G0 Blood Pressure Monitoring System (G0)

K250415510(k) TraditionalCleared Jul 2, 2025139-day reviewClass II21 CFR 870.1130

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Aktiia SA · DXN System, Measurement, Blood-Pressure, Non-Invasive

Cleared on Jul 2, 2025 as substantially equivalent to K190792 (Biobeat Technologies, Ltd., 2019) and 1 other predicate after a 139-day review. Its family tree reaches back 11 generations to 1997. 1 later device descends from it.

Decided
Jul 2, 2025
Review
139 days
Class
Class II
Product code
DXN
Recalls
None
Predicates
2

Its longest line of descent runs back 11 generations to K962603 (1997). 83 ancestors in all, the oldest K822055 (1982). 1 later device descends from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2024: 1 other ancestor2023: 1 other ancestor2018: 2 other ancestors2017: 2 other ancestors2021: 1 other ancestor2016: 1 other ancestor2015: 1 other ancestor2007: 3 other ancestors2006: 4 other ancestors2005: 5 other ancestors2001: 3 other ancestors2014: 1 other ancestor2013: 2 other ancestors2012: 3 other ancestors2004: 2 other ancestors2003: 4 other ancestors2002: 3 other ancestors2000: 8 other ancestors1996: 4 other ancestors1990: 2 other ancestors1986: 1 other ancestor1997: 1 other ancestor1994: 2 other ancestors1993: 3 other ancestors1991: 2 other ancestors1988: 1 other ancestor1982: 1 other ancestor1999: 2 other ancestors1995: 2 other ancestors2011: 1 other ancestor1998: 1 other ancestor1992: 1 other ancestor1989: 1 other ancestor2026: 1 descendant clearedK962603 · Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & Cables · Masimo Corp. · 1997K973887 · Masimo Set MS-1P Pulse Oximeter and the Lnop Series of Sensors and Cables · Masimo Corp. · 1998K982255 · Model 2000 Pulse Oximeter · Ivy Biomedical Systems, Inc. · 1998K991823 · N-395 Pulse Oximeter, Model N-395 · Nellcor Puritan Bennett, Inc. · 1999×2: 2 generations cleared close together, K012891 (2002), K021090 (2002) K031740 · VSM - Vital Signs Monitor, Model 53000 Series · Welch Allyn Protocol, Inc. · 2003K033378 · Vital Signs Monitor, Propaq LT · Welch Allyn Protocol, Inc. · 2004 · recalledK101674 · WVSM (Wireless Vital Signs Monitor), Model 5.0 · Athena Gtx · 2010K113165 · Mini-Medic · Athena Gtx · 2012K190792 · BB-613WP · Biobeat Technologies, Ltd. · 201919801990200020102020
  1. Root · 11 generations backK962603 Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & CablesMasimo Corp. · 1997
  2. 7 generations between
  3. 3 generations backK101674 WVSM (Wireless Vital Signs Monitor), Model 5.0Athena Gtx · 2010
  4. 2 generations backK113165 Mini-MedicAthena Gtx · 2012
  5. 1 generation backK190792 BB-613WPBiobeat Technologies, Ltd. · 2019
  6. This deviceK250415 G0 Blood Pressure Monitoring System (G0)Aktiia SA · 2025
  7. Descendant · cites it directly · 0 of its ownK254000 Happy Health Blood Pressure MonitorHappy Health, Inc. · 2026
Show all 11 generations
  1. Root · 11 generations backK962603 Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & CablesMasimo Corp. · 1997
  2. 10 generations backK973887 Masimo Set MS-1P Pulse Oximeter and the Lnop Series of Sensors and…Masimo Corp. · 1998
  3. 9 generations backK982255 Model 2000 Pulse OximeterIvy Biomedical Systems, Inc. · 1998
  4. 8 generations backK991823 N-395 Pulse Oximeter, Model N-395Nellcor Puritan Bennett, Inc. · 1999
  5. 7 generations backK012891 Oximax Pulse Oximetry System with N-595 Pulse Oximeter and Oximax…Nellcor Puritan Bennett, Inc. · 2002 · Recalled
  6. 6 generations backK021090 Oximax N-550 Pulse Oximeter,Model N-550 Pulse OximeterNellcor Puritan Bennett, Inc. · 2002
  7. 5 generations backK031740 VSM - Vital Signs Monitor, Model 53000 SeriesWelch Allyn Protocol, Inc. · 2003
  8. 4 generations backK033378 Vital Signs Monitor, Propaq LTWelch Allyn Protocol, Inc. · 2004 · Recalled
  9. 3 generations backK101674 WVSM (Wireless Vital Signs Monitor), Model 5.0Athena Gtx · 2010
  10. 2 generations backK113165 Mini-MedicAthena Gtx · 2012
  11. 1 generation backK190792 BB-613WPBiobeat Technologies, Ltd. · 2019
  12. This deviceK250415 G0 Blood Pressure Monitoring System (G0)Aktiia SA · 2025

