Cdi (TM) System 100 Monitoring System

K902654510(k) TraditionalCleared Nov 5, 1990143-day reviewClass II21 CFR 870.4410Recalls 5

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

3M Health Care, Ltd. · DTY Sensor, Blood-Gas, In-Line, Cardiopulmonary Bypass

Cleared on Nov 5, 1990 after a 143-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 2 of the 15 devices in its family, counting itself, have a Class I recall. 14 later devices descend from it. It has been recalled 5 times.

Decided
Nov 5, 1990
Review
143 days
Class
Class II
Product code
DTY
Recalls
5
Predicates
0

14 later devices descend from it, 3 citing it directly. 2 of the 15 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 1997: 2 descendants cleared1998: 1 descendant cleared2012: 1 descendant cleared2009: 1 descendant cleared2011: 2 descendants cleared, 1 with a Class I recall12014: 2 descendants cleared, 1 with a Class I recall12018: 2 descendants cleared2013: 1 descendant cleared2020: 1 descendant cleared2021: 1 descendant cleared1990200020102020
  1. This deviceK902654 Cdi (TM) System 100 Monitoring System3M Health Care, Ltd. · 1990
  2. Descendant · cites it directly · 10 of its ownK972962 3M Cdi Blood Parameter Monitoring System 5003M Company · 1997 · Recalled
  3. Descendant · 2 generations on · 4 of its ownK110957 M4 MonitorSpectrum Medical, Llp · 2011
  4. Descendant · 3 generations on · 3 of its ownK173591 Quantum Diagnostic Module, Quantum Diagnostic Module - No GasSpectrum Medical, Ltd. · 2018
  5. Descendant · 2 generations on · 3 of its ownK090147 Blood Monitoring Unit, Model Bmu 40Maquet Cardiopulmonary, AG · 2009 · Recalled
  6. Descendant · 4 generations on · 2 of its ownK181942 Quantum Ventilation ModuleSpectrum Medical, Ltd. · 2018

Showing 5 of 14 descendants. See all descendants

Descendants cleared per year

1997: 21998: 12009: 12011: 22012: 12013: 12014: 22018: 22020: 12021: 1
1997high 2 (1997) · low 0 (1999)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 14 · See all
  • K9729623M Cdi Blood Parameter Monitoring System 5003M Company · 1997 · cites it directly10descendants
  • K110957M4 MonitorSpectrum Medical, Llp · 2011 · 2 generations on4descendants
  • K173591Quantum Diagnostic Module, Quantum Diagnostic Module - No GasSpectrum Medical, Ltd. · 2018 · 3 generations on3descendants
  • K090147Blood Monitoring Unit, Model Bmu 40Maquet Cardiopulmonary, AG · 2009 · 2 generations on3descendants
  • K181942Quantum Ventilation ModuleSpectrum Medical, Ltd. · 2018 · 4 generations on2descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K902654Cdi (TM) System 100 Monitoring System3M Health Care, Ltd.1990—
+1K9729623M Cdi Blood Parameter Monitoring System 5003M Company199726
+2K110957M4 MonitorSpectrum Medical, Llp2011—
+3K173591Quantum Diagnostic Module, Quantum Diagnostic Module - No GasSpectrum Medical, Ltd.2018—
+2K090147Blood Monitoring Unit, Model Bmu 40Maquet Cardiopulmonary, AG20091
+4K181942Quantum Ventilation ModuleSpectrum Medical, Ltd.2018—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
199720
199810
200910
201121
201210
201310
201421
201820
202010
202110
FIG. 1 — Lineage of K902654, 1990–2021Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it? (14)

See all 14

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K9729623M Cdi Blood Parameter Monitoring System 5003M Company199710
K9807323M Cdi Quick-Cells with Heparin Treatment3M Healthcare19981
K9734723M Cdi H/S Cuvettes with Heparon Treatment3M Healthcare19970

Has anything in its family been recalled? (2 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 28, 2011Not classifiedZ-2610-2011Device DesignTerumo discovered through its own testing that the presence of electromagnetic interference (EMI) in the operating room may affect the accuracy of the values displayed by the CDI"101 Hematocrit/Oxygen Saturation Monitoring System…
Apr 21, 2011Not classifiedZ-2684-2011Device DesignInaccurate readings may result after Methylene Blue, an intravascular dye, is administered to the patient. The presence of Methylene Blue may interfere with the readings from the shunt sensor for the CDI 500 system in a way that would…
Apr 21, 2011Not classifiedZ-2683-2011Device DesignInaccurate readings may result after Methylene Blue, an intravascular dye, is administered to the patient. The presence of Methylene Blue may interfere with the readings from the shunt sensor for the CDI 500 system in a way that would…
Nov 1, 2010Not classifiedZ-1943-2011Device DesignInaccurate result. Possibility for displayed values for potassium, CO2, hematocrit, O2 saturation, pH, hemoglobin and other measured blood parameters to be inaccurate.
Nov 1, 2010Not classifiedZ-1944-2011Device DesignInaccurate result. Possibility for displayed values for potassium, CO2, hematocrit, O2 saturation, pH, hemoglobin and other measured blood parameters to be inaccurate.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K102726Cardiohelp SystemMaquet Cardiopulmonary, AG · 2011 · 3 generations on · ◆ in the 2011 descendants bar of FIG. 11 Class I recall
  • K133598Cardiohelp SystemMaquet Cardiopulmonary, AG · 2014 · 4 generations on · ◆ in the 2014 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • Of its descendants, K972962 (3M Company, 1997) has gone furthest, with 10 descendants of its own.

How does its category compare?

All DTY devices

DTY Sensor, Blood-Gas, In-Line, Cardiopulmonary Bypass: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · DTY

1in 2010

2010: 1
2007high 1 (2010) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DTY

Too few DTY 510(k)s in any year since 2007 to give a median.

Recalls per year · DTY

2in 2010

2010: 2
2007high 2 (2010) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DTY

3in 2015

2010: 1302011: 1582012: 202013: 182014: 82015: 3
2007high 158 (2011) · low 0 (2007)2025

MAUDE reports naming product code DTY, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20101—2130
20110—0158
20120—020
20130—018
20140—08
20150—03
20160—00
20170—00
20180—00
20190—00
20200—00
20210—00
20220—00
20230—00
20240—00
20250—00
20260—0—
Cite this page

510(k) Report. “K902654: Cdi (TM) System 100 Monitoring System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K902654

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026