510(k) Report

510(k) · Cleared

Cleared

Dinamap MPS Portable Monitor

K971569510(k) TraditionalCleared Sep 19, 1997143-day reviewClass II21 CFR 870.2300

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Johnson & Johnson Medical, Inc. · MSX System, Network And Communication, Physiological Monitors

Cleared on Sep 19, 1997 as substantially equivalent to K955113 (Johnson & Johnson Medical, Inc., 1996) after a 143-day review. Its family tree reaches back 2 generations to 1989. 1 of the 54 devices in its family, counting itself, has a Class I recall. 43 later devices descend from it.

Decided
Sep 19, 1997
Review
143 days
Class
Class II
Product code
MSX
Recalls
None
Predicates
1

Its longest line of descent runs back 2 generations to K896030 (1989). 10 ancestors in all. 43 later devices descend from it, 3 citing it directly. 1 of the 54 devices in this family, counting this one, has a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 1995: 1 other ancestor1994: 1 other ancestor1992: 1 other ancestor1991: 4 other ancestors1990: 1 other ancestor1998: 1 descendant cleared2000: 1 descendant cleared1999: 1 descendant cleared2002: 4 descendants cleared2007: 3 descendants cleared2006: 1 descendant cleared2010: 3 descendants cleared2009: 1 descendant cleared2014: 4 descendants cleared, 1 with a Class I recall12013: 3 descendants cleared2021: 2 descendants cleared2012: 1 descendant cleared2022: 2 descendants cleared2017: 4 descendants cleared2016: 1 descendant cleared2015: 3 descendants cleared2019: 2 descendants cleared2024: 2 descendants cleared2020: 2 descendants cleared2018: 1 descendant cleared2025: 1 descendant clearedK896030 · Hewlett-Packard Model M1016A CO2 & M1017A FI02 · Hewlett-Packard Co. · 1989K955113 · Dinamap Select Multi-Parameter System · Johnson & Johnson Medical, Inc. · 19961990200020102020
Other ancestors, by generationDescendants, by generation 2 generations back: 8 other ancestors 1 generation forward: 3 descendants2 generations forward: 4 descendants3 generations forward: 6 descendants4 generations forward: 5 descendants5 generations forward: 5 descendants6 generations forward: 7 descendants7 generations forward: 5 descendants, 1 with a Class I recall18 generations forward: 6 descendants9 generations forward: 2 descendants K955113 · Dinamap Select Multi-Parameter System · Johnson & Johnson Medical, Inc. · 1996K896030 · Hewlett-Packard Model M1016A CO2 & M1017A FI02 · Hewlett-Packard Co. · 19890+5
  1. Root · 2 generations backK896030 Hewlett-Packard Model M1016A CO2 & M1017A FI02Hewlett-Packard Co. · 1989
  2. 1 generation backK955113 Dinamap Select Multi-Parameter SystemJohnson & Johnson Medical, Inc. · 1996
  3. This deviceK971569 Dinamap MPS Portable MonitorJohnson & Johnson Medical, Inc. · 1997
  4. Descendant · cites it directly · 40 of its ownK992638 Dinamap Pro Series Monitor, Models 100, 200, 300, 400Critikon Company, LLC · 2000
  5. Descendant · 2 generations on · 22 of its ownK014255 Dinamap Procare Series Monitor; Model 100, 200, 300, 400Ge Medical Systems Information Technologies · 2002
  6. Descendant · 3 generations on · 21 of its ownK022193 Dinamap Procare Series Monitor, Models 100N, 200N, 300N, 400NGe Medical Systems Information Technologies · 2002
  7. Descendant · 4 generations on · 20 of its ownK073203 Carescape V100Ge Medical Systems Information Technologies · 2007
  8. Descendant · 2 generations on · 13 of its ownK063055 VS-800 Vital Signs MonitorShenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2006

Showing 5 of 43 descendants. See all descendants

Descendants cleared per year

1998: 11999: 12000: 12002: 42006: 12007: 32009: 12010: 32012: 12013: 32014: 42015: 32016: 12017: 42018: 12019: 22020: 22021: 22022: 22024: 22025: 1
1998high 4 (2002) · low 0 (2001)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 43 · See all
  • K992638Dinamap Pro Series Monitor, Models 100, 200, 300, 400Critikon Company, LLC · 2000 · cites it directly40descendants
  • K014255Dinamap Procare Series Monitor; Model 100, 200, 300, 400Ge Medical Systems Information Technologies · 2002 · 2 generations on22descendants
  • K022193Dinamap Procare Series Monitor, Models 100N, 200N, 300N, 400NGe Medical Systems Information Technologies · 2002 · 3 generations on21descendants
  • K073203Carescape V100Ge Medical Systems Information Technologies · 2007 · 4 generations on20descendants
  • K063055VS-800 Vital Signs MonitorShenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2006 · 2 generations on13descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K896030Hewlett-Packard Model M1016A CO2 & M1017A FI02Hewlett-Packard Co.1989—
−1K955113Dinamap Select Multi-Parameter SystemJohnson & Johnson Medical, Inc.1996—
0 (this device)K971569Dinamap MPS Portable MonitorJohnson & Johnson Medical, Inc.1997—
+1K992638Dinamap Pro Series Monitor, Models 100, 200, 300, 400Critikon Company, LLC2000—
+2K014255Dinamap Procare Series Monitor; Model 100, 200, 300, 400Ge Medical Systems Information Technologies2002—
+3K022193Dinamap Procare Series Monitor, Models 100N, 200N, 300N, 400NGe Medical Systems Information Technologies2002—
+4K073203Carescape V100Ge Medical Systems Information Technologies2007—
+2K063055VS-800 Vital Signs MonitorShenzhen Mindray Bio-Medical Electronics Co., Ltd.2006—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199010
199140
199210
199410
199510

