510(k) Report

510(k) · Cleared

Cleared

BB-613-BPM

K241066510(k) TraditionalCleared Jan 14, 2025271-day reviewClass II21 CFR 870.1130

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Biobeat Technologies, Ltd. · DXN System, Measurement, Blood-Pressure, Non-Invasive

Cleared on Jan 14, 2025 as substantially equivalent to K190792 (Biobeat Technologies, Ltd., 2019) and 1 other predicate after a 271-day review. Its family tree reaches back 22 generations to 1997. 2 of the 179 devices in its family, counting itself, have a Class I recall.

Decided
Jan 14, 2025
Review
271 days
Class
Class II
Product code
DXN
Recalls
None
Predicates
2

Its longest line of descent runs back 22 generations to K962603 (1997). 178 ancestors in all, the oldest K822055 (1982). No later device cites it as a predicate yet. 2 of the 179 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year cleared 2019: 2 other ancestors2018: 3 other ancestors2017: 3 other ancestors2012: 5 other ancestors2016: 2 other ancestors, 1 with a Class I recall12015: 2 other ancestors2014: 4 other ancestors2010: 6 other ancestors2007: 4 other ancestors2006: 8 other ancestors2005: 12 other ancestors2001: 3 other ancestors2013: 2 other ancestors2011: 3 other ancestors2004: 9 other ancestors2003: 8 other ancestors2002: 10 other ancestors2000: 15 other ancestors1996: 4 other ancestors1990: 3 other ancestors1986: 2 other ancestors2008: 3 other ancestors1997: 5 other ancestors, 1 with a Class I recall11994: 2 other ancestors1993: 4 other ancestors1991: 5 other ancestors1988: 1 other ancestor1982: 1 other ancestor2009: 1 other ancestor1999: 4 other ancestors1995: 5 other ancestors1998: 7 other ancestors1992: 4 other ancestors1985: 1 other ancestor1989: 1 other ancestor1987: 2 other ancestorsK962603 · Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & Cables · Masimo Corp. · 1997K973887 · Masimo Set MS-1P Pulse Oximeter and the Lnop Series of Sensors and Cables · Masimo Corp. · 1998K982255 · Model 2000 Pulse Oximeter · Ivy Biomedical Systems, Inc. · 1998K991823 · N-395 Pulse Oximeter, Model N-395 · Nellcor Puritan Bennett, Inc. · 1999×2: 2 generations cleared close together, K012891 (2002), K021090 (2002) K033048 · Cas 740 Series Monitor, Models 740-1, 740M-1, 740-2MS, 740M-2MS, 740-2NL, 740M-2NL, 740-2NN, 740M-2NN, 740-3MS, 740M-3MS · Cas Medical Systems, Inc. · 2004 · recalledK050844 · 750C Series Monitor, Models 750C-2MS, 750CM-2MS, 750C-NNL, 750CM-2NL, 750C-3MS, 750CM-3MS, 750C-3NL, 750CM-3NL · Cas Medical Systems, Inc. · 2005K060065 · CAPNOSTREAM20 Two Parameter Bedside Monitor & CAPNOSTREAM10 One Parameter Bedside Monitor · Oridion Capnography, Inc. · 2006 · recalledK072295 · Capnostream 20 with A2 Adaptive Averaging Software · Oridion Capnography, Inc. · 2007K082268 · CAPNOSTREAM20 with Integrated Pulmonary Index · Oridion Capnography, Inc. · 2009 · recalledK091241 · Masimo Rainbow Set Radical 7 Pulse CO-Oximeter; Rad 87 Pulse CO-Oximeter · Masimo Corporation · 2009K100428 · Masimo Rainbow Set Radical 7R Pulse CO-Oximeter · Masimo Corporation · 2010×2: 2 generations cleared close together, K110516 (2011), K112687 (2011) K121013 · Connex Vital Signs Monitor 6000 Series · Welch Allyn, Inc. · 2012×2: 2 generations cleared close together, K140188 (2014), K142394 (2014) K150314 · Masimo MightySat Rx Fingertip Pulse Oximeter · Masimo Corporation · 2015K181956 · Masimo MightySat Rx Fingertip Pulse Oximeter · Masimo Corporation · 2019K212153 · Biobeat Platform, BB-613WP Patch · Biobeat Technologies, Ltd. · 2022K222010 · Biobeat Platform-2 and BB-613WP Patch · Biobeat Technologies, Ltd. · 202219801990200020102020
