510(k) Report

510(k) · Cleared

Recalled · 2011

Sams 8000 Cardioplegia Monitor

K960916510(k) TraditionalCleared Jun 4, 199690-day reviewClass II21 CFR 870.1100Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

3M Health Care, Sarns · DSJ Alarm, Blood-Pressure

Cleared on Jun 4, 1996 as substantially equivalent to K915183 (3M Health Care, Sarns, 1992) and 1 other predicate after a 90-day review. Its family tree reaches back 1 generation to 1987. It has been recalled once.

Decided
Jun 4, 1996
Review
90 days
Class
Class II
Product code
DSJ
Recalls
1
Predicates
2

Its longest line of descent runs back 1 generation to K871131 (1987). 2 ancestors in all. No later device cites it as a predicate yet.

Other ancestors, by year cleared 1992: 1 other ancestorK871131 · Sarns 9000 Heart-Lung Console · 3M Health Care, Sarns · 1987 · recalled1986198819901992199419961998
Other ancestors, by generation 1 generation back: 1 other ancestor K871131 · Sarns 9000 Heart-Lung Console · 3M Health Care, Sarns · 1987 · recalled0
  1. Root · 1 generation backK871131 Sarns 9000 Heart-Lung Console3M Health Care, Sarns · 1987 · Recalled
  2. This deviceK960916 Sams 8000 Cardioplegia Monitor3M Health Care, Sarns · 1996
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K871131Sarns 9000 Heart-Lung Console3M Health Care, Sarns198714
0 (this device)K960916Sams 8000 Cardioplegia Monitor3M Health Care, Sarns1996—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199210
FIG. 1 — Lineage of K960916, 1987–1996Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DSJ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K915183Sarns 8000 Modular Perfusion System
Summary page 1
…blood pump" (21 CFR 870.4370). Predicate Devices: The new Sarns 8000 Cardioplegia Monitor is substantially equivalent to the existing Sarns 8000 Cardioplegia Monitor (K915183) and the Sarns 9000 Cardioplegia Control (K871131). Device Description: The Sarns 8000 Cardioplegia Monitor is only compatible…
3M Health Care, Sarns199245
*K871131Sarns 9000 Heart-Lung Console
Summary page 1
…: The new Sarns 8000 Cardioplegia Monitor is substantially equivalent to the existing Sarns 8000 Cardioplegia Monitor (K915183) and the Sarns 9000 Cardioplegia Control (K871131). Device Description: The Sarns 8000 Cardioplegia Monitor is only compatible with the Sarns 8000 Modular Perfusion System. The…
3M Health Care, Sarns198714

Also cleared against K915183: 2 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 29, 2011Not classifiedZ-3214-2011Component change controlThe Cardioplegia Monitor for the Sarns Modular Perfusion System 8000 may fail to track cardioplegia delivery volume.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • 2 of the 2 ancestors have a recall of any class.
  • 2 other devices also cite its predicate K915183.

How does its category compare?

All DSJ devices

DSJ Alarm, Blood-Pressure: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · DSJ

1in 2019

2014: 12015: 12017: 12019: 1
2007high 1 (2014) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DSJ

Too few DSJ 510(k)s in any year since 2007 to give a median.

Recalls per year · DSJ

No recall names DSJ since 2007.

Adverse event reports per year · DSJ

6in 2025

2015: 22017: 12019: 22021: 322022: 272023: 162024: 302025: 6
2007high 32 (2021) · low 0 (2007)2025

MAUDE reports naming product code DSJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20100—00
20110—00
20120—00
20130—00
20141—00
20151—02
20160—00
20171—01
20180—00
20191—02
20200—00
20210—032
20220—027
20230—016
20240—030
20250—06
20260—0—

What is it indicated for?

The Sarns 8000 Cardioplegia Monitor is indicated to measure extracorporeal line pressure and signal when the pressure exceeds the operator-set limit, either providing an alert or triggering an alarm which stops the cardioplegia pump only. The monitor also measures temperatures of the patient or extracorporeal circuit, counts the time during and between cardioplegia doses, and tracks the volume of cardioplegia delivered.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K960916: Sams 8000 Cardioplegia Monitor.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K960916

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026