter VISICLEAR Modifications: We have provided an acc… | 2023 |
| † | K131402 | Visiclear Smoke Evacuation System | 2014 |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (3 Class I)
No FDA recall names this device.
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K122180High Flow Insufflantion UnitOlympus Medical Systems Corporation · 2012 · 6 generations back · ◆ on the line of descent in FIG. 11 Class I recall
- K110294High Flow Insufflation Unit Uhi-4Olympus Medical Systems Corporation · 2011 · 7 generations back · ◆ on the line of descent in FIG. 11 Class I recall
- K014166Uhi-3 High Flow Insufflation UnitOlympus Optical Co., Ltd. · 2002 · 8 generations back · ◆ in the 2002 ancestors bar of FIG. 11 Class I recall
What does this lineage say?
- The line of descent spans 27 years and passes through 7 companies; a new generation was cleared every 2.3 years on median.
- 9 of the 34 ancestors have a recall of any class; 3 have a Class I recall.
- 1 other device also cites its predicate K222901.
How does its category compare?
All HIF devicesHIF Insufflator, Laparoscopic: FDA decisions, review times, recalls and adverse event reports over 20 years. 2024, this device’s year, is highlighted.
FDA decisions per year · HIF
3in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · HIF
210in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · HIF
12so far in 2026, the most since 2007
Class I (1 since 2007) Class II/III not classified
Adverse event reports per year · HIF
1,507in 2025
MAUDE reports naming product code HIF, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 4 | 140 | 0 | 8 |
| 2008 | 2 | — | 1 | 14 |
| 2009 | 4 | 191 | 0 | 21 |
| 2010 | 2 | — | 0 | 41 |
| 2011 | 5 | 128 | 0 | 44 |
| 2012 | 5 | 138 | 0 | 38 |
| 2013 | 2 | — | 0 | 40 |
| 2014 | 2 | — | 0 | 43 |
| 2015 | 2 | — | 0 | 30 |
| 2016 | 3 | 88 | 1 | 72 |
| 2017 | 5 | 242 | 0 | 68 |
| 2018 | 3 | 207 | 1 | 97 |
| 2019 | 2 | — | 0 | 128 |
| 2020 | 4 | 64 | 0 | 360 |
| 2021 | 1 | — | 0 | 654 |
| 2022 | 1 | — | 0 | 667 |
| 2023 | 7 | 134 | 1 (1) | 890 |
| 2024 | 4 | 170 | 0 | 1,483 |
| 2025 | 3 | 210 | 0 | 1,507 |
| 2026 | 3 | 265 | 12 | — |
What is it indicated for?
The EVA15 Insufflator is intended for use in diagnostic and/or therapeutic endoscopic and laparoscopic procedures to
distend the abdomen, rectum, colon, esophagus, stomach or thoracic cavity with up to 25 mmHg pressure, by filling it
with gas and to evacuate surgical smoke.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K232902 · printed Nov 4, 2026
Cite this page
510(k) Report. “K232902: EVA15 insufflator.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K232902
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026