Other ancestors, by generation 1 generation back: 1 other ancestor2 generations back: 5 other ancestors3 generations back: 13 other ancestors, 1 with a Class I recall14 generations back: 20 other ancestors5 generations back: 22 other ancestors6 generations back: 18 other ancestors7 generations back: 20 other ancestors8 generations back: 16 other ancestors9 generations back: 13 other ancestors10 generations back: 15 other ancestors, 1 with a Class I recall111 generations back: 5 other ancestors12 generations back: 5 other ancestors13 generations back: 3 other ancestors K222010 · Biobeat Platform-2 and BB-613WP Patch · Biobeat Technologies, Ltd. · 2022K212153 · Biobeat Platform, BB-613WP Patch · Biobeat Technologies, Ltd. · 2022K181956 · Masimo MightySat Rx Fingertip Pulse Oximeter · Masimo Corporation · 2019K150314 · Masimo MightySat Rx Fingertip Pulse Oximeter · Masimo Corporation · 2015K142394 · Masimo Root Monitoring System and Accesories · Masimo Corporation · 2014K140188 · Masimo Root Monitoring System · Masimo Corporation · 2014K121013 · Connex Vital Signs Monitor 6000 Series · Welch Allyn, Inc. · 2012K112687 · Connex (R) Vital Signs Monitor 6000 Series · Welch Allyn, Inc. · 2011K110516 · Connex Vital Signs Monitor 6000 Series · Welch Allyn, Inc. · 2011K100428 · Masimo Rainbow Set Radical 7R Pulse CO-Oximeter · Masimo Corporation · 2010K091241 · Masimo Rainbow Set Radical 7 Pulse CO-Oximeter; Rad 87 Pulse CO-Oximeter · Masimo Corporation · 2009K082268 · CAPNOSTREAM20 with Integrated Pulmonary Index · Oridion Capnography, Inc. · 2009 · recalledK072295 · Capnostream 20 with A2 Adaptive Averaging Software · Oridion Capnography, Inc. · 2007K060065 · CAPNOSTREAM20 Two Parameter Bedside Monitor & CAPNOSTREAM10 One Parameter Bedside Monitor · Oridion Capnography, Inc. · 2006 · recalledK050844 · 750C Series Monitor, Models 750C-2MS, 750CM-2MS, 750C-NNL, 750CM-2NL, 750C-3MS, 750CM-3MS, 750C-3NL, 750CM-3NL · Cas Medical Systems, Inc. · 2005K033048 · Cas 740 Series Monitor, Models 740-1, 740M-1, 740-2MS, 740M-2MS, 740-2NL, 740M-2NL, 740-2NN, 740M-2NN, 740-3MS, 740M-3MS · Cas Medical Systems, Inc. · 2004 · recalledK021090 · Oximax N-550 Pulse Oximeter,Model N-550 Pulse Oximeter · Nellcor Puritan Bennett, Inc. · 2002K012891 · Oximax Pulse Oximetry System with N-595 Pulse Oximeter and Oximax Sensors and Cables (Aka Accessories) · Nellcor Puritan Bennett, Inc. · 2002 · recalledK991823 · N-395 Pulse Oximeter, Model N-395 · Nellcor Puritan Bennett, Inc. · 1999K982255 · Model 2000 Pulse Oximeter · Ivy Biomedical Systems, Inc. · 1998K973887 · Masimo Set MS-1P Pulse Oximeter and the Lnop Series of Sensors and Cables · Masimo Corp. · 1998K962603 · Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & Cables · Masimo Corp. · 1997−20−15−10−50
