510(k) Report

510(k) · Cleared

Cleared

Amsco Century Steam Sterilizer

K030789510(k) SpecialCleared Mar 26, 200314-day reviewClass II21 CFR 880.6880

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

STERIS Corporation · FLE Sterilizer, Steam

Cleared on Mar 26, 2003 as substantially equivalent to K964332 (Amsco Intl., Inc., 1998) and 1 other predicate after a 14-day review. Its family tree reaches back 4 generations to 1982. 8 later devices descend from it.

Decided
Mar 26, 2003
Review
14 days
Class
Class II
Product code
FLE
Recalls
None
Predicates
2

Its longest line of descent runs back 4 generations to K820783 (1982). 5 ancestors in all. 8 later devices descend from it, 3 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1998: 1 other ancestor2012: 1 descendant cleared2011: 1 descendant cleared2014: 2 descendants cleared2021: 1 descendant cleared2023: 2 descendants cleared2013: 1 descendant clearedK820783 · Castle Microcomputer Sterilizer Control · Sybron Corp. · 1982K970907 · Castle Series 100 Steam Sterilizers (Straightline) · Getinge/Castle, Inc. · 1997K994314 · Castle Series 100HC Steam Sterilizers · Getinge/Castle, Inc. · 2000K012573 · Castle 400HC/500HC Series Steam Sterilizer, Models 422HC, 433HC, 522HC, and 533HC · Getinge/Castle, Inc. · 200119801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor 1 generation forward: 3 descendants2 generations forward: 4 descendants3 generations forward: 1 descendant K012573 · Castle 400HC/500HC Series Steam Sterilizer, Models 422HC, 433HC, 522HC, and 533HC · Getinge/Castle, Inc. · 2001K994314 · Castle Series 100HC Steam Sterilizers · Getinge/Castle, Inc. · 2000K970907 · Castle Series 100 Steam Sterilizers (Straightline) · Getinge/Castle, Inc. · 1997K820783 · Castle Microcomputer Sterilizer Control · Sybron Corp. · 19820
  1. Root · 4 generations backK820783 Castle Microcomputer Sterilizer ControlSybron Corp. · 1982
  2. 3 generations backK970907 Castle Series 100 Steam Sterilizers (Straightline)Getinge/Castle, Inc. · 1997
  3. 2 generations backK994314 Castle Series 100HC Steam SterilizersGetinge/Castle, Inc. · 2000
  4. 1 generation backK012573 Castle 400HC/500HC Series Steam Sterilizer, Models 422HC, 433HC…Getinge/Castle, Inc. · 2001
  5. This deviceK030789 Amsco Century Steam SterilizerSTERIS Corporation · 2003
  6. Descendant · cites it directly · 3 of its ownK111223 Amsco Chimeron Small Steam SterilizerSTERIS Corporation · 2011 · Recalled
  7. Descendant · cites it directly · 2 of its ownK120149 Skyron Integrity 215 Steam Sterilizer Skytron Integrity 215 SG Steam…Sakura Seiki Co., Ltd. · 2012
  8. Descendant · 2 generations on · 1 of its ownK221474 PRIMUS Healthcare Sterilizer (Model PSS11-HC)Primus Sterilizer Company, LLC · 2023
  9. Descendant · 3 generations on · 0 of its ownK233132 PRIMUS Healthcare Sterilizer (Model PSS11-HA & PSS11-HB)Primus Sterilizer Company, LLC · 2023
  10. Descendant · 2 generations on · 0 of its ownK212424 AMSCO 400 Small Steam Sterilizers, AMSCO 400 Medium Steam SterilizersSTERIS Corporation · 2021

