510(k) Report

510(k) · Cleared

Recalled · Class II · 2017

Legacy/Legacy-D Table

K973039510(k) TraditionalCleared Oct 16, 199763-day reviewClass II21 CFR 892.1980Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

GE Medical Systems · IXR Table, Radiographic, Tilting

Cleared on Oct 16, 1997 after a 63-day review. The FDA summary names no predicate by number. 5 later devices descend from it. It has been recalled 4 times.

Decided
Oct 16, 1997
Review
63 days
Class
Class II
Product code
IXR
Recalls
4
Predicates
0

5 later devices descend from it, 1 citing it directly.

Descendants, by year cleared 2001: 1 descendant cleared2008: 1 descendant cleared2011: 1 descendant cleared2014: 1 descendant cleared2015: 1 descendant cleared1995200020052010201520202025
  1. This deviceK973039 Legacy/Legacy-D TableGE Medical Systems · 1997
  2. Descendant · cites it directly · 4 of its ownK011624 Expedio 500D R&F X-Ray SystemGE Medical Systems · 2001 · Recalled
  3. Descendant · 2 generations on · 3 of its ownK081091 Precision 500D R&F X-Ray SystemGE Healthcare · 2008 · Recalled
  4. Descendant · 3 generations on · 2 of its ownK110785 Kalare Diagnostic X-Ray Table SystemToshiba Medical Systems Coporation, Japan · 2011 · Recalled
  5. Descendant · 4 generations on · 1 of its ownK133553 KalareToshiba America Medical Systems, In.C · 2014 · Recalled
  6. Descendant · 5 generations on · 0 of its ownK143281 Ultimax-iToshibamedical Systems Corporation · 2015

Most influential descendants

Showing 5 of 5 · See all
  • K011624Expedio 500D R&F X-Ray SystemGE Medical Systems · 2001 · cites it directly4descendants
  • K081091Precision 500D R&F X-Ray SystemGE Healthcare · 2008 · 2 generations on3descendants
  • K110785Kalare Diagnostic X-Ray Table SystemToshiba Medical Systems Coporation, Japan · 2011 · 3 generations on2descendants
  • K133553KalareToshiba America Medical Systems, In.C · 2014 · 4 generations on1descendant
  • K143281Ultimax-iToshibamedical Systems Corporation · 2015 · 5 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K973039Legacy/Legacy-D TableGE Medical Systems1997—
+1K011624Expedio 500D R&F X-Ray SystemGE Medical Systems200111
+2K081091Precision 500D R&F X-Ray SystemGE Healthcare20089
+3K110785Kalare Diagnostic X-Ray Table SystemToshiba Medical Systems Coporation, Japan20111
+4K133553KalareToshiba America Medical Systems, In.C20143
+5K143281Ultimax-iToshibamedical Systems Corporation2015—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200110
200810
201110
201410
201510
FIG. 1 — Lineage of K973039, 1997–2015Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in IXR? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against?

The FDA summary names no predicate by number.

What descends from it? (5)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K011624Expedio 500D R&F X-Ray SystemGE Medical Systems20014

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 6, 2017Class IIZ-2179-2017Device DesignReported incidents of a patient step detaching from the table. A fall from a patient step detaching while in use could result in an injury to a patient or operator.
Jun 12, 2015Class IIZ-1966-2015Nonconforming Material/ComponentThe SFD/IDD safety mechanism may not engage properly at the lower range of SFD/IDD travel. This mechanism is installed to hold and prevent the SFD/IDD from falling due to a counterweight cable failure. The lower range of travel is 1.1 in…
Aug 30, 2013Class IIZ-0012-2014Employee errorGE Healthcare has recently become aware of a potential safety issue involving the spotfilm device power assist handle. There is a potential for the internal bolts on the power assist handle to be loose or to loosen and/or fall out. This…
Jan 2, 2008Not classifiedZ-1611-2008Nonconforming Material/ComponentGE Healthcare was notified that 2 recently installed Precision 500D systems were not in compliance with portions of the Performance Standard for Diagnostic X-ray Equipment. X-ray production was possible from the fluoroscopic X-ray tube…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K011624 (GE Medical Systems, 2001) has gone furthest, with 4 descendants of its own.

How does its category compare?

All IXR devices

IXR Table, Radiographic, Tilting: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · IXR

FDA has decided no IXR device since 2007.

Median 510(k) review, days · IXR

Too few IXR 510(k)s in any year since 2007 to give a median.

Recalls per year · IXR

1in 2025

2007: 12008: 42010: 52016: 12017: 12019: 12025: 1
2007high 5 (2010) · low 0 (2009)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · IXR

1in 2024

2007: 42008: 82009: 12010: 122011: 102012: 1642013: 1962014: 1202015: 822016: 582017: 252018: 172019: 82020: 122021: 52022: 22023: 12024: 1
2007high 196 (2013) · low 0 (2025)2025

MAUDE reports naming product code IXR, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—14
20080—48
20090—01
20100—512
20110—010
20120—0164
20130—0196
20140—0120
20150—082
20160—158
20170—125
20180—017
20190—18
20200—012
20210—05
20220—02
20230—01
20240—01
20250—10
20260—0—

What is it indicated for?

This table is intended for use in a diagnostic x-ray system to support patients during general purpose radiological procedures in the horizontal, vertical and trendelenburg positions.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K973039: Legacy/Legacy-D Table.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K973039

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026