  1. Root · 22 generations backK962603 Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & CablesMasimo Corp. · 1997
  2. 18 generations between
  3. 3 generations backK181956 Masimo MightySat Rx Fingertip Pulse OximeterMasimo Corporation · 2019
  4. 2 generations backK212153 Biobeat Platform, BB-613WP PatchBiobeat Technologies, Ltd. · 2022
  5. 1 generation backK222010 Biobeat Platform-2 and BB-613WP PatchBiobeat Technologies, Ltd. · 2022
  6. This deviceK241066 BB-613-BPMBiobeat Technologies, Ltd. · 2025
Show all 22 generations
  1. Root · 22 generations backK962603 Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & CablesMasimo Corp. · 1997
  2. 21 generations backK973887 Masimo Set MS-1P Pulse Oximeter and the Lnop Series of Sensors and…Masimo Corp. · 1998
  3. 20 generations backK982255 Model 2000 Pulse OximeterIvy Biomedical Systems, Inc. · 1998
  4. 19 generations backK991823 N-395 Pulse Oximeter, Model N-395Nellcor Puritan Bennett, Inc. · 1999
  5. 18 generations backK012891 Oximax Pulse Oximetry System with N-595 Pulse Oximeter and Oximax…Nellcor Puritan Bennett, Inc. · 2002 · Recalled
  6. 17 generations backK021090 Oximax N-550 Pulse Oximeter,Model N-550 Pulse OximeterNellcor Puritan Bennett, Inc. · 2002
  7. 16 generations backK033048 Cas 740 Series Monitor, Models 740-1, 740M-1, 740-2MS, 740M-2MS…Cas Medical Systems, Inc. · 2004 · Recalled
  8. 15 generations backK050844 750C Series Monitor, Models 750C-2MS, 750CM-2MS, 750C-NNL…Cas Medical Systems, Inc. · 2005
  9. 14 generations backK060065 CAPNOSTREAM20 Two Parameter Bedside Monitor & CAPNOSTREAM10 One…Oridion Capnography, Inc. · 2006 · Recalled
  10. 13 generations backK072295 Capnostream 20 with A2 Adaptive Averaging SoftwareOridion Capnography, Inc. · 2007
  11. 12 generations backK082268 CAPNOSTREAM20 with Integrated Pulmonary IndexOridion Capnography, Inc. · 2009 · Recalled
  12. 11 generations backK091241 Masimo Rainbow Set Radical 7 Pulse CO-Oximeter; Rad 87 Pulse…Masimo Corporation · 2009
  13. 10 generations backK100428 Masimo Rainbow Set Radical 7R Pulse CO-OximeterMasimo Corporation · 2010
  14. 9 generations backK110516 Connex Vital Signs Monitor 6000 SeriesWelch Allyn, Inc. · 2011
  15. 8 generations backK112687 Connex (R) Vital Signs Monitor 6000 SeriesWelch Allyn, Inc. · 2011
  16. 7 generations backK121013 Connex Vital Signs Monitor 6000 SeriesWelch Allyn, Inc. · 2012
  17. 6 generations backK140188 Masimo Root Monitoring SystemMasimo Corporation · 2014
  18. 5 generations backK142394 Masimo Root Monitoring System and AccesoriesMasimo Corporation · 2014
  19. 4 generations backK150314 Masimo MightySat Rx Fingertip Pulse OximeterMasimo Corporation · 2015
  20. 3 generations backK181956 Masimo MightySat Rx Fingertip Pulse OximeterMasimo Corporation · 2019
  21. 2 generations backK212153 Biobeat Platform, BB-613WP PatchBiobeat Technologies, Ltd. · 2022
  22. 1 generation backK222010 Biobeat Platform-2 and BB-613WP PatchBiobeat Technologies, Ltd. · 2022
  23. This deviceK241066 BB-613-BPMBiobeat Technologies, Ltd. · 2025