Most influential descendants

Showing 1 of 1 · See all
  • K254000Happy Health Blood Pressure MonitorHappy Health, Inc. · 2026 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−11K962603Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & CablesMasimo Corp.1997—
−10K973887Masimo Set MS-1P Pulse Oximeter and the Lnop Series of Sensors and CablesMasimo Corp.1998—
−9K982255Model 2000 Pulse OximeterIvy Biomedical Systems, Inc.1998—
−8K991823N-395 Pulse Oximeter, Model N-395Nellcor Puritan Bennett, Inc.1999—
−7K012891Oximax Pulse Oximetry System with N-595 Pulse Oximeter and Oximax Sensors and…Nellcor Puritan Bennett, Inc.20021
−6K021090Oximax N-550 Pulse Oximeter,Model N-550 Pulse OximeterNellcor Puritan Bennett, Inc.2002—
−5K031740VSM - Vital Signs Monitor, Model 53000 SeriesWelch Allyn Protocol, Inc.2003—
−4K033378Vital Signs Monitor, Propaq LTWelch Allyn Protocol, Inc.20043
−3K101674WVSM (Wireless Vital Signs Monitor), Model 5.0Athena Gtx2010—
−2K113165Mini-MedicAthena Gtx2012—
−1K190792BB-613WPBiobeat Technologies, Ltd.2019—
0 (this device)K250415G0 Blood Pressure Monitoring System (G0)Aktiia SA2025—
+1K254000Happy Health Blood Pressure MonitorHappy Health, Inc.2026—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198210
198610
198810
198910
199020
199120
199210
199330
199420
199520
199640
199710
199810
199920
200080
200130
200230
200340
200420
200550
200640
200730
201110
201230
201320
201410
201510
201610
201720
201820
202110
202310
202410

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202610
FIG. 1 — Lineage of K250415, 1997–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DXN? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DXN device

  1. #1 K222125 Arm Blood Pressure Monitor, models AOJ-30A…score 18.5 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K190792BB-613WP
Summary page 5
…Non-Invasive 870.1130 DXN I Legally Marketed Predicate Devices 21 CFR 807.92(a)(3) Predicate # Predicate Trade Name (Primary Predicate is listed first) Product Code K190792 BB-613 WP DXN K240254 Blood Pressure Monitor (TMB-2296-BT) DXN Device Description Summary 21 CFR 807.92(a)(4) The Aktiia GO Blo…
Biobeat Technologies, Ltd.2019—
*K240254Blood Pressure Monitor (TMB-2296-BT)
Summary page 5
…0 DXN I Legally Marketed Predicate Devices 21 CFR 807.92(a)(3) Predicate # Predicate Trade Name (Primary Predicate is listed first) Product Code K190792 BB-613 WP DXN K240254 Blood Pressure Monitor (TMB-2296-BT) DXN Device Description Summary 21 CFR 807.92(a)(4) The Aktiia GO Blood Pressure Monitoring…
Guangdong Transtek Medical Electronics Co., Ltd.2024—

Also cleared against K190792: 4 other devices →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K254000Happy Health Blood Pressure MonitorHappy Health, Inc.20260

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 28 years and passes through 7 companies; a new generation was cleared every 1.3 years on median.
  • 12 of the 83 ancestors have a recall of any class.
  • Of its descendants, K254000 (Happy Health, Inc., 2026) has gone furthest, with 0 descendants of its own.
  • 4 other devices also cite its predicate K190792.

How does its category compare?

All DXN devices

DXN System, Measurement, Blood-Pressure, Non-Invasive: FDA decisions, review times, recalls and adverse event reports over 20 years. 2025, this device’s year, is highlighted.

FDA decisions per year · DXN

28in 2025

2007: 272008: 232009: 462010: 432011: 342012: 362013: 342014: 292015: 302016: 302017: 332018: 222019: 332020: 242021: 232022: 252023: 352024: 172025: 282026 (partial year): 29
2007high 46 (2009) · low 17 (2024)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DXN

142in 2025

2007: 118 days median2008: 82 days median2009: 73 days median2010: 94 days median2011: 72 days median2012: 116 days median2013: 150 days median2014: 133 days median2015: 97 days median2016: 144 days median2017: 184 days median2018: 235 days median2019: 156 days median2020: 198 days median2021: 190 days median2022: 137 days median2023: 182 days median2024: 107 days median2025: 142 days median2026 (partial year): 147 days median
2007high 235 (2018) · low 72 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DXN

3in 2024

2008: 12009: 32011: 12013: 12016: 12019: 12024: 3
2007high 3 (2009) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DXN

239in 2025

2007: 212008: 82009: 172010: 142011: 102012: 192013: 722014: 472015: 622016: 542017: 902018: 6172019: 962020: 1302021: 1622022: 1552023: 1622024: 2702025: 239
2007high 617 (2018) · low 8 (2008)2025

MAUDE reports naming product code DXN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200727118021
2008238218
20094673317
20104394014
20113472110
201236116019
201334150172
201429133047
20153097062
201630144154
201733184090
2018222350617
201933156196
2020241980130
2021231900162
2022251370155
2023351820162
2024171073270
2025281420239
2026291470—

What is it indicated for?

The Aktiia GO Blood pressure monitoring system consists of a wrist worn monitor and an

oscillometric cuff. The system is intended for measuring blood pressure and pulse rate based on

using a Pulse Wave technique in adults aged from 22 to 59 years old with wrist circumference

ranging from 14 cm to 21 cm following a calibration process (every 24 hours) using the oscillometric

blood pressure cuff.

The Aktiia GO Blood pressure monitoring system is intended for spot-checking of adult patients for

home use.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K250415: G0 Blood Pressure Monitoring System (G0).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K250415

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026