510(k) Report

510(k) · Cleared

Cleared

Castle Series 100 Steam Sterilizers (Straightline)

K970907510(k) TraditionalCleared Nov 12, 1997246-day reviewClass II21 CFR 880.6880

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Getinge/Castle, Inc. · FLE Sterilizer, Steam

Cleared on Nov 12, 1997 as substantially equivalent to K820783 (Sybron Corp., 1982) after a 246-day review. Its family tree reaches back 1 generation to 1982. 24 later devices descend from it.

Decided
Nov 12, 1997
Review
246 days
Class
Class II
Product code
FLE
Recalls
None
Predicates
1

Its longest line of descent runs back 1 generation to K820783 (1982). 24 later devices descend from it, 1 citing it directly.

Descendants, by year cleared 2000: 1 descendant cleared2001: 1 descendant cleared2004: 1 descendant cleared2003: 2 descendants cleared2011: 2 descendants cleared2007: 1 descendant cleared2002: 1 descendant cleared2012: 2 descendants cleared2014: 2 descendants cleared2013: 4 descendants cleared2018: 1 descendant cleared2021: 2 descendants cleared2023: 2 descendants cleared2026: 1 descendant cleared2022: 1 descendant clearedK820783 · Castle Microcomputer Sterilizer Control · Sybron Corp. · 198219801990200020102020
Descendants, by generation 1 generation forward: 1 descendant2 generations forward: 2 descendants3 generations forward: 4 descendants4 generations forward: 8 descendants5 generations forward: 7 descendants6 generations forward: 2 descendants K820783 · Castle Microcomputer Sterilizer Control · Sybron Corp. · 19820+5
  1. Root · 1 generation backK820783 Castle Microcomputer Sterilizer ControlSybron Corp. · 1982
  2. This deviceK970907 Castle Series 100 Steam Sterilizers (Straightline)Getinge/Castle, Inc. · 1997
  3. Descendant · cites it directly · 23 of its ownK994314 Castle Series 100HC Steam SterilizersGetinge/Castle, Inc. · 2000
  4. Descendant · 2 generations on · 21 of its ownK012573 Castle 400HC/500HC Series Steam Sterilizer, Models 422HC, 433HC…Getinge/Castle, Inc. · 2001
  5. Descendant · 3 generations on · 8 of its ownK030789 Amsco Century Steam SterilizerSTERIS Corporation · 2003
  6. Descendant · 3 generations on · 5 of its ownK070657 Getinge 633HC Series Vacuum/Gravity Steam SterilizerGetinge Sourcing, LLC · 2007
  7. Descendant · 4 generations on · 3 of its ownK172159 Getinge GSS67N Series Steam SterilizerMaquet GmbH · 2018

Showing 5 of 24 descendants. See all descendants

Descendants cleared per year

2000: 12001: 12002: 12003: 22004: 12007: 12011: 22012: 22013: 42014: 22018: 12021: 22022: 12023: 22026: 1
2000high 4 (2013) · low 0 (2005)2026

Most influential descendants

Showing 5 of 24 · See all
  • K994314Castle Series 100HC Steam SterilizersGetinge/Castle, Inc. · 2000 · cites it directly23descendants
  • K012573Castle 400HC/500HC Series Steam Sterilizer, Models 422HC, 433HC, 522HC, and 533HCGetinge/Castle, Inc. · 2001 · 2 generations on21descendants
  • K030789Amsco Century Steam SterilizerSTERIS Corporation · 2003 · 3 generations on8descendants
  • K070657Getinge 633HC Series Vacuum/Gravity Steam SterilizerGetinge Sourcing, LLC · 2007 · 3 generations on5descendants
  • K172159Getinge GSS67N Series Steam SterilizerMaquet GmbH · 2018 · 4 generations on3descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K820783Castle Microcomputer Sterilizer ControlSybron Corp.1982—
0 (this device)K970907Castle Series 100 Steam Sterilizers (Straightline)Getinge/Castle, Inc.1997—
+1K994314Castle Series 100HC Steam SterilizersGetinge/Castle, Inc.2000—
+2K012573Castle 400HC/500HC Series Steam Sterilizer, Models 422HC, 433HC, 522HC, and…Getinge/Castle, Inc.2001—
+3K030789Amsco Century Steam SterilizerSTERIS Corporation2003—
+3K070657Getinge 633HC Series Vacuum/Gravity Steam SterilizerGetinge Sourcing, LLC2007—
+4K172159Getinge GSS67N Series Steam SterilizerMaquet GmbH2018—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200010
200110
200210
200320
200410
200710
201120
201220
201340
201420
201810
202120
202210
202320
202610
FIG. 1 — Lineage of K970907, 1982–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FLE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FLE device

  1. #1 K250168 Life Ultra 25L Tabletop Mono Chamber Steam…score 12.7 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K820783Castle Microcomputer Sterilizer Control
Summary page 1
…team Sterilizer (per 21 CFR 880.6880) Per 21CFR 807.92(a)(3) We believe this product to be substantially equivalent to our Castle Microcomputer Contros and Sterilizers K820783. Per 21CFR 807.92(a)(4), (5) & (6) The Castle Series 100 Sterilizers (Straightline) is an upgrade of an existing microcomputer…
Sybron Corp.1982—

Also cleared against K820783: 2 other devices →

What descends from it? (24)

See all 24

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K994314Castle Series 100HC Steam SterilizersGetinge/Castle, Inc.200023

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has no recall.
  • Of its descendants, K994314 (Getinge/Castle, Inc., 2000) has gone furthest, with 23 descendants of its own.
  • 2 other devices also cite its predicate K820783.

