510(k) Report

510(k) · Cleared

Cleared

DEKA ACE Pump System

K233952510(k) TraditionalCleared Mar 13, 202489-day reviewClass II21 CFR 880.5730

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Deka Research and Development · QFG Alternate Controller Enabled Insulin Infusion Pump

Cleared on Mar 13, 2024 as substantially equivalent to K213536 (Deka Research and Development, 2023) after an 89-day review. Its family tree reaches back 2 generations to 2019. 3 later devices descend from it.

Decided
Mar 13, 2024
Review
89 days
Class
Class II
Product code
QFG
Recalls
None
Predicates
1

Its longest line of descent runs back 2 generations to DEN180058 (2019). 3 later devices descend from it, 2 citing it directly.

Descendants, by year cleared 2025: 1 descendant cleared2024: 1 descendant cleared2026: 1 descendant clearedDEN180058 · t:slim X2 insulin pump with interoperable technology · Tandem Diabetes Care, Inc. · 2019K213536 · DEKA ACE Pump System · Deka Research and Development · 2023201620182020202220242026
Descendants, by generation 1 generation forward: 2 descendants2 generations forward: 1 descendant K213536 · DEKA ACE Pump System · Deka Research and Development · 2023DEN180058 · t:slim X2 insulin pump with interoperable technology · Tandem Diabetes Care, Inc. · 20190
  1. Root · 2 generations backDEN180058 t:slim X2 insulin pump with interoperable technologyTandem Diabetes Care, Inc. · 2019
  2. 1 generation backK213536 DEKA ACE Pump SystemDeka Research and Development · 2023
  3. This deviceK233952 DEKA ACE Pump SystemDeka Research and Development · 2024
  4. Descendant · cites it directly · 1 of its ownK250930 twiist systemDeka Research & Development Corp. · 2025
  5. Descendant · 2 generations on · 0 of its ownK261672 twiist Automated Insulin Delivery (AID) SystemDeka Research and Development · 2026
  6. Descendant · cites it directly · 0 of its ownK241178 DEKA ACE Pump SystemDeka Research and Development · 2024

Most influential descendants

Showing 3 of 3 · See all
  • K250930twiist systemDeka Research & Development Corp. · 2025 · cites it directly1descendant
  • K261672twiist Automated Insulin Delivery (AID) SystemDeka Research and Development · 2026 · 2 generations on0descendants
  • K241178DEKA ACE Pump SystemDeka Research and Development · 2024 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2DEN180058t:slim X2 insulin pump with interoperable technologyTandem Diabetes Care, Inc.2019—
−1K213536DEKA ACE Pump SystemDeka Research and Development2023—
0 (this device)K233952DEKA ACE Pump SystemDeka Research and Development2024—
+1K250930twiist systemDeka Research & Development Corp.2025—
+2K261672twiist Automated Insulin Delivery (AID) SystemDeka Research and Development2026—
+1K241178DEKA ACE Pump SystemDeka Research and Development2024—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202410
202510
202610
FIG. 1 — Lineage of K233952, 2019–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in QFG? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new QFG device

  1. #1 K251032 MiniMed 780G insulin pumpscore 19.6 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K213536DEKA ACE Pump System
Summary page 4
…: QFG Class: II Device Panel: Clinical Chemistry Predicate Device The predicate device for this submission is the DEKA ACE Pump System cleared on 7/25/2023 under 510(k) K213536. Device Description The DEKA ACE Pump system is a modification of the previously cleared DEKA ACE Pump System (K213536). The mo…
Deka Research and Development2023—

What descends from it? (3)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K250930twiist systemDeka Research & Development Corp.20251
K241178DEKA ACE Pump SystemDeka Research and Development20240

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 5 years and passes through 2 companies; a new generation was cleared every 4.4 years on median.
  • None of the 2 ancestors has a recall.
  • Of its descendants, K250930 (Deka Research & Development Corp., 2025) has gone furthest, with 1 descendant of its own.

How does its category compare?

All QFG devices

QFG Alternate Controller Enabled Insulin Infusion Pump: FDA decisions, review times, recalls and adverse event reports over 20 years. 2024, this device’s year, is highlighted.

FDA decisions per year · QFG

5in 2025

2019: 22020: 12022: 22023: 92024: 52025: 52026 (partial year): 6
2007high 9 (2023) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · QFG

44in 2025

2023: 150 days median2024: 89 days median2025: 44 days median2026 (partial year): 103 days median
2007high 150 (2023) · low 44 (2025)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · QFG

8so far in 2026, the most since 2007

2022: 12023: 1, 1 Class I2025: 3, 1 Class I2026 (partial year): 8, 1 Class I
2007high 8 (2026) · low 0 (2007)2026

Class I (3 since 2007) Class II/III not classified

Adverse event reports per year · QFG

324,360in 2025

2020: 30,5222021: 74,7012022: 129,1192023: 188,0142024: 273,1802025: 324,360
2007high 324,360 (2025) · low 0 (2007)2025

MAUDE reports naming product code QFG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20100—00
20110—00
20120—00
20130—00
20140—00
20150—00
20160—00
20170—00
20180—00
20192—00
20201—030,522
20210—074,701
20222—1129,119
202391501 (1)188,014
20245890273,180
20255443 (1)324,360
202661038 (1)—

What is it indicated for?

The DEKA ACE Pump System is intended for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin, ages six and above. The pump is able to reliably and securely communicate with compatible, digitally connected devices, including automated insulin dosing software, to receive, execute, and confirm commands from these devices. The pump is intended for single patient, home use and…

Read the full statement (117 words)

The DEKA ACE Pump System is intended for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin, ages six and above. The pump is able to reliably and securely communicate with compatible, digitally connected devices, including automated insulin dosing software, to receive, execute, and confirm commands from these devices. The pump is intended for single patient, home use and requires a prescription.

The bolus calculator is indicated for use for aiding the user in determining the bolus insulin dosage for management of diabetes mellitus based on consumed carbohydrates, operator entered blood glucose, insulin sensitivity, insulin to carbohydrate ratio, target glucose values, and current insulin on board.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K233952: DEKA ACE Pump System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K233952

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026