Class I recalls in this family

View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−22K962603Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & CablesMasimo Corp.1997—
−21K973887Masimo Set MS-1P Pulse Oximeter and the Lnop Series of Sensors and CablesMasimo Corp.1998—
−20K982255Model 2000 Pulse OximeterIvy Biomedical Systems, Inc.1998—
−19K991823N-395 Pulse Oximeter, Model N-395Nellcor Puritan Bennett, Inc.1999—
−18K012891Oximax Pulse Oximetry System with N-595 Pulse Oximeter and Oximax Sensors and…Nellcor Puritan Bennett, Inc.20021
−17K021090Oximax N-550 Pulse Oximeter,Model N-550 Pulse OximeterNellcor Puritan Bennett, Inc.2002—
−16K033048Cas 740 Series Monitor, Models 740-1, 740M-1, 740-2MS, 740M-2MS, 740-2NL…Cas Medical Systems, Inc.20041
−15K050844750C Series Monitor, Models 750C-2MS, 750CM-2MS, 750C-NNL, 750CM-2NL…Cas Medical Systems, Inc.2005—
−14K060065CAPNOSTREAM20 Two Parameter Bedside Monitor & CAPNOSTREAM10 One Parameter…Oridion Capnography, Inc.20063
−13K072295Capnostream 20 with A2 Adaptive Averaging SoftwareOridion Capnography, Inc.2007—
−12K082268CAPNOSTREAM20 with Integrated Pulmonary IndexOridion Capnography, Inc.20099
−11K091241Masimo Rainbow Set Radical 7 Pulse CO-Oximeter; Rad 87 Pulse CO-OximeterMasimo Corporation2009—
−10K100428Masimo Rainbow Set Radical 7R Pulse CO-OximeterMasimo Corporation2010—
−9K110516Connex Vital Signs Monitor 6000 SeriesWelch Allyn, Inc.2011—
−8K112687Connex (R) Vital Signs Monitor 6000 SeriesWelch Allyn, Inc.2011—
−7K121013Connex Vital Signs Monitor 6000 SeriesWelch Allyn, Inc.2012—
−6K140188Masimo Root Monitoring SystemMasimo Corporation2014—
−5K142394Masimo Root Monitoring System and AccesoriesMasimo Corporation2014—
−4K150314Masimo MightySat Rx Fingertip Pulse OximeterMasimo Corporation2015—
−3K181956Masimo MightySat Rx Fingertip Pulse OximeterMasimo Corporation2019—
−2K212153Biobeat Platform, BB-613WP PatchBiobeat Technologies, Ltd.2022—
−1K222010Biobeat Platform-2 and BB-613WP PatchBiobeat Technologies, Ltd.2022—
0 (this device)K241066BB-613-BPMBiobeat Technologies, Ltd.2025—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198210
198510
198620
198720
198810
198910
199030
199150
199240
199340
199420
199550
199640
199751
199870
199940
2000150
200130
2002100
200380
200490
2005120
200680
200740
200830
200910
201060
201130
201250
201320
201440
201520
201621
201730
201830
201920
FIG. 1 — Lineage of K241066, 1997–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DXN device

  1. #1 K222125 Arm Blood Pressure Monitor, models AOJ-30A…score 18.5 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K190792BB-613WP
Summary page 5
…XG  870.2700 Oximeter, DQA  870.2910 Radiofrequency physiological signal transmitter and receiver, DRG Regulatory Class: Class II Predicate Devices BB-613 WP (K190792) Biobeat Platform-2 (K222010) System Description The Biobeat BB-613-BPM includes a sensor device that is attached t…
Biobeat Technologies, Ltd.2019—
*K222010Biobeat Platform-2 and BB-613WP Patch
Summary page 5
…10 Radiofrequency physiological signal transmitter and receiver, DRG Regulatory Class: Class II Predicate Devices BB-613 WP (K190792) Biobeat Platform-2 (K222010) System Description The Biobeat BB-613-BPM includes a sensor device that is attached to the patient’s chest to colle…
Biobeat Technologies, Ltd.2022—

