510(k) Report

510(k) · Cleared

Cleared

Babyroo TN300

K243606510(k) TraditionalCleared Jan 17, 202557-day reviewClass II21 CFR 880.5130

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Draeger Medical Systems, Inc. · FMT Warmer, Infant Radiant

Cleared on Jan 17, 2025 as substantially equivalent to K230278 (Draeger Medical Systems, Inc., 2023) after a 57-day review. Its family tree reaches back 3 generations to 1989.

Decided
Jan 17, 2025
Review
57 days
Class
Class II
Product code
FMT
Recalls
None
Predicates
1

Its longest line of descent runs back 3 generations to K892885 (1989). 5 ancestors in all. No later device cites it as a predicate yet.

  1. Root · 3 generations backK892885 Resubmitted RD1000 Neopuff Infant ResuscitatorFisher & Paykel Electronics, Ltd. · 1989
  2. 2 generations backK120642 Resuscitaire with AutobreathDraeger Medical Systems, Inc. · 2012 · Recalled
  3. 1 generation backK230278 Babyroo TN300Draeger Medical Systems, Inc. · 2023
  4. This deviceK243606 Babyroo TN300Draeger Medical Systems, Inc. · 2025
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K892885Resubmitted RD1000 Neopuff Infant ResuscitatorFisher & Paykel Electronics, Ltd.1989—
−2K120642Resuscitaire with AutobreathDraeger Medical Systems, Inc.20121
−1K230278Babyroo TN300Draeger Medical Systems, Inc.2023—
0 (this device)K243606Babyroo TN300Draeger Medical Systems, Inc.2025—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199010
200010
FIG. 1 — Lineage of K243606, 1989–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FMT? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FMT device

  1. #1 K230278 Babyroo TN300score 8.7 · 2 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K230278Babyroo TN300
Summary page 5
…ices The purpose of this premarket notification is the modification of the subject Babyroo TN300 infant warmer device, which was originally cleared under K230278. The modification consists of the activation, for the US market, of existing optional software functionality to f…
Draeger Medical Systems, Inc.2023—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 36 years and passes through 2 companies; a new generation was cleared every 10.6 years on median.
  • 2 of the 5 ancestors have a recall of any class.

How does its category compare?

All FMT devices

FMT Warmer, Infant Radiant: FDA decisions, review times, recalls and adverse event reports over 20 years. 2025, this device’s year, is highlighted.

FDA decisions per year · FMT

1in 2025

2007: 22009: 12010: 12011: 12012: 42013: 12018: 12020: 22021: 12022: 12023: 12025: 1
2007high 4 (2012) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FMT

103in 2012

2012: 103 days median
2007high 103 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FMT

4in 2019

2008: 12009: 22013: 22014: 12015: 12016: 12018: 42019: 4, 4 Class I2026 (partial year): 2, 2 Class I
2007high 4 (2018) · low 0 (2007)2026

Class I (6 since 2007) Class II/III not classified

Adverse event reports per year · FMT

359in 2025

2007: 212008: 1142009: 802010: 1062011: 792012: 542013: 552014: 792015: 562016: 562017: 462018: 792019: 1,4372020: 1,6052021: 4732022: 1532023: 1752024: 2572025: 359
2007high 1,605 (2020) · low 21 (2007)2025

MAUDE reports naming product code FMT, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—021
20080—1114
20091—280
20101—0106
20111—079
20124103054
20131—255
20140—179
20150—156
20160—156
20170—046
20181—479
20190—4 (4)1,437
20202—01,605
20211—0473
20221—0153
20231—0175
20240—0257
20251—0359
20260—2 (2)—

What is it indicated for?

The Babyroo TN300 is an open care radiant warmer that provides a controlled source of heat and regulation of skin temperature for neonates and infants. The optional integrated resuscitation module provides emergency respiratory support administered by clinicians and includes the functionality of suction. Additionally, the device provides weighing (optional) and pulse oximetry (optional) of neonates and infants. The device is designed for use with patients with a body weight up to 10 kg (22 lb).

The device is indicated for thermoregulation, skin temperature regulation, weighing (optional), pulse oximetry (optional), and resuscitation (optional) of neonates and infants.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K243606: Babyroo TN300.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K243606

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026