510(k) Report

510(k) · Cleared

Recalled · Class II · 2026

Swan-Ganz Pacing Probe and Catheters; Swan-Ganz Bipolar Pacing Catheter (D97120F5); Swan-Ganz Bipolar Pacing Catheter (D97130F5); Swan-Ganz Pacing Thermodilution TD Catheter (D200F7); Chandler Transluminal V-Pacing Probe (D98100)

K233983510(k) TraditionalCleared Jun 27, 2024192-day reviewClass II21 CFR 870.1240Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Edwards Lifesciences · DYG Catheter, Flow Directed

Cleared on Jun 27, 2024 as substantially equivalent to K791183 (Edwards Lifesciences, LLC, 1979) and 2 other predicates after a 192-day review. Its family tree reaches back 2 generations to 1982. It has been recalled once.

Decided
Jun 27, 2024
Review
192 days
Class
Class II
Product code
DYG
Recalls
1
Predicates
3

Its longest line of descent runs back 2 generations to K813521 (1982). 4 ancestors in all, the oldest K791183 (1979). No later device cites it as a predicate yet.

Other ancestors, by year cleared 1982: 1 other ancestor1979: 1 other ancestorK813521 · Transluminal Bipolar Pacing Probe · American Edwards Laboratories · 1982K152633 · Chandler Transluminal Bipolar Pacing Probe, Flextip Transluminal Bipolar Pacing Probe · Edwards Lifesciences, LLC · 201519801990200020102020
  1. Root · 2 generations backK813521 Transluminal Bipolar Pacing ProbeAmerican Edwards Laboratories · 1982
  2. 1 generation backK152633 Chandler Transluminal Bipolar Pacing Probe, Flextip Transluminal…Edwards Lifesciences, LLC · 2015
  3. This deviceK233983 Swan-Ganz Pacing Probe and Catheters; Swan-Ganz Bipolar Pacing…Edwards Lifesciences · 2024
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K813521Transluminal Bipolar Pacing ProbeAmerican Edwards Laboratories1982—
−1K152633Chandler Transluminal Bipolar Pacing Probe, Flextip Transluminal Bipolar…Edwards Lifesciences, LLC2015—
0 (this device)K233983Swan-Ganz Pacing Probe and Catheters; Swan-Ganz Bipolar Pacing Catheter…Edwards Lifesciences2024—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
197910
198210
FIG. 1 — Lineage of K233983, 1982–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DYG? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DYG device

  1. #1 K242451 HemoSphere Alta Advanced Monitoring Platform…score 24.2 · 39 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K791183Swan-Ganz Flow-Directed Catheter
Summary page 5
…Cardiovascular permanent or temporary pacemaker electrode Regulation Class Class II Swan-Ganz Flow Directed Catheter (K791183, cleared on July 12, 1979) Predicate Device Swan-Ganz Bipolar Pacing Catheter (K822723, cleared on…
Edwards Lifesciences, LLC1979—
*K822723Swan-Ganz Bipolar Pacing Catheter
Summary page 5
…-Ganz Flow Directed Catheter (K791183, cleared on July 12, 1979) Predicate Device Swan-Ganz Bipolar Pacing Catheter (K822723, cleared on Oct 15, 1982) Chandler Transluminal Bipolar Pacing Probe (K152633, clea…
American Edwards Laboratories1982—
*K152633Chandler Transluminal Bipolar Pacing Probe, Flextip Transluminal Bipolar Pacing Probe
Summary page 5
…polar Pacing Catheter (K822723, cleared on Oct 15, 1982) Chandler Transluminal Bipolar Pacing Probe (K152633, cleared Oct 15, 2015) The Swan-Ganz Pacing Catheters and Pacing Probes are a series of…
Edwards Lifesciences, LLC2015—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 8, 2026Class IIZ-2204-2026Process change controlCatheters may, due to a manufacturing process and material changes, leak/break from the (blue) proximal injectate lumen hub, which may lead to infection, medication loss, and/or blood loss.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 42 years and passes through 3 companies; a new generation was cleared every 33.6 years on median.
  • None of the 4 ancestors has a recall.

How does its category compare?

All DYG devices

DYG Catheter, Flow Directed: FDA decisions, review times, recalls and adverse event reports over 20 years. 2024, this device’s year, is highlighted.

FDA decisions per year · DYG

3in 2024

2008: 12011: 12016: 12020: 12022: 12023: 12024: 3
2007high 3 (2024) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DYG

188in 2024

2024: 188 days median
2007high 188 (2024)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DYG

2in 2019

2007: 12009: 32013: 12018: 4, 4 Class I2019: 22026 (partial year): 2
2007high 4 (2018) · low 0 (2008)2026

Class I (4 since 2007) Class II/III not classified

Adverse event reports per year · DYG

268in 2025

2007: 1682008: 4002009: 1452010: 1472011: 2492012: 2212013: 1492014: 1782015: 1632016: 1542017: 1152018: 1332019: 1442020: 1042021: 972022: 1302023: 2292024: 2262025: 268
2007high 400 (2008) · low 97 (2021)2025

MAUDE reports naming product code DYG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—1168
20081—0400
20090—3145
20100—0147
20111—0249
20120—0221
20130—1149
20140—0178
20150—0163
20161—0154
20170—0115
20180—4 (4)133
20190—2144
20201—0104
20210—097
20221—0130
20231—0229
202431880226
20250—0268
20260—2—

What is it indicated for?

The Swan-Ganz pacing-TD catheters are indicated for use for temporary, atrial, ventricular, or A-V sequential cardiac pacing of adult surgical and/or critically ill patients. This includes, but is not limited to, patients with arrythmias, acute myocardial infarction, cardiac surgery, and coronary angiography. In addition, these catheters are indicated for use in hemodynamic monitoring of patients experiencing medical conditions such as, but not limited…

Read the full statement (156 words)

The Swan-Ganz pacing-TD catheters are indicated for use for temporary, atrial, ventricular, or A-V sequential cardiac pacing of adult surgical and/or critically ill patients. This includes, but is not limited to, patients with arrythmias, acute myocardial infarction, cardiac surgery, and coronary angiography. In addition, these catheters are indicated for use in hemodynamic monitoring of patients experiencing medical conditions such as, but not limited to, post-major surgical recovery, trauma, sepsis, burns, pulmonary disease, pulmonary failure, stroke, and cardiac disease including heart failure.

The Swan-Ganz bipolar pacing catheters are indicated for use in surgical and/or critically ill cardiac patients who may benefit from temporary, transvenous right ventricular pacing.

The Chandler transluminal V-pacing probe is a transluminal bipolar pacing probe, which must be used with a Swan-Ganz Paceport catheter is indicated for use in adult surgical and/or critically ill cardiac patients. This includes, but is not limited to, patients with arrhythmias, acute myocardial infarction, cardiac surgery and coronary angiography.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K233983: Swan-Ganz Pacing Probe and Catheters; Swan-Ganz Bipolar Pacing Catheter (D97120F5); Swan-Ganz Bipolar Pacing Catheter (D97130F5); Swan-Ganz Pacing Thermodilution TD Catheter (D200F7); Chandler Transluminal V-Pacing Probe (D98100).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K233983

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026