510(k) Report

510(k) · Cleared

Recalled · Class II · 2017

Medi-Therm (Mta-5900) & Detachable Hoses

K912051510(k) TraditionalCleared Jul 30, 199182-day reviewClass II21 CFR 870.5900Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Gaymar Industries, Inc. · DWJ System, Thermal Regulating

Cleared on Jul 30, 1991 after an 82-day review. FDA lists a 510(k) statement for it but has not posted the file, so its predicates are not known. 46 later devices descend from it. It has been recalled 2 times.

Decided
Jul 30, 1991
Review
82 days
Class
Class II
Product code
DWJ
Recalls
2
Predicates
0

46 later devices descend from it, 5 citing it directly.

Descendants, by year cleared 2003: 2 descendants cleared2007: 4 descendants cleared2011: 2 descendants cleared2006: 1 descendant cleared2010: 4 descendants cleared2014: 3 descendants cleared2009: 1 descendant cleared2016: 4 descendants cleared2012: 4 descendants cleared2005: 2 descendants cleared2008: 2 descendants cleared2004: 1 descendant cleared2023: 3 descendants cleared2025: 5 descendants cleared2017: 2 descendants cleared2018: 2 descendants cleared2020: 1 descendant cleared2015: 1 descendant cleared2021: 1 descendant cleared2022: 1 descendant cleared1990200020102020
  1. This deviceK912051 Medi-Therm (Mta-5900) & Detachable HosesGaymar Industries, Inc. · 1991
  2. Descendant · cites it directly · 20 of its ownK071341 Modification to Arctic Sun Temperature Management System, Model 2000Medivance, Inc. · 2007 · Recalled
  3. Descendant · cites it directly · 18 of its ownK061023 Thermosuit SystemLife Recovery Systems Hd, LLC · 2006 · Recalled
  4. Descendant · 2 generations on · 16 of its ownK083662 CriticoolMtre Advanced Technology, Ltd. · 2009 · Recalled
  5. Descendant · cites it directly · 16 of its ownK022366 Celsius Control SystemInnercool Therapies · 2003
  6. Descendant · 2 generations on · 15 of its ownK030421 Alsius Coolgard 3000 & Alsius Catheter SystemAlsius Corp. · 2003 · Recalled

Showing 5 of 46 descendants. See all descendants

Descendants cleared per year

2003: 22004: 12005: 22006: 12007: 42008: 22009: 12010: 42011: 22012: 42014: 32015: 12016: 42017: 22018: 22020: 12021: 12022: 12023: 32025: 5
2003high 5 (2025) · low 0 (2013)2026

Most influential descendants

Showing 5 of 46 · See all
  • K071341Modification to Arctic Sun Temperature Management System, Model 2000Medivance, Inc. · 2007 · cites it directly20descendants
  • K061023Thermosuit SystemLife Recovery Systems Hd, LLC · 2006 · cites it directly18descendants
  • K083662CriticoolMtre Advanced Technology, Ltd. · 2009 · 2 generations on16descendants
  • K022366Celsius Control SystemInnercool Therapies · 2003 · cites it directly16descendants
  • K030421Alsius Coolgard 3000 & Alsius Catheter SystemAlsius Corp. · 2003 · 2 generations on15descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K912051Medi-Therm (Mta-5900) & Detachable HosesGaymar Industries, Inc.1991—
+1K071341Modification to Arctic Sun Temperature Management System, Model 2000Medivance, Inc.20071
+1K061023Thermosuit SystemLife Recovery Systems Hd, LLC20061
+2K083662CriticoolMtre Advanced Technology, Ltd.20091
+1K022366Celsius Control SystemInnercool Therapies2003—
+2K030421Alsius Coolgard 3000 & Alsius Catheter SystemAlsius Corp.20034

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200320
200410
200520
200610
200740
200820
200910
201040
201120
201240
201430
201510
201640
201720
201820
202010
202110
202210
202330
202550
FIG. 1 — Lineage of K912051, 1991–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DWJ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DWJ device

  1. #1 K260278 ThermAffyx (TM) Patient Safety Systemscore 12.8 · 5 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA lists a 510(k) statement for it but has not posted the file, so its predicates are not known.

What descends from it? (46)

See all 46

5 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K071341Modification to Arctic Sun Temperature Management System, Model 2000Medivance, Inc.200720
K061023Thermosuit SystemLife Recovery Systems Hd, LLC200618
K022366Celsius Control SystemInnercool Therapies200316
K100585Medi-Therm Hyper/Hypothermia System, Models MTA6900, MTA7900Gaymar Industries, Inc.20119
K101589Blanketrol Iii, Model 233 and Coolblue, Model 25-01 Hyper-Hypothermia SystemsCincinnati Sub-Zero Products, Inc.20104

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 29, 2017Class IIZ-2058-2017Software designThe water may reheat too quickly if the unit is in "Automatic" mode, switched to "Manual" mode to view water temperature, and then switched back to "Automatic" mode.
Dec 27, 2013Class IIZ-0933-2014Device DesignStryker Medical has identified that on some units of the MTA7900, there is a lack of a water temperature display while the unit is in Auto mode. While in Auto mode, only the patient temperature is displayed along with the set point. The…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K071341 (Medivance, Inc., 2007) has gone furthest, with 20 descendants of its own.

How does its category compare?

All DWJ devices

DWJ System, Thermal Regulating: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · DWJ

3in 2025

2007: 42008: 32009: 32010: 82011: 42012: 52013: 52014: 42016: 22017: 52018: 32019: 22020: 52021: 22022: 22023: 62024: 22025: 32026 (partial year): 5
2007high 8 (2010) · low 0 (2015)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DWJ

131in 2025

2007: 106 days median2008: 73 days median2009: 106 days median2010: 134 days median2011: 180 days median2012: 69 days median2013: 77 days median2014: 131 days median2017: 153 days median2018: 95 days median2020: 30 days median2023: 42 days median2025: 131 days median2026 (partial year): 99 days median
2007high 180 (2011) · low 30 (2020)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DWJ

3in 2024

2009: 62010: 22011: 72012: 12013: 22015: 12017: 32018: 52019: 32020: 22021: 12022: 52024: 32026 (partial year): 3
2007high 7 (2011) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DWJ

2,384in 2025

2007: 522008: 402009: 712010: 1042011: 1362012: 1062013: 732014: 1162015: 2452016: 1,4002017: 1,4732018: 1,5172019: 2,2942020: 2,5462021: 2,6892022: 3,2262023: 3,4552024: 3,3292025: 2,384
2007high 3,455 (2023) · low 40 (2008)2025

MAUDE reports naming product code DWJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20074106052
2008373040
20093106671
201081342104
201141807136
20125691106
2013577273
201441310116
20150—1245
20162—01,400
2017515331,473
201839551,517
20192—32,294
202053022,546
20212—12,689
20222—53,226
202364203,455
20242—33,329
2025313102,384
20265993—
Cite this page

510(k) Report. “K912051: Medi-Therm (Mta-5900) & Detachable Hoses.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K912051

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026