How does its category compare?

All FLE devices

FLE Sterilizer, Steam: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · FLE

4in 2025

2007: 52008: 12009: 82010: 22011: 52012: 52013: 52014: 22015: 32016: 12017: 32018: 82019: 52020: 22021: 52022: 102023: 72024: 22025: 42026 (partial year): 3
2007high 10 (2022) · low 1 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FLE

184in 2025

2007: 67 days median2009: 109 days median2011: 141 days median2012: 111 days median2013: 241 days median2015: 302 days median2017: 233 days median2018: 146 days median2019: 192 days median2021: 147 days median2022: 133 days median2023: 88 days median2025: 184 days median2026 (partial year): 85 days median
2007high 302 (2015) · low 67 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FLE

1in 2021

2007: 12008: 12011: 12012: 12013: 22015: 12018: 12019: 22021: 12026 (partial year): 1
2007high 2 (2013) · low 0 (2009)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FLE

71in 2025

2007: 192008: 212009: 192010: 202011: 632012: 692013: 1122014: 1132015: 682016: 762017: 632018: 902019: 912020: 872021: 812022: 702023: 662024: 642025: 71
2007high 113 (2014) · low 19 (2007)2025

MAUDE reports naming product code FLE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007567119
20081—121
20098109019
20102—020
20115141163
20125111169
201352412112
20142—0113
20153302168
20161—076
20173233063
20188146190
20195192291
20202—087
20215147181
202210133070
2023788066
20242—064
20254184071
20263851—

What is it indicated for?

The Getinge/Castle Series 100 Steam Sterilizers (Straightline) with the PACS 2000 controls is an upgrade of an existing microcomputer control system for a comprehensive line of health care reusable medical reprocessing equipment. The PACS 2000 controls intended to be used for the control of a variety of equipment including washers, washing/disinfectors, washer/sterilizers and sterilizers. This submittal is made for the PACS 2000 controls applied to the…

Read the full statement (401 words)

The Getinge/Castle Series 100 Steam Sterilizers (Straightline) with the PACS 2000 controls is an upgrade of an existing microcomputer control system for a comprehensive line of health care reusable medical reprocessing equipment. The PACS 2000 controls intended to be used for the control of a variety of equipment including washers, washing/disinfectors, washer/sterilizers and sterilizers. This submittal is made for the PACS 2000 controls applied to the Getinge/Castle (Straightline) line of steam sterilizer models (see attachment Section 18 - Competitive Comparison). The Series 100 (Straightline) sterilizer line includes two sizes, A 16” x 16” x 26” chamber having a volume 3.9 cubic feet and a 20” x 20” x 38” chamber having a volume 8.8 cubic feet. Both chamber sizes include single and double door sterilizers with a power door offered on the single door model only. The Series 100 Steam Sterilizer (Straightline) includes gravity (Model 122) and prevacuum (Model 133) steam sterilizers for healthcare facilities and moist heat (Model 123) sterilizes for laboratories and industrial uses. The latter applications are addressed in this submittal for the rare occasion health care facilities may use them. The to be cleared for each of the models are as follows:

For the Model 122

P1 - GRAV WRAP 1 - Gravity Cycle - 30 minute exposure at 250°F and 30 minute dry

P2 - GRAV WRAP 2 - Gravity Cycle - 10 minute exposure at 275°F and 30 minute dry

P3 - FLASH 3+ - Gravity Cycle with 3 minute exposure at 275°F and 0 minute dry

P4 - FLASH 10+ - Gravity Cycle with 10 minute exposure at 275°F and 0 minute dry

For the Model 123

P1 - GRAV WRAP 1 - Gravity Cycle - 30 minute exposure at 250°F and 30 minute dry

P4 - FLASH 3+ - Gravity Cycle - 3 minute exposure at 275°F and o minutes dry

For the Model 133

P1 - PREVA WRAP - Prevacuum Cycle - 3 minute exposure at 275°F and 16 minute dry

P2 - PREVA WRP1 - Prevacuum Cycle - 3 minute exposure at 275°F and 3 minute dry

P3 - FLASH 3+ - Gravity Cycle - 3 minute exposure at 275°F and 0 minute dry

P4 - FLASH 10+ - Gravity Cycle - 10 minute exposure at 275°F and 0 minute dry

P5 - VAC LEAK TST - Air leak Test - 3 minute exposure at 268°F and 15 min. leak test

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K970907: Castle Series 100 Steam Sterilizers (Straightline).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K970907

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026