Showing 5 of 8 descendants. See all descendants

Descendants cleared per year

2011: 12012: 12013: 12014: 22021: 12023: 2
2011high 2 (2014) · low 0 (2015)2026

Most influential descendants

Showing 5 of 8 · See all
  • K111223Amsco Chimeron Small Steam SterilizerSTERIS Corporation · 2011 · cites it directly3descendants
  • K120149Skyron Integrity 215 Steam Sterilizer Skytron Integrity 215 SG Steam Sterilizer Sakura…Sakura Seiki Co., Ltd. · 2012 · cites it directly2descendants
  • K221474PRIMUS Healthcare Sterilizer (Model PSS11-HC)Primus Sterilizer Company, LLC · 2023 · 2 generations on1descendant
  • K233132PRIMUS Healthcare Sterilizer (Model PSS11-HA & PSS11-HB)Primus Sterilizer Company, LLC · 2023 · 3 generations on0descendants
  • K212424AMSCO 400 Small Steam Sterilizers, AMSCO 400 Medium Steam SterilizersSTERIS Corporation · 2021 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K820783Castle Microcomputer Sterilizer ControlSybron Corp.1982—
−3K970907Castle Series 100 Steam Sterilizers (Straightline)Getinge/Castle, Inc.1997—
−2K994314Castle Series 100HC Steam SterilizersGetinge/Castle, Inc.2000—
−1K012573Castle 400HC/500HC Series Steam Sterilizer, Models 422HC, 433HC, 522HC, and…Getinge/Castle, Inc.2001—
0 (this device)K030789Amsco Century Steam SterilizerSTERIS Corporation2003—
+1K111223Amsco Chimeron Small Steam SterilizerSTERIS Corporation20113
+1K120149Skyron Integrity 215 Steam Sterilizer Skytron Integrity 215 SG Steam…Sakura Seiki Co., Ltd.2012—
+2K221474PRIMUS Healthcare Sterilizer (Model PSS11-HC)Primus Sterilizer Company, LLC2023—
+3K233132PRIMUS Healthcare Sterilizer (Model PSS11-HA & PSS11-HB)Primus Sterilizer Company, LLC2023—
+2K212424AMSCO 400 Small Steam Sterilizers, AMSCO 400 Medium Steam SterilizersSTERIS Corporation2021—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199810

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201110
201210
201310
201420
202110
202320
FIG. 1 — Lineage of K030789, 1982–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FLE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FLE device

  1. #1 K250168 Life Ultra 25L Tabletop Mono Chamber Steam…score 12.7 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K964332Eagle Century Steam Sterilizer
Summary page 1
…stabile materials used in healthcare facilities. The Amsco Century Steam Sterilizer is substantially equivalent with the predicate device Eagle Century Steam Sterilizer, K964332, cleared June 1,1998 and also substantially equivalent to Castle 400HC/500HC Series Steam Sterilizer, K012573, cleared September…
Amsco Intl., Inc.19982
*K012573Castle 400HC/500HC Series Steam Sterilizer, Models 422HC, 433HC, 522HC, and 533HC
Summary page 1
…t with the predicate device Eagle Century Steam Sterilizer, K964332, cleared June 1,1998 and also substantially equivalent to Castle 400HC/500HC Series Steam Sterilizer, K012573, cleared September 5,2001 for the 135°C (275°F) for 3 minutes prevac cycle claim. The Amsco Century Prevacuum Steam Sterilizer (…
Getinge/Castle, Inc.2001—

Also cleared against K964332: 1 other device →

What descends from it? (8)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K111223Amsco Chimeron Small Steam SterilizerSTERIS Corporation20113
K120149Skyron Integrity 215 Steam Sterilizer Skytron Integrity 215 SG Steam Sterilizer Sakura…Sakura Seiki Co., Ltd.20122
K141009Fort Defiance Automated Steam SterilizerFort Defiance Industries, Inc.20140

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 21 years and passes through 3 companies; a new generation was cleared every 2.3 years on median.
  • 1 of the 5 ancestors has a recall of any class.
  • Of its descendants, K111223 (STERIS Corporation, 2011) has gone furthest, with 3 descendants of its own.
  • 1 other device also cites its predicate K964332.

How does its category compare?

