510(k) Report

510(k) · Cleared

Recalled · Class III · 2018

Thumper

K972525510(k) TraditionalCleared Oct 2, 199787-day reviewClass II21 CFR 870.5200Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Michigan Instruments, Inc. · DRM Compressor, Cardiac, External

Cleared on Oct 2, 1997 as substantially equivalent to K851139 (Michigan Instruments, Inc., 1985) after an 87-day review. Its family tree reaches back 1 generation to 1985. 1 of the 13 devices in its family, counting itself, has a Class I recall. 11 later devices descend from it. It has been recalled 3 times.

Decided
Oct 2, 1997
Review
87 days
Class
Class II
Product code
DRM
Recalls
3
Predicates
1

Its longest line of descent runs back 1 generation to K851139 (1985). 11 later devices descend from it, 5 citing it directly. 1 of the 13 devices in this family, counting this one, has a Class I recall.

Descendants, by year cleared 2010: 1 descendant cleared2004: 1 descendant cleared2008: 1 descendant cleared2006: 2 descendants cleared2009: 1 descendant cleared2016: 2 descendants cleared2014: 1 descendant cleared2018: 1 descendant cleared2021: 1 descendant cleared, 1 with a Class I recall1K851139 · Thumper Cardiopulmonary Resuscitator 1005 · Michigan Instruments, Inc. · 19851990200020102020
Descendants, by generation 1 generation forward: 5 descendants2 generations forward: 2 descendants3 generations forward: 2 descendants4 generations forward: 1 descendant5 generations forward: 1 descendant, 1 with a Class I recall1 K851139 · Thumper Cardiopulmonary Resuscitator 1005 · Michigan Instruments, Inc. · 19850+5
  1. Root · 1 generation backK851139 Thumper Cardiopulmonary Resuscitator 1005Michigan Instruments, Inc. · 1985
  2. This deviceK972525 ThumperMichigan Instruments, Inc. · 1997
  3. Descendant · cites it directly · 7 of its ownK053403 LUCASJolife AB · 2006
  4. Descendant · cites it directly · 6 of its ownK073079 Thumper, Model: 1008Michigan Instruments, Inc. · 2008
  5. Descendant · cites it directly · 6 of its ownK062401 Modification to LucasJolife AB · 2006
  6. Descendant · 2 generations on · 5 of its ownK090422 LUCAS 2Jolife AB · 2009 · Recalled
  7. Descendant · 3 generations on · 2 of its ownK161768 LUCAS 3 Chest Compression SystemJolife AB · 2016 · Recalled

Showing 5 of 11 descendants. See all descendants

Descendants cleared per year

2004: 12006: 22008: 12009: 12010: 12014: 12016: 22018: 12021: 1
2004high 2 (2006) · low 0 (2005)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 11 · See all
  • K053403LUCASJolife AB · 2006 · cites it directly7descendants
  • K073079Thumper, Model: 1008Michigan Instruments, Inc. · 2008 · cites it directly6descendants
  • K062401Modification to LucasJolife AB · 2006 · cites it directly6descendants
  • K090422LUCAS 2Jolife AB · 2009 · 2 generations on5descendants
  • K161768LUCAS 3 Chest Compression SystemJolife AB · 2016 · 3 generations on2descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K851139Thumper Cardiopulmonary Resuscitator 1005Michigan Instruments, Inc.1985—
0 (this device)K972525ThumperMichigan Instruments, Inc.1997—
+1K053403LUCASJolife AB2006—
+1K073079Thumper, Model: 1008Michigan Instruments, Inc.2008—
+1K062401Modification to LucasJolife AB2006—
+2K090422LUCAS 2Jolife AB20093
+3K161768LUCAS 3 Chest Compression SystemJolife AB20162

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200410
200620
200810
200910
201010
201410
201620
201810
202111
FIG. 1 — Lineage of K972525, 1985–2021Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DRM device

  1. #1 K221700 AutoPulse NXT Resuscitation Systemscore -7.5 · 6 generations · Class I recall
  2. #2 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K851139Thumper Cardiopulmonary Resuscitator 1005
Summary page 1
…870.5200 . Substantially equivalent devices - Thumper® Model 1004: a “preamendment device.” Thumper® Model 1005: 510(k) Clearance dated May 15, 1985) reference number K851139. . Description of Thumper® Model 1007 The Thumper® Model 1007 is a pneumatically powered external cardiac compressor used on a p…
Michigan Instruments, Inc.1985—
*—Thumper® Model 1004
Named without a submission number
—

Also cleared against K851139: 4 other devices →

What descends from it? (11)

See all 11

5 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K053403LUCASJolife AB20067
K073079Thumper, Model: 1008Michigan Instruments, Inc.20086
K062401Modification to LucasJolife AB20066
K102068Miniaturize Chest Compressor (MCC)Resuscitation International20101
K024215Enertron CPR JackEnertron Engineering Co.20040

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 13, 2018Class IIIZ-1632-2018Nonconforming Material/ComponentA performance issue with the chest depth needle valve (CDNV) may arise where turning the knob will not deliver any compression depth to the piston during the first 5 revolutions of the knob.The last ¿ to ¿ revolution of the knob could…
Sep 14, 2006Not classifiedZ-0178-2007OtherFailure to initiate compressions when turned on.
Aug 4, 2006Not classifiedZ-0119-2007OtherFailure to initiate compressions when first turned on, if improperly shut down-Operator Manual updated to include proper shut down procedures

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K211289RMU-2000 Automated Chest Compression SystemDefibtech, LLC · 2021 · 5 generations on · ◆ in the 2021 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • Its one ancestor has no recall.
  • Of its descendants, K053403 (Jolife AB, 2006) has gone furthest, with 7 descendants of its own.
  • 4 other devices also cite its predicate K851139.

How does its category compare?

All DRM devices

DRM Compressor, Cardiac, External: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · DRM

1in 2023

2008: 22009: 12010: 12012: 12014: 12016: 22018: 12021: 12023: 1
2007high 2 (2008) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DRM

Too few DRM 510(k)s in any year since 2007 to give a median.

Recalls per year · DRM

1in 2022

2008: 12011: 12012: 12022: 12026 (partial year): 1
2007high 1 (2008) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DRM

690in 2025

2007: 82008: 132009: 92010: 72011: 162012: 4832013: 1,2682014: 6642015: 6742016: 6392017: 6792018: 7832019: 7062020: 8442021: 7552022: 7302023: 7652024: 7102025: 690
2007high 1,268 (2013) · low 7 (2010)2025

MAUDE reports naming product code DRM, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—08
20082—113
20091—09
20101—07
20110—116
20121—1483
20130—01,268
20141—0664
20150—0674
20162—0639
20170—0679
20181—0783
20190—0706
20200—0844
20211—0755
20220—1730
20231—0765
20240—0710
20250—0690
20260—1—

What is it indicated for?

This device is used to perform Cardiopulmonary Resuscitation (CPR) on adult patients and only adult patients in cases of clinical death as defined by a lack of spontaneous breathing and pulse.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K972525: Thumper.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K972525

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026