510(k) Report

510(k) · Cleared

Cleared

Stat IM 5000 Cassette Autoclave

K962179510(k) TraditionalCleared Jan 13, 1997221-day reviewClass II21 CFR 880.6880

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Scican · FLE Sterilizer, Steam

Cleared on Jan 13, 1997 as substantially equivalent to K915054 (Scican, 1992) after a 221-day review. Its family tree reaches back 1 generation to 1992. 2 later devices descend from it.

Decided
Jan 13, 1997
Review
221 days
Class
Class II
Product code
FLE
Recalls
None
Predicates
1

Its longest line of descent runs back 1 generation to K915054 (1992). 2 later devices descend from it, 1 citing it directly.

Descendants, by year cleared 2007: 1 descendant cleared2009: 1 descendant clearedK915054 · Statim Cassett Autoclave, Modification · Scican · 19921990200020102020
Descendants, by generation 1 generation forward: 1 descendant2 generations forward: 1 descendant K915054 · Statim Cassett Autoclave, Modification · Scican · 19920
  1. Root · 1 generation backK915054 Statim Cassett Autoclave, ModificationScican · 1992
  2. This deviceK962179 Stat IM 5000 Cassette AutoclaveScican · 1997
  3. Descendant · cites it directly · 1 of its ownK072466 Statim 7000 (Without Usb Port), Model 01-702100, Statim 7000 (with…Scican, Ltd. · 2007
  4. Descendant · 2 generations on · 0 of its ownK090670 Midmark M3 Ultrafast Automatic SterilizerMidmark Corp. · 2009

Most influential descendants

Showing 2 of 2 · See all
  • K072466Statim 7000 (Without Usb Port), Model 01-702100, Statim 7000 (with Usb Port), Model…Scican, Ltd. · 2007 · cites it directly1descendant
  • K090670Midmark M3 Ultrafast Automatic SterilizerMidmark Corp. · 2009 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K915054Statim Cassett Autoclave, ModificationScican1992—
0 (this device)K962179Stat IM 5000 Cassette AutoclaveScican1997—
+1K072466Statim 7000 (Without Usb Port), Model 01-702100, Statim 7000 (with Usb Port)…Scican, Ltd.2007—
+2K090670Midmark M3 Ultrafast Automatic SterilizerMidmark Corp.2009—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200710
200910
FIG. 1 — Lineage of K962179, 1992–2009Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FLE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FLE device

  1. #1 K250168 Life Ultra 25L Tabletop Mono Chamber Steam…score 12.7 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K915054Statim Cassett Autoclave, Modification
Summary page 1
…ign, composition and function to the STATIM Cassette Autoclave manufactured by SciCan, Division of Lux and Zwingenberger. The STAT/M has an approved 510 (k) application (K915054) and is currently being marketed in the USA. The main difference between the two autoclaves is that the STAT/M 5000 sterilization…
Scican1992—

Also cleared against K915054: 1 other device →

What descends from it? (2)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K072466Statim 7000 (Without Usb Port), Model 01-702100, Statim 7000 (with Usb Port), Model…Scican, Ltd.20071

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has no recall.
  • Of its descendants, K072466 (Scican, Ltd., 2007) has gone furthest, with 1 descendant of its own.
  • 1 other device also cites its predicate K915054.

How does its category compare?

All FLE devices

FLE Sterilizer, Steam: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · FLE

4in 2025

2007: 52008: 12009: 82010: 22011: 52012: 52013: 52014: 22015: 32016: 12017: 32018: 82019: 52020: 22021: 52022: 102023: 72024: 22025: 42026 (partial year): 3
2007high 10 (2022) · low 1 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FLE

184in 2025

2007: 67 days median2009: 109 days median2011: 141 days median2012: 111 days median2013: 241 days median2015: 302 days median2017: 233 days median2018: 146 days median2019: 192 days median2021: 147 days median2022: 133 days median2023: 88 days median2025: 184 days median2026 (partial year): 85 days median
2007high 302 (2015) · low 67 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FLE

1in 2021

2007: 12008: 12011: 12012: 12013: 22015: 12018: 12019: 22021: 12026 (partial year): 1
2007high 2 (2013) · low 0 (2009)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FLE

71in 2025

2007: 192008: 212009: 192010: 202011: 632012: 692013: 1122014: 1132015: 682016: 762017: 632018: 902019: 912020: 872021: 812022: 702023: 662024: 642025: 71
2007high 113 (2014) · low 19 (2007)2025

MAUDE reports naming product code FLE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007567119
20081—121
20098109019
20102—020
20115141163
20125111169
201352412112
20142—0113
20153302168
20161—076
20173233063
20188146190
20195192291
20202—087
20215147181
202210133070
2023788066
20242—064
20254184071
20263851—

What is it indicated for?

The STAT/M 5000 Cassette Autoclave is a table-top steam sterilizer designed for the reprocessing of reusable instruments in a clinical setting. The STAT/M 5000 has four microprocessor-controlled, pre-programmed cycles. The first cycle operates at 132 °C for 3.5 minutes and is used for light loads of up to 0.5 kg solid stainless steel instruments free from deep cavities or holes, for example, scalers, forceps or scissors. The second cycle operates at 132…

Read the full statement (169 words)

The STAT/M 5000 Cassette Autoclave is a table-top steam sterilizer designed for the reprocessing of reusable instruments in a clinical setting. The STAT/M 5000 has four microprocessor-controlled, pre-programmed cycles. The first cycle operates at 132 °C for 3.5 minutes and is used for light loads of up to 0.5 kg solid stainless steel instruments free from deep cavities or holes, for example, scalers, forceps or scissors. The second cycle operates at 132 °C for 6 minutes and is for up to 1.5 kg of stainless steel instruments, such as those described above, and instruments of more complex construction, such as handpieces, wrapped in paper/paper or paper/plastic-film packs. The third cycle operates at 121 °C for 35 minutes and is for up to 0.4 kg of rubber, plastic or heat-sensitive items, for example, plastic mouth mirrors, suction tips or impression trays. The fourth cycle is designed for up to 1.5 kg of unwrapped stainless steel instrtuments, such as those described above, and instruments of more complex construction, such as handpieces.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K962179: Stat IM 5000 Cassette Autoclave.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K962179

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026