All FLE devices

FLE Sterilizer, Steam: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · FLE

4in 2025

2007: 52008: 12009: 82010: 22011: 52012: 52013: 52014: 22015: 32016: 12017: 32018: 82019: 52020: 22021: 52022: 102023: 72024: 22025: 42026 (partial year): 3
2007high 10 (2022) · low 1 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FLE

184in 2025

2007: 67 days median2009: 109 days median2011: 141 days median2012: 111 days median2013: 241 days median2015: 302 days median2017: 233 days median2018: 146 days median2019: 192 days median2021: 147 days median2022: 133 days median2023: 88 days median2025: 184 days median2026 (partial year): 85 days median
2007high 302 (2015) · low 67 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FLE

1in 2021

2007: 12008: 12011: 12012: 12013: 22015: 12018: 12019: 22021: 12026 (partial year): 1
2007high 2 (2013) · low 0 (2009)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FLE

71in 2025

2007: 192008: 212009: 192010: 202011: 632012: 692013: 1122014: 1132015: 682016: 762017: 632018: 902019: 912020: 872021: 812022: 702023: 662024: 642025: 71
2007high 113 (2014) · low 19 (2007)2025

MAUDE reports naming product code FLE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007567119
20081—121
20098109019
20102—020
20115141163
20125111169
201352412112
20142—0113
20153302168
20161—076
20173233063
20188146190
20195192291
20202—087
20215147181
202210133070
2023788066
20242—064
20254184071
20263851—

What is it indicated for?

The Am~coCentury Steam Sterilizer is designed for efficient steriltzation of non-porous and porous, heat and moisture-stabile materials used in healthcare facilities. The Amsco Century Steam Sterilizer is available in the following configurations: 16” x 16” x 26” Single Door Gravity* 16” x 16” x 26” Single Door Prevacuum 16” x 16” x 26” Double Door Gravity 16” x 16” x 26” Double Door Prevacuum 20” x 20” x 38” Single Door Gravity 20” x 20” x 38” Single…

Read the full statement (342 words)

The Am~coCentury Steam Sterilizer is designed for efficient steriltzation of non-porous and porous, heat and moisture-stabile materials used in healthcare facilities. The Amsco Century Steam Sterilizer is available in the following configurations:

16” x 16” x 26” Single Door Gravity* 16” x 16” x 26” Single Door Prevacuum

16” x 16” x 26” Double Door Gravity 16” x 16” x 26” Double Door Prevacuum

20” x 20” x 38” Single Door Gravity 20” x 20” x 38” Single Door Prevacuum

20” x 20’’ x 38” Double Door Gravity 20” x 20” x 38” Double Door Prevacuum

*Note: No changes are being made to the gravity sterilizers program.

The Amsco Century Steam Sterilizer is equipped with the following factory-programmed set sterilization cyde values:

Cycle — Recommended Loads — Stetilize Temp — Sterilize Time — Dry Time

Flash — Unwrapped instrument tray with a single instrument — 270°F (132°C) — 3 minutes — 1 minute

Flash — Unwrapped instrument tray with multiple non-porous instruments, maximum weight 17 pounds per tray. — 270°F (1 32OC) — 10 minutes — 1 minute

Express — Single wrapped instrument tray with a single instrument. Non-porous goods only. — 270°F (132°C) — 4 minutes — 3 minutes

Prevac — Up to two double wrapped instrument trays, maximum weight 17 pounds per tray. Up to six fabric packs. — 270°F (132°C) — 4 minutes — 20 minutes’

Prevac — Up to two double wrapped instrument trays, maximum weight 17 pounds per tray. — 275°F (135°C) — 3 minutes — 16 minutes

Cycle — Recommended Loads — Sterilize Temp — Sterilize Time — Dry Time

Flash — Unwrapped instrument tray with a single instrument — 270°F (132°C) — 3 minutes

Flash — Unwrapped instrument tray with multiple non-porous instruments, maximum weight 17 pounds per tray. — 270°F (132OC) — 10 minutes

Gravity — Up to two double wrapped instrument trays, maximum weight 17 pounds per tray — 270°F (132°C) — 15 minutes

Gravity — Fabric Packs u,p-b5 ; xr r d - ~ f /&b — 250°F (121°C) — 30 minutes2

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K030789: Amsco Century Steam Sterilizer.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K030789

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026