510(k) Report

510(k) · Cleared

Cleared

Quantim 16 Autoclave

K021064510(k) TraditionalCleared Aug 8, 2002128-day reviewClass II21 CFR 880.6880

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Scican · FLE Sterilizer, Steam

Cleared on Aug 8, 2002 as substantially equivalent to K990189 (Midmark Corp., 1999) after a 128-day review. Its family tree reaches back 2 generations to 1994.

Decided
Aug 8, 2002
Review
128 days
Class
Class II
Product code
FLE
Recalls
None
Predicates
1

Its longest line of descent runs back 2 generations to K936181 (1994). No later device cites it as a predicate yet.

  1. Root · 2 generations backK936181 Midmark M-11 Ultraclave Steam SterilizerMidmark Corp. · 1994 · Recalled
  2. 1 generation backK990189 Modification of: Midmark M-11 Ultraclave Steam SterilizerMidmark Corp. · 1999
  3. This deviceK021064 Quantim 16 AutoclaveScican · 2002
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K936181Midmark M-11 Ultraclave Steam SterilizerMidmark Corp.19941
−1K990189Modification of: Midmark M-11 Ultraclave Steam SterilizerMidmark Corp.1999—
0 (this device)K021064Quantim 16 AutoclaveScican2002—
FIG. 1 — Lineage of K021064, 1994–2002Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FLE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FLE device

  1. #1 K250168 Life Ultra 25L Tabletop Mono Chamber Steam…score 12.7 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K990189Modification of: Midmark M-11 Ultraclave Steam SterilizerMidmark Corp.1999—

Also cleared against K990189: 4 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 8 years and passes through 2 companies; a new generation was cleared every 4.4 years on median.
  • 1 of the 2 ancestors has a recall of any class.
  • 4 other devices also cite its predicate K990189.

How does its category compare?

All FLE devices

FLE Sterilizer, Steam: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · FLE

4in 2025

2007: 52008: 12009: 82010: 22011: 52012: 52013: 52014: 22015: 32016: 12017: 32018: 82019: 52020: 22021: 52022: 102023: 72024: 22025: 42026 (partial year): 3
2007high 10 (2022) · low 1 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FLE

184in 2025

2007: 67 days median2009: 109 days median2011: 141 days median2012: 111 days median2013: 241 days median2015: 302 days median2017: 233 days median2018: 146 days median2019: 192 days median2021: 147 days median2022: 133 days median2023: 88 days median2025: 184 days median2026 (partial year): 85 days median
2007high 302 (2015) · low 67 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FLE

1in 2021

2007: 12008: 12011: 12012: 12013: 22015: 12018: 12019: 22021: 12026 (partial year): 1
2007high 2 (2013) · low 0 (2009)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FLE

71in 2025

2007: 192008: 212009: 192010: 202011: 632012: 692013: 1122014: 1132015: 682016: 762017: 632018: 902019: 912020: 872021: 812022: 702023: 662024: 642025: 71
2007high 113 (2014) · low 19 (2007)2025

MAUDE reports naming product code FLE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007567119
20081—121
20098109019
20102—020
20115141163
20125111169
201352412112
20142—0113
20153302168
20161—076
20173233063
20188146190
20195192291
20202—087
20215147181
202210133070
2023788066
20242—064
20254184071
20263851—

What is it indicated for?

The QUANTlM 16 tabletop autoclave is a gravity steam autoclave designed to process medical and dental instruments to achieve successful sterilization in a clinical setting. It utilizes saturated stearn at high pressures in order to attain an effective kill of infectious bio-organisms and prevent cross infection. The unit is intended to sterilize heat and moisture stable medical and dental instruments (including dental handpieces) which are commonly found…

Read the full statement (258 words)

The QUANTlM 16 tabletop autoclave is a gravity steam autoclave designed to process medical and dental instruments to achieve successful sterilization in a clinical setting. It utilizes saturated stearn at high pressures in order to attain an effective kill of infectious bio-organisms and prevent cross infection.

The unit is intended to sterilize heat and moisture stable medical and dental instruments (including dental handpieces) which are commonly found in medical and dental offices, hospitals, clinics, and other facilities. The instruments must be suitable for steam sterilization at 132°C (270°F). Two additional cycles are also specified for (i) the sterilimtion of textiles and wrapped surgical packs and (ii) liquid loads contained in open top glass flasks. Hovever, the QUANTIM 16 Table-top Autoclave is not intended nor recommended for sterilization of liquids intended for direct patient contact.

The sterilization cycles as to established time, temperatures, load configurations, drying time and indicated uses for the QUANTM 16 Table-top Autoclave are as follows:

CYCLE | STERILIZATION TEMPERATURE | MAXMUM LOAD | STERILIZATION TIME | DRYING TIME | INTENDED USE

Unwrapped | 133°C (270°F) | 2k-g (4.41bs) | 3 min. | 30 min. | Flash cycle for solid and hinged instruments, dental handpieces

Wrapped | 132°C (270°F) | lkg (2.21bs) | 15 min. | 30 min. | Solid and hinged instruments, dental handpieces

Packs | 121”~ (250°F) | Ikg (2.21bs) | 30 rnin. | 30 rnin. | Textiles and wrapped surEical~packs

Liquids | | | 30 rnin. | N/A (no drying phase) | Liquids suitable for sterilization in open top flasks (not intended for

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K021064: Quantim 16 Autoclave.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K021064

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026