Also cleared against K190792: 4 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (2 Class I)

No FDA recall names this device.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K160552EV1000 Clinical Platform with ClearSight Finger Cuff or ClearSight System, EV1000 Clinical PlatformEdwards Lifesciences, LLC · 2016 · 3 generations back · ◆ in the 2016 ancestors bar of FIG. 11 Class I recall
  • K970209Capnocheck PlusBci Intl., Inc. · 1997 · 10 generations back · ◆ in the 1997 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 28 years and passes through 8 companies; a new generation was cleared every 11 months on median.
  • 27 of the 178 ancestors have a recall of any class; 2 have a Class I recall.
  • 4 other devices also cite its predicate K190792.

How does its category compare?

All DXN devices

DXN System, Measurement, Blood-Pressure, Non-Invasive: FDA decisions, review times, recalls and adverse event reports over 20 years. 2025, this device’s year, is highlighted.

FDA decisions per year · DXN

28in 2025

2007: 272008: 232009: 462010: 432011: 342012: 362013: 342014: 292015: 302016: 302017: 332018: 222019: 332020: 242021: 232022: 252023: 352024: 172025: 282026 (partial year): 29
2007high 46 (2009) · low 17 (2024)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DXN

142in 2025

2007: 118 days median2008: 82 days median2009: 73 days median2010: 94 days median2011: 72 days median2012: 116 days median2013: 150 days median2014: 133 days median2015: 97 days median2016: 144 days median2017: 184 days median2018: 235 days median2019: 156 days median2020: 198 days median2021: 190 days median2022: 137 days median2023: 182 days median2024: 107 days median2025: 142 days median2026 (partial year): 147 days median
2007high 235 (2018) · low 72 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DXN

3in 2024

2008: 12009: 32011: 12013: 12016: 12019: 12024: 3
2007high 3 (2009) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DXN

239in 2025

2007: 212008: 82009: 172010: 142011: 102012: 192013: 722014: 472015: 622016: 542017: 902018: 6172019: 962020: 1302021: 1622022: 1552023: 1622024: 2702025: 239
2007high 617 (2018) · low 8 (2008)2025

MAUDE reports naming product code DXN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200727118021
2008238218
20094673317
20104394014
20113472110
201236116019
201334150172
201429133047
20153097062
201630144154
201733184090
2018222350617
201933156196
2020241980130
2021231900162
2022251370155
2023351820162
2024171073270
2025281420239
2026291470—

What is it indicated for?

The Biobeat BB-613-BPM is a noninvasive blood pressure monitor which provides a notification to the user every 15 minutes over a 24 hour period for user- initiated measurements. It includes a device and a platform (BPM) for use in adults by healthcare professionals, for collection of pulse rate and tracking changes in blood pressure based on Pulse Wave Transit Time (PWTT) in hospitals, clinics, long-term care and at home. The BB-613-BPM tracks changes in…

Read the full statement (124 words)

The Biobeat BB-613-BPM is a noninvasive blood pressure monitor which provides a notification to the user every 15 minutes over a 24 hour period for user- initiated measurements. It includes a device and a platform (BPM) for use in adults by healthcare professionals, for collection of pulse rate and tracking changes in blood pressure based on Pulse Wave Transit Time (PWTT) in hospitals, clinics, long-term care and at home. The BB-613-BPM tracks changes in blood pressure and requires calibration using an FDA-cleared oscillometric blood pressure monitor. The data from the Biobeat BB-613-BPM can be accessed through the Biobeat platform for review by a healthcare professional or can be downloaded as a report. The device is not intended to be used on critical care patients.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K241066: BB-613-BPM.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K241066

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026