Coulter Gen-S System

K962988510(k) TraditionalCleared Oct 30, 199690-day reviewClass II21 CFR 864.5200Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Coulter Corp. · GKL Counter, Cell, Automated (Particle Counter)

Cleared on Oct 30, 1996 as substantially equivalent to K885093 (Coulter Electronics, Inc., 1988) and 1 other predicate after a 90-day review. Its family tree reaches back 1 generation to 1988. 6 later devices descend from it. It has been recalled 4 times.

Decided
Oct 30, 1996
Review
90 days
Class
Class II
Product code
GKL
Recalls
4
Predicates
2

Its longest line of descent runs back 1 generation to K885093 (1988). 2 ancestors in all. 6 later devices descend from it, 3 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1993: 1 other ancestor2004: 1 descendant cleared2000: 1 descendant cleared2002: 1 descendant cleared2005: 1 descendant cleared2008: 1 descendant cleared2015: 1 descendant clearedK885093 · Coulter STKS · Coulter Electronics, Inc. · 19881990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor 1 generation forward: 3 descendants2 generations forward: 2 descendants3 generations forward: 1 descendant K885093 · Coulter STKS · Coulter Electronics, Inc. · 19880
  1. Root · 1 generation backK885093 Coulter STKSCoulter Electronics, Inc. · 1988
  2. This deviceK962988 Coulter Gen-S SystemCoulter Corp. · 1996
  3. Descendant · cites it directly · 2 of its ownK022200 Pentra 120/Pentra 120 Retic Options SPS, Model Pentra 120Abx Diagnostics · 2002 · Recalled
  4. Descendant · 2 generations on · 1 of its ownK050719 Abx Pentra DX 120: Option SPS EvolutionHoriba Abx · 2005 · Recalled
  5. Descendant · cites it directly · 1 of its ownK041882 Flowcare SystemPointcare Technologies, Inc. · 2004
  6. Descendant · 3 generations on · 0 of its ownK151133 Pentra XLRHoriba Abx Sas · 2015
  7. Descendant · 2 generations on · 0 of its ownK072869 Pointcare Now SystemPointcare Technologies, Inc. · 2008 · Recalled

Showing 5 of 6 descendants. See all descendants

Descendants cleared per year

2000: 12002: 12004: 12005: 12008: 12015: 1
2000high 1 (2000) · low 0 (2001)2026

Most influential descendants

Showing 5 of 6 · See all
  • K022200Pentra 120/Pentra 120 Retic Options SPS, Model Pentra 120Abx Diagnostics · 2002 · cites it directly2descendants
  • K050719Abx Pentra DX 120: Option SPS EvolutionHoriba Abx · 2005 · 2 generations on1descendant
  • K041882Flowcare SystemPointcare Technologies, Inc. · 2004 · cites it directly1descendant
  • K151133Pentra XLRHoriba Abx Sas · 2015 · 3 generations on0descendants
  • K072869Pointcare Now SystemPointcare Technologies, Inc. · 2008 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K885093Coulter STKSCoulter Electronics, Inc.1988—
0 (this device)K962988Coulter Gen-S SystemCoulter Corp.1996—
+1K022200Pentra 120/Pentra 120 Retic Options SPS, Model Pentra 120Abx Diagnostics20022
+2K050719Abx Pentra DX 120: Option SPS EvolutionHoriba Abx20051
+1K041882Flowcare SystemPointcare Technologies, Inc.2004—
+3K151133Pentra XLRHoriba Abx Sas2015—
+2K072869Pointcare Now SystemPointcare Technologies, Inc.20081

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199310

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200010
200210
200410
200510
200810
201510
FIG. 1 — Lineage of K962988, 1988–2015Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in GKL? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K885093Coulter STKS
Summary page 1
…864.5850 Device Class: Class I, Exempt 510(k) numbers of Coulter devices to which substantial equivalence is claimed: COULTER® STKS Analyzer with Reticulocyte Analysis, K885093 and K932030 The product is an Automated Differential Cell Counter which like the predicate device COULTER® STKS Analyzer with Re…
Coulter Electronics, Inc.1988—
*K932030Coulter STKS with Reticulocyte Analysis Capability
Summary page 1
…ice Class: Class I, Exempt 510(k) numbers of Coulter devices to which substantial equivalence is claimed: COULTER® STKS Analyzer with Reticulocyte Analysis, K885093 and K932030 The product is an Automated Differential Cell Counter which like the predicate device COULTER® STKS Analyzer with Reticulocyte A…
Coulter Corp.1993—

Also cleared against K885093: 3 other devices →

What descends from it? (6)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K022200Pentra 120/Pentra 120 Retic Options SPS, Model Pentra 120Abx Diagnostics20022
K041882Flowcare SystemPointcare Technologies, Inc.20041
K993356Coulter Hematology Analyzers with Irf & MRV AnalysesCoulter Corp.20000

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 31, 2011Not classifiedZ-1956-2011Software designThis recall was initiated because Beckman Coulter identified a problem with some date fields that are transmitted to the Laboratory Information System (LIS). Starting Jan 1, 2011, the date fields will transmit as a blank field to the LIS…
Apr 18, 2007Not classifiedZ-0315-2008Software designPatient Mis-identification can occur: 1)- When a positive identifier (Sample ID or Cassette/Position) is manually edited to a positive identifier that is already in the To Do list. The workstation will accept the entry and no error…
Dec 15, 2005Not classifiedZ-0475-06OtherThere is a risk of sample misidentification when processing samples in the Manual aspiration mode.
Dec 10, 2002Not classifiedZ-0571-03OtherDevice design allows for cross contamination issue to occur, possibly affecting assays and results reports.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • None of the 2 ancestors has a recall.
  • Of its descendants, K022200 (Abx Diagnostics, 2002) has gone furthest, with 2 descendants of its own.
  • 3 other devices also cite its predicate K885093.

How does its category compare?

All GKL devices

GKL Counter, Cell, Automated (Particle Counter): FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · GKL

1in 2016

2007: 12010: 12016: 1
2007high 1 (2007) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · GKL

Too few GKL 510(k)s in any year since 2007 to give a median.

Recalls per year · GKL

1in 2018

2007: 12011: 12014: 12018: 1
2007high 1 (2007) · low 0 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · GKL

1in 2025

2007: 252008: 42009: 12011: 312012: 142013: 112016: 12017: 12018: 22020: 132021: 112022: 72023: 12025: 1
2007high 31 (2011) · low 0 (2010)2025

MAUDE reports naming product code GKL, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—125
20080—04
20090—01
20101—00
20110—131
20120—014
20130—011
20140—10
20150—00
20161—01
20170—01
20180—12
20190—00
20200—013
20210—011
20220—07
20230—01
20240—00
20250—01
20260—0—

What is it indicated for?

The COULTER® GEN'S system is a quantitative, automated hematology analyzer and leukocyte differential counter For In Vitro Diagnostic Use in clinical laboratories. The GEN-S System also provides automated Reticulocyte analysis. The COULTER GEN'S SM™ INTEGRATED SLIDEMAKER Option prepares a blood film on a clean microscope slide.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K962988: Coulter Gen-S System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K962988

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026