510(k) Report

510(k) · Cleared

Recalled · Class II · 2020

Lifepak 500

K955854510(k) TraditionalCleared Nov 4, 1996312-day reviewClass III21 CFR 870.5310Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Physio-Control Corp. · MKJ Automated External Defibrillators (Non-Wearable)

Cleared on Nov 4, 1996 as substantially equivalent to K943301 (Physio-Control Corp., 1995) and 1 other predicate after a 312-day review. Its family tree reaches back 1 generation to 1995. 4 of the 30 devices in its family, counting itself, have a Class I recall. 27 later devices descend from it. It has been recalled 2 times.

Decided
Nov 4, 1996
Review
312 days
Class
Class III
Product code
MKJ
Recalls
2
Predicates
2

Its longest line of descent runs back 1 generation to K943301 (1995). 2 ancestors in all. 27 later devices descend from it, 4 citing it directly. 4 of the 30 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 1995: 1 other ancestor2001: 1 descendant cleared1999: 2 descendants cleared2006: 3 descendants cleared, 1 with a Class I recall12012: 2 descendants cleared, 1 with a Class I recall12003: 1 descendant cleared2008: 2 descendants cleared2002: 2 descendants cleared2005: 1 descendant cleared2013: 2 descendants cleared, 1 with a Class I recall12009: 1 descendant cleared2010: 1 descendant cleared2004: 2 descendants cleared2007: 2 descendants cleared2014: 1 descendant cleared, 1 with a Class I recall12011: 1 descendant cleared2015: 1 descendant cleared2024: 1 descendant cleared2025: 1 descendant clearedK943301 · Quik Combo · Physio-Control Corp. · 19951995200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor 1 generation forward: 4 descendants2 generations forward: 6 descendants, 1 with a Class I recall13 generations forward: 12 descendants, 2 with a Class I recall24 generations forward: 3 descendants, 1 with a Class I recall15 generations forward: 1 descendant6 generations forward: 1 descendant K943301 · Quik Combo · Physio-Control Corp. · 19950+5
  1. Root · 1 generation backK943301 Quik ComboPhysio-Control Corp. · 1995
  2. This deviceK955854 Lifepak 500Physio-Control Corp. · 1996
  3. Descendant · cites it directly · 25 of its ownK983393 Lifepak 500 BiphasicPhysio-Control Corp. · 1999 · Recalled
  4. Descendant · 2 generations on · 22 of its ownK991910 Lifepak 12 BiphasicPhysio-Control Corp. · 1999 · Recalled
  5. Descendant · 3 generations on · 21 of its ownK012274 Medtronic Physio-Control Lifepak 20 Defibrillator/Monitor/PacemakerMedtronic Physio-Control Corp. · 2002 · Recalled
  6. Descendant · cites it directly · 21 of its ownK012428 Modification to: Lifepak 500 Automated External DefibrillatorMedtronic Physio-Control Corp. · 2001 · Recalled
  7. Descendant · 2 generations on · 20 of its ownK033275 Lifepak 12, 20, 500, CR PlusMedtronic Physio-Control Corp. · 2003 · Recalled

Showing 5 of 27 descendants. See all descendants

Descendants cleared per year

1999: 22001: 12002: 22003: 12004: 22005: 12006: 32007: 22008: 22009: 12010: 12011: 12012: 22013: 22014: 12015: 12024: 12025: 1
1999high 3 (2006) · low 0 (2000)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 27 · See all
  • K983393Lifepak 500 BiphasicPhysio-Control Corp. · 1999 · cites it directly25descendants
  • K991910Lifepak 12 BiphasicPhysio-Control Corp. · 1999 · 2 generations on22descendants
  • K012274Medtronic Physio-Control Lifepak 20 Defibrillator/Monitor/PacemakerMedtronic Physio-Control Corp. · 2002 · 3 generations on21descendants
  • K012428Modification to: Lifepak 500 Automated External DefibrillatorMedtronic Physio-Control Corp. · 2001 · cites it directly21descendants
  • K033275Lifepak 12, 20, 500, CR PlusMedtronic Physio-Control Corp. · 2003 · 2 generations on20descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K943301Quik ComboPhysio-Control Corp.1995—
0 (this device)K955854Lifepak 500Physio-Control Corp.1996—
+1K983393Lifepak 500 BiphasicPhysio-Control Corp.19991
+2K991910Lifepak 12 BiphasicPhysio-Control Corp.19993
+3K012274Medtronic Physio-Control Lifepak 20 Defibrillator/Monitor/PacemakerMedtronic Physio-Control Corp.200217
+1K012428Modification to: Lifepak 500 Automated External DefibrillatorMedtronic Physio-Control Corp.20012
+2K033275Lifepak 12, 20, 500, CR PlusMedtronic Physio-Control Corp.20031

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199510

Descendants by year cleared

Year clearedDevicesWith a Class I recall
199920
200110
200220
200310
200420
200510
200631
200720
200820
200910
201010
201110
201221
201321
201411
201510
202410
202510
FIG. 1 — Lineage of K955854, 1995–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

