510(k) · Cleared

Recalled · Class II · 2018

Vitalab Flexor/Vitalab Selectra 2/Vitalab Viva

K973628510(k) TraditionalCleared Mar 13, 1998170-day reviewClass I21 CFR 862.2170Recalls 5

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Vital Scientific N.V. · JJF Analyzer, Chemistry, Micro, For Clinical Use

Cleared on Mar 13, 1998 after a 170-day review. The FDA summary names its predicate without a number. 1 later device descends from it. It has been recalled 5 times.

Decided
Mar 13, 1998
Review
170 days
Class
Class I
Product code
JJF
Recalls
5
Predicates
0

1 later device descends from it, 1 citing it directly.

Descendants, by year cleared 2010: 1 descendant cleared1995200020052010201520202025
  1. This deviceK973628 Vitalab Flexor/Vitalab Selectra 2/Vitalab VivaVital Scientific N.V. · 1998
  2. Descendant · cites it directly · 0 of its ownK093883 Ast/Got 4+1 SL, Ast/Got 4+1 SL, Elical 2, Elitrol I, Elitrol II…Seppim S.A.S. · 2010

Most influential descendants

Showing 1 of 1 · See all
  • K093883Ast/Got 4+1 SL, Ast/Got 4+1 SL, Elical 2, Elitrol I, Elitrol II, Model Assl-0250…Seppim S.A.S. · 2010 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K973628Vitalab Flexor/Vitalab Selectra 2/Vitalab VivaVital Scientific N.V.1998—
+1K093883Ast/Got 4+1 SL, Ast/Got 4+1 SL, Elical 2, Elitrol I, Elitrol II, Model…Seppim S.A.S.2010—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201010
FIG. 1 — Lineage of K973628, 1998–2010Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JJF? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—Roche Cobas Mira
Named without a submission number
—

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K093883Ast/Got 4+1 SL, Ast/Got 4+1 SL, Elical 2, Elitrol I, Elitrol II, Model Assl-0250…Seppim S.A.S.20100

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Nov 15, 2018Class IIZ-0547-2020Nonconforming Material/ComponentDuring in-house testing it was determined that their is a possibility that during a sample run, the instrument may aspirate a sample from the wrong sample container resulting in a sample result and patient ID mismatch. If not detected…
Jan 18, 2016Class IIZ-2558-2019Software Manufacturing/Software DeploymentInstrument stopped working due to a software lockup, and no patient results are produced.
Jan 10, 2012Class IIZ-0156-2020Component design/selectionThe action was initiated due power supply failure. During a power supply failure the instruments malfunctions and no patient results are produced.
Jan 10, 2012Class IIZ-0155-2020Component design/selectionThe action was initiated due power supply failure. During a power supply failure the instruments malfunctions and no patient results are produced.
Jan 10, 2012Class IIZ-0154-2020Component design/selectionThe action was initiated due power supply failure. During a power supply failure the instruments malfunctions and no patient results are produced.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K093883 (Seppim S.A.S., 2010) has gone furthest, with 0 descendants of its own.

How does its category compare?

All JJF devices

JJF Analyzer, Chemistry, Micro, For Clinical Use: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · JJF

FDA has decided no JJF device since 2007.

Median 510(k) review, days · JJF

Too few JJF 510(k)s in any year since 2007 to give a median.

Recalls per year · JJF

2in 2018

2007: 22012: 32018: 2
2007high 3 (2012) · low 0 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JJF

2in 2025

2007: 102008: 22009: 32010: 212011: 192012: 82013: 12014: 32022: 12023: 12025: 2
2007high 21 (2010) · low 0 (2015)2025

MAUDE reports naming product code JJF, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—210
20080—02
20090—03
20100—021
20110—019
20120—38
20130—01
20140—03
20150—00
20160—00
20170—00
20180—20
20190—00
20200—00
20210—00
20220—01
20230—01
20240—00
20250—02
20260—0—

What is it indicated for?

This notification is for an in vitro diagnostic device, which is an automated chemistry analyzer, micro type, intended for clinical use in conjunction with certain materials to measure a variety of analytes, including applications in clinical chemistry, monitoring drugs of abuse and in therapeutic drug monitoring.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K973628: Vitalab Flexor/Vitalab Selectra 2/Vitalab Viva.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K973628

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026