Giraffe Omnibed Carestation CS1

K213551510(k) TraditionalCleared Mar 2, 2022114-day reviewClass II21 CFR 880.5130Recalls 5 (2 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Datex Ohmeda, Inc. · FMT Warmer, Infant Radiant

Cleared on Mar 2, 2022 as substantially equivalent to K152809 (Ohmeda Medical, A Division of Datex-Ohmeda, Inc., 2015) after a 114-day review. Its family tree reaches back 4 generations to 1997. 3 of the 10 devices in its family, counting itself, have a Class I recall. 2 later devices descend from it. It has been recalled 5 times, including 2 Class I recalls.

Decided
Mar 2, 2022
Review
114 days
Class
Class II
Product code
FMT
Recalls
5 (2 Class I)
Predicates
1

Its longest line of descent runs back 4 generations to K954204 (1997). 7 ancestors in all. 2 later devices descend from it, 1 citing it directly. 3 of the 10 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2010: 1 other ancestor, 1 with a Class I recall12001: 1 other ancestor2000: 1 other ancestor2025: 1 descendant cleared2026: 1 descendant clearedK954204 · Drager Incubator 8000 IC · Drager, Inc. · 1997K020547 · Ohmeda Medical Giraffe Incubator · Ohmeda Medical · 2002 · recalledK072512 · Modification to: Ohmeda Medical Giraffe Incubator (Giraffe and Panda Uninterruptible Power Supply) · Datex-Ohmeda, Inc. · 2007K152809 · Giraffe Incubator Carestation CS1 · Ohmeda Medical, A Division of Datex-Ohmeda, Inc. · 2015 · Class I recall1995200020052010201520202025
Other ancestors, by generationDescendants, by generation 2 generations back: 2 other ancestors, 1 with a Class I recall13 generations back: 1 other ancestor 1 generation forward: 1 descendant2 generations forward: 1 descendant K152809 · Giraffe Incubator Carestation CS1 · Ohmeda Medical, A Division of Datex-Ohmeda, Inc. · 2015 · Class I recallK072512 · Modification to: Ohmeda Medical Giraffe Incubator (Giraffe and Panda Uninterruptible Power Supply) · Datex-Ohmeda, Inc. · 2007K020547 · Ohmeda Medical Giraffe Incubator · Ohmeda Medical · 2002 · recalledK954204 · Drager Incubator 8000 IC · Drager, Inc. · 19970
  1. Root · 4 generations backK954204 Drager Incubator 8000 ICDrager, Inc. · 1997
  2. 3 generations backK020547 Ohmeda Medical Giraffe IncubatorOhmeda Medical · 2002 · Recalled
  3. 2 generations backK072512 Modification to: Ohmeda Medical Giraffe Incubator (Giraffe and Panda…Datex-Ohmeda, Inc. · 2007
  4. 1 generation backK152809 Giraffe Incubator Carestation CS1Ohmeda Medical, A Division of Datex-Ohmeda, Inc. · 2015 · Recalled (Class I)
  5. This deviceK213551 Giraffe Omnibed Carestation CS1Datex Ohmeda, Inc. · 2022
  6. Descendant · cites it directly · 1 of its ownK251663 Giraffe OmniBed Carestation (CS1); Giraffe Incubator Carestation (CS1)Datex-Ohmeda, Inc. · 2025
  7. Descendant · 2 generations on · 0 of its ownK260326 Giraffe™ OmniBed™ Carestation (CS1); Giraffe™ Incubator Carestation…Datex-Ohmeda, Inc. · 2026

Class I recalls in this family

Most influential descendants

Showing 2 of 2 · See all
  • K251663Giraffe OmniBed Carestation (CS1); Giraffe Incubator Carestation (CS1)Datex-Ohmeda, Inc. · 2025 · cites it directly1descendant
  • K260326Giraffe™ OmniBed™ Carestation (CS1); Giraffe™ Incubator Carestation (CS1)Datex-Ohmeda, Inc. · 2026 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K954204Drager Incubator 8000 ICDrager, Inc.1997—
−3K020547Ohmeda Medical Giraffe IncubatorOhmeda Medical20024
−2K072512Modification to: Ohmeda Medical Giraffe Incubator (Giraffe and Panda…Datex-Ohmeda, Inc.2007—
−1K152809Giraffe Incubator Carestation CS1Ohmeda Medical, A Division of Datex-Ohmeda, Inc.2015Class I
0 (this device)K213551Giraffe Omnibed Carestation CS1Datex Ohmeda, Inc.2022—
+1K251663Giraffe OmniBed Carestation (CS1); Giraffe Incubator Carestation (CS1)Datex-Ohmeda, Inc.2025—
+2K260326Giraffe™ OmniBed™ Carestation (CS1); Giraffe™ Incubator Carestation (CS1)Datex-Ohmeda, Inc.2026—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200010
200110
201011

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202510
202610
FIG. 1 — Lineage of K213551, 1997–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FMT device