4 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K943301Quik Combo
Summary page 1
…ubstantial Equivalence: This defibrillator is substantially equivalent to the currently marketed Physio-Control LIFEPAK 300 automated external defibrillator (510(k) no. K943301) and the SurVIVAlink automated external defibrillator (510(k) no. K940445). Description: The LIFEPAK 500 Automated External Def…
Physio-Control Corp.1995—
*K940445Vivalink Aed Defibrillator
Summary page 1
…the currently marketed Physio-Control LIFEPAK 300 automated external defibrillator (510(k) no. K943301) and the SurVIVAlink automated external defibrillator (510(k) no. K940445). Description: The LIFEPAK 500 Automated External Defibrillator (AED) is a portable battery powered device which applies a brief…
Survivalink Corp.1995—

What descends from it? (27)

See all 27

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K983393Lifepak 500 BiphasicPhysio-Control Corp.199925
K012428Modification to: Lifepak 500 Automated External DefibrillatorMedtronic Physio-Control Corp.200121
K052057Lifepak 500 Automated External DefibrillatorMedtronic Emergency Response Systems, Inc.20064
K120350Extended Life BatteryAed Battery Exchange, LLC20120

Has anything in its family been recalled? (4 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 20, 2020Class IIZ-1633-2020Device DesignAutomated External Defibrillators may not detect a patient connection due to mechanical wear-through of the contact plating on the therapy connector. The wear-through could result in the device not recognizing a patient is connected which…
Jun 13, 2007Not classifiedZ-1148-2007OtherReduced Shock- the device may deliver 100 Joule (J) to a patient, which is less than the recommended factory minimum default setting of 200 J.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K052465Reduced Energy Pediatric Electrode Pads for the Samaritan Aed and Pad, Models Sde 251, SPP 351Heartsine Technologies, Inc. · 2006 · 2 generations on · ◆ in the 2006 descendants bar of FIG. 11 Class I recall
  • K130454Lifepak 20 Defibrillator/Monitor, Lifepak 20E Defibrillator/MonitorPhysio-Control, Inc. · 2013 · 3 generations on · ◆ in the 2013 descendants bar of FIG. 11 Class I recall
  • K122600Lifepak 1000 DefibrillatorPhysio-Control, Inc. · 2012 · 3 generations on · ◆ in the 2012 descendants bar of FIG. 11 Class I recall
  • K142430LIFEPAK 15 monitor/defibrillatorPhysio-Control, Inc. · 2014 · 4 generations on · ◆ in the 2014 descendants bar of FIG. 13 Class I recalls

What does this lineage say?

  • None of the 2 ancestors has a recall.
  • Of its descendants, K983393 (Physio-Control Corp., 1999) has gone furthest, with 25 descendants of its own.

How does its category compare?

All MKJ devices

MKJ Automated External Defibrillators (Non-Wearable): FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · MKJ

48in 2025

2007: 72008: 132009: 82010: 42011: 112012: 102013: 72014: 112015: 62017: 42018: 202019: 252020: 522021: 412022: 192023: 242024: 272025: 482026 (partial year): 31
2007high 52 (2020) · low 0 (2016)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MKJ

156in 2015

2007: 69 days median2008: 141 days median2009: 165 days median2010: 250 days median2011: 36 days median2012: 197 days median2013: 191 days median2014: 196 days median2015: 156 days median
2007high 250 (2010) · low 36 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MKJ

1in 2025

2007: 122008: 242009: 372010: 352011: 72012: 5, 1 Class I2013: 42014: 24, 15 Class I2015: 12016: 2, 1 Class I2017: 6, 4 Class I2018: 32019: 6, 2 Class I2020: 22021: 42022: 32023: 112024: 52025: 12026 (partial year): 3
2007high 37 (2009) · low 1 (2015)2026

Class I (23 since 2007) Class II/III not classified

Adverse event reports per year · MKJ

13,315in 2025

2007: 3,9582008: 4,7522009: 6,7822010: 7,3992011: 8,1702012: 9,7322013: 14,9552014: 15,3992015: 16,1292016: 17,0622017: 16,6842018: 16,5912019: 16,8612020: 16,1972021: 16,3002022: 14,8672023: 13,7662024: 15,1092025: 13,315
2007high 17,062 (2016) · low 3,958 (2007)2025

MAUDE reports naming product code MKJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007769123,958
200813141244,752
20098165376,782
20104250357,399
2011113678,170
2012101975 (1)9,732
20137191414,955
20141119624 (15)15,399
20156156116,129
20160—2 (1)17,062
20174—6 (4)16,684
201820—316,591
201925—6 (2)16,861
202052—216,197
202141—416,300
202219—314,867
202324—1113,766
202427—515,109
202548—113,315
202631—3—

What is it indicated for?

The LIFEPAK 500 Automated External Defibrillator may be used in the hospital or pre-hospital setting by authorized personnel to terminate certain potentially fatal cardiac arrhythmias.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K955854: Lifepak 500.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K955854

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026