510(k) Report

510(k) · Cleared

Recalled · Class I · 2026

Panda iRes Warmer, Giraffe Warmer

K201628510(k) TraditionalCleared Feb 12, 2021241-day reviewClass II21 CFR 880.5130Recalls 3 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

GE Healthcare · FMT Warmer, Infant Radiant

Cleared on Feb 12, 2021 as substantially equivalent to K122267 (Ohmeda Medical, A Division of Datex-Ohmeda, Inc., 2012) after a 241-day review. Its family tree reaches back 2 generations to 2010. 3 of the 3 devices in its family, counting itself, have a Class I recall. It has been recalled 3 times, including a Class I recall.

Decided
Feb 12, 2021
Review
241 days
Class
Class II
Product code
FMT
Recalls
3 (1 Class I)
Predicates
1

Its longest line of descent runs back 2 generations to K101804 (2010). No later device cites it as a predicate yet. 3 of the 3 devices in this family, counting this one, have a Class I recall.

  1. Root · 2 generations backK101804 Giraffe and Panda WarmersOhmeda Medical, A Division of Datex-Ohmeda, Inc. · 2010 · Recalled (Class I)
  2. 1 generation backK122267 Giraffe and Panda WarmersOhmeda Medical, A Division of Datex-Ohmeda, Inc. · 2012 · Recalled (Class I)
  3. This deviceK201628 Panda iRes Warmer, Giraffe WarmerGE Healthcare · 2021

Class I recalls in this family

View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K101804Giraffe and Panda WarmersOhmeda Medical, A Division of Datex-Ohmeda, Inc.2010Class I
−1K122267Giraffe and Panda WarmersOhmeda Medical, A Division of Datex-Ohmeda, Inc.2012Class I
0 (this device)K201628Panda iRes Warmer, Giraffe WarmerGE Healthcare2021—
FIG. 1 — Lineage of K201628, 2010–2021Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FMT device

  1. #1 K243606 Babyroo TN300score 9.3 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K122267Giraffe and Panda Warmers
Summary page 4
…21 CFR 880.5130 Infant radiant warmer. Classification II Product Code: FMT Predicate Device: Panda Warmer, Giraffe Warmer (K122267) R-81 Indications for U…
Ohmeda Medical, A Division of Datex-Ohmeda, Inc.2012Class I

Also cleared against K122267: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (3 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 8, 2026Class IZ-2358-2026Process controlBlender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not match the intended concentration set by the user, potentially resulting in hypoxia or…
Aug 9, 2024Class IIZ-3080-2024Process controlGE HealthCare has become aware that complete electrical safety testing was not conducted during manufacturing of certain Giraffe and Panda iRes Warmers. There are multiple electrical safety protections included in the product design.…
Aug 9, 2024Class IIZ-3081-2024Process controlGE HealthCare has become aware that complete electrical safety testing was not conducted during manufacturing of certain Giraffe and Panda iRes Warmers. There are multiple electrical safety protections included in the product design.…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K122267Giraffe and Panda WarmersOhmeda Medical, A Division of Datex-Ohmeda, Inc. · 2012 · Direct predicate · ◆ on the line of descent in FIG. 11 Class I recall
  • K101804Giraffe and Panda WarmersOhmeda Medical, A Division of Datex-Ohmeda, Inc. · 2010 · 2 generations back · ◆ on the line of descent in FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 11 years and passes through 2 companies; a new generation was cleared every 8.4 years on median.
  • 2 of the 2 ancestors have a recall of any class; 2 have a Class I recall.
  • 1 other device also cites its predicate K122267.

How does its category compare?

All FMT devices

FMT Warmer, Infant Radiant: FDA decisions, review times, recalls and adverse event reports over 20 years. 2021, this device’s year, is highlighted.

FDA decisions per year · FMT

1in 2025

2007: 22009: 12010: 12011: 12012: 42013: 12018: 12020: 22021: 12022: 12023: 12025: 1
2007high 4 (2012) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FMT

103in 2012

2012: 103 days median
2007high 103 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FMT

4in 2019

2008: 12009: 22013: 22014: 12015: 12016: 12018: 42019: 4, 4 Class I2026 (partial year): 2, 2 Class I
2007high 4 (2018) · low 0 (2007)2026

Class I (6 since 2007) Class II/III not classified

Adverse event reports per year · FMT

359in 2025

2007: 212008: 1142009: 802010: 1062011: 792012: 542013: 552014: 792015: 562016: 562017: 462018: 792019: 1,4372020: 1,6052021: 4732022: 1532023: 1752024: 2572025: 359
2007high 1,605 (2020) · low 21 (2007)2025

MAUDE reports naming product code FMT, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—021
20080—1114
20091—280
20101—0106
20111—079
20124103054
20131—255
20140—179
20150—156
20160—156
20170—046
20181—479
20190—4 (4)1,437
20202—01,605
20211—0473
20221—0153
20231—0175
20240—0257
20251—0359
20260—2 (2)—

What is it indicated for?

Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermoregulate based on their own physiology. Infant radiant warmers may be used to facilitate the neonate's transition to the external environment or to provide a controlled open environment. An optional integrated SpO2 monitoring feature may be used for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (Sp02) and…

Read the full statement (114 words)

Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermoregulate based on their own physiology. Infant radiant warmers may be used to facilitate the neonate's transition to the external environment or to provide a controlled open environment. An optional integrated SpO2 monitoring feature may be used for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (Sp02) and pulse rate (measured by a SpO2 sensor). An optional integrated resuscitation system may be used to provide the basic equipment required for pulmonary resuscitation of infants. Pulmonary resuscitation includes practices necessary to establish a clear airway and provide oxygen or air/oxygen mixtures and/or manual ventilation to the infant.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K201628: Panda iRes Warmer, Giraffe Warmer.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K201628

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026