Descendants by year cleared

Year clearedDevicesWith a Class I recall
199810
199910
200010
200240
200610
200730
200910
201030
201210
201330
201441
201530
201610
201740
201810
201920
202020
202120
202220
202420
202510
FIG. 1 — Lineage of K971569, 1989–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MSX device

  1. #1 K242750 Central Stationscore 21.5 · 31 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K955113Dinamap Select Multi-Parameter System
Summary page 2
…onitor is substantially equivalent to the currently-marketed DINAMAP MPS Select Monitor (“Select Monitor”) which received marketing clearance August 15, 1996, via 510(k) K955113. The DINAMAP MPS Portable Monitor is a prescription device intended for use only by health care professionals. It can be used in…
Johnson & Johnson Medical, Inc.1996—

Also cleared against K955113: 2 other devices →

What descends from it? (43)

See all 43

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K992638Dinamap Pro Series Monitor, Models 100, 200, 300, 400Critikon Company, LLC200040
K982641Advanced Nibp ModuleJohnson & Johnson Medical, Inc.19990
K982342Dinamap MPS Select Portable MonitorCritikon Company, LLC19980

Has anything in its family been recalled? (1 Class I)

No FDA recall names this device.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K133576Monitor B40Ge Medical Systems China Co., Ltd. · 2014 · 7 generations on · ◆ in the 2014 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 8 years and passes through 2 companies; a new generation was cleared every 6.6 years on median.
  • 1 of the 10 ancestors has a recall of any class.
  • Of its descendants, K992638 (Critikon Company, LLC, 2000) has gone furthest, with 40 descendants of its own.
  • 2 other devices also cite its predicate K955113.

How does its category compare?

All MSX devices

MSX System, Network And Communication, Physiological Monitors: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · MSX

4in 2025

2007: 42008: 12009: 12010: 32011: 72012: 42013: 22014: 32015: 52016: 32017: 22018: 42019: 42020: 52021: 32022: 52023: 32024: 42025: 42026 (partial year): 5
2007high 7 (2011) · low 1 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MSX

295in 2025

2007: 58 days median2010: 43 days median2011: 72 days median2012: 145 days median2014: 165 days median2015: 138 days median2016: 174 days median2018: 111 days median2019: 92 days median2020: 193 days median2021: 227 days median2022: 177 days median2023: 244 days median2024: 212 days median2025: 295 days median2026 (partial year): 145 days median
2007high 295 (2025) · low 43 (2010)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MSX

1in 2025

2007: 12008: 12009: 22010: 12011: 12012: 22015: 22017: 22018: 12022: 12025: 12026 (partial year): 2
2007high 2 (2009) · low 0 (2013)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MSX

19in 2025

2007: 52008: 62009: 72010: 142011: 32012: 42013: 172014: 322015: 312016: 112017: 162018: 152019: 152020: 252021: 172022: 182023: 102024: 322025: 19
2007high 32 (2014) · low 3 (2011)2025

MAUDE reports naming product code MSX, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200745815
20081—16
20091—27
2010343114
201177213
2012414524
20132—017
20143165032
20155138231
20163174011
20172—216
20184111115
2019492015
20205193025
20213227017
20225177118
20233244010
20244212032
20254295119
202651452—

What is it indicated for?

The DINAMAP MPS* Select* Portable Monitor is intended to monitor a single patient’s vital signs in the hospital, outpatient surgery and healthcare practitioner facilities. The patient populations include adult, pediatric and neonatal. The Portable Monitor networking capabilities are identical to the predicate device (K955113) and include connection to the OBSERVER* Central Station via VHF, spread spectrum or hardwire communication; host communications…

Read the full statement (129 words)

The DINAMAP MPS* Select* Portable Monitor is intended to monitor a single patient’s vital signs in the hospital, outpatient surgery and healthcare practitioner facilities. The patient populations include adult, pediatric and neonatal. The Portable Monitor networking capabilities are identical to the predicate device (K955113) and include connection to the OBSERVER* Central Station via VHF, spread spectrum or hardwire communication; host communications for use on the auxiliary serial port or RS-232 serial port; and remote view protocol over Ethernet enabling communication with other devices such as currently-marketed DINAMAP* Monitors, remote display, data collection or hospital information system, or remote alarm. In addition, the Portable Monitor may be operated from internal NiMH batteries making the device portable. This device is intended for use by qualified healthcare personnel trained in its use.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K971569: Dinamap MPS Portable Monitor.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K971569

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026