510(k) Report

510(k) · Cleared

Recalled · Class II · 2015

Creatinine Meth for Use on the Dimension Clin Chem

K925668510(k) TraditionalCleared Dec 28, 199248-day reviewClass II21 CFR 862.1225Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

E.I. Dupont DE Nemours & Co., Inc. · CGX Alkaline Picrate, Colorimetry, Creatinine

Cleared on Dec 28, 1992 after a 48-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 3 later devices descend from it. It has been recalled 3 times.

Decided
Dec 28, 1992
Review
48 days
Class
Class II
Product code
CGX
Recalls
3
Predicates
0

3 later devices descend from it, 2 citing it directly.

Descendants, by year cleared 2006: 1 descendant cleared2014: 2 descendants cleared1990200020102020
  1. This deviceK925668 Creatinine Meth for Use on the Dimension Clin ChemE.I. Dupont DE Nemours & Co., Inc. · 1992
  2. Descendant · cites it directly · 1 of its ownK061238 Dimension Vista Carbon Dioxide, Creatinine and Lactate Dehydrogenase…Dade Behring, Inc. · 2006 · Recalled
  3. Descendant · cites it directly · 0 of its ownK132638 Dimension Creatinine (CRE2) Flex Reagent CartridgeSiemens Healthcare Diagnostics, Inc. · 2014 · Recalled
  4. Descendant · 2 generations on · 0 of its ownK133728 Dimension Vista Creatinine (CRE2) Flex Reagent CartridgeSiemens Healthcare Diagnostics, Inc. · 2014

Most influential descendants

Showing 3 of 3 · See all
  • K061238Dimension Vista Carbon Dioxide, Creatinine and Lactate Dehydrogenase Flex Reagent…Dade Behring, Inc. · 2006 · cites it directly1descendant
  • K132638Dimension Creatinine (CRE2) Flex Reagent CartridgeSiemens Healthcare Diagnostics, Inc. · 2014 · cites it directly0descendants
  • K133728Dimension Vista Creatinine (CRE2) Flex Reagent CartridgeSiemens Healthcare Diagnostics, Inc. · 2014 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K925668Creatinine Meth for Use on the Dimension Clin ChemE.I. Dupont DE Nemours & Co., Inc.1992—
+1K061238Dimension Vista Carbon Dioxide, Creatinine and Lactate Dehydrogenase Flex…Dade Behring, Inc.20061
+1K132638Dimension Creatinine (CRE2) Flex Reagent CartridgeSiemens Healthcare Diagnostics, Inc.20141
+2K133728Dimension Vista Creatinine (CRE2) Flex Reagent CartridgeSiemens Healthcare Diagnostics, Inc.2014—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200610
201420
FIG. 1 — Lineage of K925668, 1992–2014Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in CGX? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CGX device

  1. #1 K242685 Atellica® CH Creatinine_3 (Crea3)score 12.1 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it? (3)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K061238Dimension Vista Carbon Dioxide, Creatinine and Lactate Dehydrogenase Flex Reagent…Dade Behring, Inc.20061
K132638Dimension Creatinine (CRE2) Flex Reagent CartridgeSiemens Healthcare Diagnostics, Inc.20140

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jul 15, 2015Class IIZ-2741-2015Device DesignSiemens has confirmed that the Dimension CREA and Dimension Vista CREA may produce falsely depressed results up to - 12.7% and -22% respectively, for a creatinine concentration of 1.5 mg/dL when hemolysis exceeds 200 mg/dL and 300 mg/dL…
Nov 22, 2011Not classifiedZ-0772-2012Under Investigation by firmSiemens has confirmed the potential for falsely elevated creatinine results with CREA FLex(R) reagent cartridge lots GA2196, GA2203, DC2221, DB2221, GA2229 and FA2237 when processing tests from a well that has been punctured.
Dec 27, 2010Not classifiedZ-1729-2011Material/Component ContaminationFirm has confirmed the potential for falsely elevated creatinine results with specified CREA Flex(R) reagent cartridge lots when processing tests from a well that has been punctured. This issue may affect creatinine results for quality…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K061238 (Dade Behring, Inc., 2006) has gone furthest, with 1 descendant of its own.

How does its category compare?

All CGX devices

CGX Alkaline Picrate, Colorimetry, Creatinine: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · CGX

1in 2025

2007: 42009: 12012: 12013: 12014: 22016: 12022: 12024: 22025: 12026 (partial year): 1
2007high 4 (2007) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CGX

25in 2007

2007: 25 days median
2007high 25 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CGX

2in 2016

2008: 12009: 22010: 22011: 12012: 22016: 22026 (partial year): 1
2007high 2 (2009) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · CGX

34in 2025

2007: 152008: 272009: 52010: 52011: 322012: 132013: 242014: 132015: 132016: 32017: 212018: 142019: 252020: 192021: 132022: 232023: 192024: 322025: 34
2007high 34 (2025) · low 3 (2016)2025

MAUDE reports naming product code CGX, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007425015
20080—127
20091—25
20100—25
20110—132
20121—213
20131—024
20142—013
20150—013
20161—23
20170—021
20180—014
20190—025
20200—019
20210—013
20221—023
20230—019
20242—032
20251—034
20261—1—
Cite this page

510(k) Report. “K925668: Creatinine Meth for Use on the Dimension Clin Chem.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K925668

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026