  1. #1 K243606 Babyroo TN300score 9.3 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K152809Giraffe Incubator Carestation CS1
Summary page 4
…Classification Names: Neonatal Incubator Regulation 21 CFR 880.5400 Neonatal Incubator Classification II Product Code: FMZ Predicate Device: Giraffe Incubator (K152809) Indications for Use: The Giraffe Incubator Carestation is an Infant Incubator. Incubators provide heat in a controlled manner…
Ohmeda Medical, A Division of Datex-Ohmeda, Inc.2015Class I

What descends from it? (2)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K251663Giraffe OmniBed Carestation (CS1); Giraffe Incubator Carestation (CS1)Datex-Ohmeda, Inc.20251

Has anything in its family been recalled? (3 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 1, 2026Class IIZ-2477-2026Component design/selectionGE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain conditions, for example, impact during transport, one or more of the fasteners that secure the…
May 1, 2026Class IIZ-2476-2026Component design/selectionGE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain conditions, for example, impact during transport, one or more of the fasteners that secure the…
Sep 3, 2024Class IZ-0048-2025Use errorGE HealthCare has become aware that for certain Giraffe OmniBed and Giraffe OmniBed Carestation devices, the screw that secures the doors that cover the warmer heaters may not have been torqued to specification. This may lead to the doors…
Jun 6, 2023Class IZ-0140-2025Component design/selectionPrior to current established manufacturing practices, higher levels of airborne formaldehyde exposure from newly manufactured incubators could be seen during clinical use that would rapidly decrease over a week. Affected customers were…
May 19, 2023Class IIZ-2100-2023Nonconforming Material/ComponentGE HealthCare has become aware of the potential that an incorrect type of secondary latch could have been installed on the northeast (NE) and/or northwest (NW) bedside panels for certain Giraffe OmniBed and Giraffe OmniBed Carestation…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K152809Giraffe Incubator Carestation CS1Ohmeda Medical, A Division of Datex-Ohmeda, Inc. · 2015 · Direct predicate · ◆ on the line of descent in FIG. 12 Class I recalls
  • K101778Giraffe IncubatorOhmeda Medical · 2010 · 2 generations back · ◆ in the 2010 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 25 years and passes through 5 companies; a new generation was cleared every 6.2 years on median.
  • 4 of the 7 ancestors have a recall of any class; 2 have a Class I recall.
  • Of its descendants, K251663 (Datex-Ohmeda, Inc., 2025) has gone furthest, with 1 descendant of its own.

How does its category compare?

All FMT devices

FMT Warmer, Infant Radiant: FDA decisions, review times, recalls and adverse event reports over 20 years. 2022, this device’s year, is highlighted.

FDA decisions per year · FMT

1in 2025

2007: 22009: 12010: 12011: 12012: 42013: 12018: 12020: 22021: 12022: 12023: 12025: 1
2007high 4 (2012) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FMT

103in 2012

2012: 103 days median
2007high 103 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FMT

4in 2019

2008: 12009: 22013: 22014: 12015: 12016: 12018: 42019: 4, 4 Class I2026 (partial year): 2, 2 Class I
2007high 4 (2018) · low 0 (2007)2026

Class I (6 since 2007) Class II/III not classified

Adverse event reports per year · FMT

359in 2025

2007: 212008: 1142009: 802010: 1062011: 792012: 542013: 552014: 792015: 562016: 562017: 462018: 792019: 1,4372020: 1,6052021: 4732022: 1532023: 1752024: 2572025: 359
2007high 1,605 (2020) · low 21 (2007)2025

MAUDE reports naming product code FMT, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—021
20080—1114
20091—280
20101—0106
20111—079
20124103054
20131—255
20140—179
20150—156
20160—156
20170—046
20181—479
20190—4 (4)1,437
20202—01,605
20211—0473
20221—0153
20231—0175
20240—0257
20251—0359
20260—2 (2)—

What is it indicated for?

The Giraffe OmniBed Carestation is a combination of an infant incubator and an infant warmer. The device can be operated as an incubator or as a warmer and can transition from one mode to the other on user's demand. It cannot be operated in both modes at the same time. Incubators and warmers provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Incubators provide an enclosed…

Read the full statement (138 words)

The Giraffe OmniBed Carestation is a combination of an infant incubator and an infant warmer. The device can be operated as an incubator or as a warmer and can transition from one mode to the other on user's demand. It cannot be operated in both modes at the same time. Incubators and warmers provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology.

Incubators provide an enclosed, temperature-controlled environment and warmers provide infrared heat in an open environment. They may also be used for short periods of time to facilitate the neonate's transition from the uterus to the external environment.

This device may incorporate a Servo Controlled Oxygen Delivery System. This is indicated to provide stable oxygen concentration within the infant compartment at the value set by the operator (21-65%).

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K213551: Giraffe Omnibed Carestation CS1.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K213551

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026