Recalled · 2010Synchon CX Delta Clin Syst (CX4, CX5, CX7 Delta)
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Beckman Instruments, Inc. · JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Cleared on May 8, 1995 after an 82-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 19 later devices descend from it. It has been recalled 4 times.
- Decided
- May 8, 1995
- Review
- 82 days
- Class
- Class I
- Product code
- JJE
- Recalls
- 4
- Predicates
- 0
19 later devices descend from it, 3 citing it directly.
- This deviceK950958 Synchon CX Delta Clin Syst (CX4, CX5, CX7 Delta)Beckman Instruments, Inc. · 1995
- Descendant · cites it directly · 17 of its ownK965240 Synchron LX 20 Clinical Chemistry System, Synchron LX Albumin…Beckman Instruments, Inc. · 1997 · Recalled
- Descendant · 2 generations on · 11 of its ownK042291 Unicel DXC 600 and 800 Synchron SystemsBeckman Coulter, Inc. · 2004 · Recalled
- Descendant · 3 generations on · 5 of its ownK183688 i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc. · 2020
- Descendant · 3 generations on · 4 of its ownK183678 i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc. · 2020
- Descendant · 3 generations on · 4 of its ownK191298 i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc. · 2020
Showing 5 of 19 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 19 · See all- K965240Synchron LX 20 Clinical Chemistry System, Synchron LX Albumin Reagent, Synchron LX…Beckman Instruments, Inc. · 1997 · cites it directly17descendants
- K042291Unicel DXC 600 and 800 Synchron SystemsBeckman Coulter, Inc. · 2004 · 2 generations on11descendants
- K183688i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc. · 2020 · 3 generations on5descendants
- K183678i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc. · 2020 · 3 generations on4descendants
- K191298i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc. · 2020 · 3 generations on4descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| 0 (this device) | K950958 | Synchon CX Delta Clin Syst (CX4, CX5, CX7 Delta) | Beckman Instruments, Inc. | 1995 | — |
| +1 | K965240 | Synchron LX 20 Clinical Chemistry System, Synchron LX Albumin Reagent… | Beckman Instruments, Inc. | 1997 | 12 |
| +2 | K042291 | Unicel DXC 600 and 800 Synchron Systems | Beckman Coulter, Inc. | 2004 | 34 |
| +3 | K183688 | i-STAT CHEM8+ cartridge with the i-STAT 1 System | Abbott Point of Care, Inc. | 2020 | — |
| +3 | K183678 | i-STAT CHEM8+ cartridge with the i-STAT 1 System | Abbott Point of Care, Inc. | 2020 | — |
| +3 | K191298 | i-STAT CHEM8+ cartridge with the i-STAT 1 System | Abbott Point of Care, Inc. | 2020 | — |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1997 | 3 | 0 |
| 2001 | 1 | 0 |
| 2004 | 2 | 0 |
| 2006 | 1 | 0 |
| 2007 | 3 | 0 |
| 2011 | 1 | 0 |
| 2020 | 3 | 0 |
| 2021 | 1 | 0 |
| 2023 | 3 | 0 |
| 2026 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in JJE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
- Candidate predicates for a device like this, rankedin the report
- Recall history across each candidate’s familyin the report
What was it cleared against?
FDA has not posted a summary or statement for it, so its predicates are not known.
What descends from it? (19)
See all 193 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Aug 23, 2010 | Not classified | Z-0128-2012 | Software design | The recall was initiated because Beckman Coulter has confirmed software limitations which affect only manually-loaded reagent cartridges on the CX4/5/7 CE, CX4/5/7 Delta and CX9 ALX Clinical Systems. |
| Aug 23, 2010 | Not classified | Z-0125-2012 | Software design | The recall was initiated because Beckman Coulter has confirmed software limitations which affect only manually-loaded reagent cartridges on the CX4/5/7 CE, CX4/5/7 Delta and CX9 ALX Clinical Systems. |
| Aug 23, 2010 | Not classified | Z-0131-2012 | Software design | The recall was initiated because Beckman Coulter has confirmed software limitations which affect only manually-loaded reagent cartridges on the CX4/5/7 CE, CX4/5/7 Delta and CX9 ALX Clinical Systems. |
| Aug 23, 2010 | Not classified | Z-0132-2012 | Software design | The recall was initiated because Beckman Coulter has confirmed software limitations which affect only manually-loaded reagent cartridges on the CX4/5/7 CE, CX4/5/7 Delta and CX9 ALX Clinical Systems. |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- Of its descendants, K965240 (Beckman Instruments, Inc., 1997) has gone furthest, with 17 descendants of its own.
How does its category compare?
All JJE devicesJJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · JJE
1in 2019
2026 to Oct 25, outlined.
Median 510(k) review, days · JJE
224in 2014
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · JJE
6in 2025
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · JJE
4,400in 2025
MAUDE reports naming product code JJE, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 0 | — | 20 | 2,181 |
| 2008 | 1 | — | 20 | 2,159 |
| 2009 | 0 | — | 14 | 3,440 |
| 2010 | 0 | — | 32 | 3,273 |
| 2011 | 2 | — | 25 | 11,606 |
| 2012 | 1 | — | 16 | 3,707 |
| 2013 | 1 | — | 21 | 2,489 |
| 2014 | 3 | 224 | 14 | 2,841 |
| 2015 | 0 | — | 15 | 2,793 |
| 2016 | 0 | — | 9 | 2,580 |
| 2017 | 0 | — | 18 | 1,869 |
| 2018 | 0 | — | 17 | 4,590 |
| 2019 | 1 | — | 7 | 3,948 |
| 2020 | 0 | — | 7 | 3,986 |
| 2021 | 0 | — | 6 | 3,920 |
| 2022 | 0 | — | 0 | 3,998 |
| 2023 | 0 | — | 12 | 3,783 |
| 2024 | 0 | — | 6 | 4,221 |
| 2025 | 0 | — | 6 | 4,400 |
| 2026 | 0 | — | 5 | — |
https://510kreport.com/device/K950958 · printed Nov 4, 2026
Cite this page
510(k) Report. “K950958: Synchon CX Delta Clin Syst (CX4, CX5, CX7 Delta).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K950958
Report an extraction error
Are you the manufacturer?