510(k) · Cleared
Recalled · Class II · 2018Vitalab Flexor/Vitalab Selectra 2/Vitalab Viva
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Vital Scientific N.V. · JJF Analyzer, Chemistry, Micro, For Clinical Use
Cleared on Mar 13, 1998 after a 170-day review. The FDA summary names its predicate without a number. 1 later device descends from it. It has been recalled 5 times.
- Decided
- Mar 13, 1998
- Review
- 170 days
- Class
- Class I
- Product code
- JJF
- Recalls
- 5
- Predicates
- 0
1 later device descends from it, 1 citing it directly.
- This deviceK973628 Vitalab Flexor/Vitalab Selectra 2/Vitalab VivaVital Scientific N.V. · 1998
- Descendant · cites it directly · 0 of its ownK093883 Ast/Got 4+1 SL, Ast/Got 4+1 SL, Elical 2, Elitrol I, Elitrol II…Seppim S.A.S. · 2010
Most influential descendants
Showing 1 of 1 · See all- K093883Ast/Got 4+1 SL, Ast/Got 4+1 SL, Elical 2, Elitrol I, Elitrol II, Model Assl-0250…Seppim S.A.S. · 2010 · cites it directly0descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| 0 (this device) | K973628 | Vitalab Flexor/Vitalab Selectra 2/Vitalab Viva | Vital Scientific N.V. | 1998 | — |
| +1 | K093883 | Ast/Got 4+1 SL, Ast/Got 4+1 SL, Elical 2, Elitrol I, Elitrol II, Model… | Seppim S.A.S. | 2010 | — |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2010 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in JJF? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
- Candidate predicates for a device like this, rankedin the report
- Recall history across each candidate’s familyin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | — | Roche Cobas Mira Named without a submission number |
What descends from it? (1)
1 later device cites it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K093883 | Ast/Got 4+1 SL, Ast/Got 4+1 SL, Elical 2, Elitrol I, Elitrol II, Model Assl-0250… | 2010 | 0 |
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Nov 15, 2018 | Class II | Z-0547-2020 | Nonconforming Material/Component | During in-house testing it was determined that their is a possibility that during a sample run, the instrument may aspirate a sample from the wrong sample container resulting in a sample result and patient ID mismatch. If not detected… |
| Jan 18, 2016 | Class II | Z-2558-2019 | Software Manufacturing/Software Deployment | Instrument stopped working due to a software lockup, and no patient results are produced. |
| Jan 10, 2012 | Class II | Z-0156-2020 | Component design/selection | The action was initiated due power supply failure. During a power supply failure the instruments malfunctions and no patient results are produced. |
| Jan 10, 2012 | Class II | Z-0155-2020 | Component design/selection | The action was initiated due power supply failure. During a power supply failure the instruments malfunctions and no patient results are produced. |
| Jan 10, 2012 | Class II | Z-0154-2020 | Component design/selection | The action was initiated due power supply failure. During a power supply failure the instruments malfunctions and no patient results are produced. |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- Of its descendants, K093883 (Seppim S.A.S., 2010) has gone furthest, with 0 descendants of its own.
How does its category compare?
All JJF devicesJJF Analyzer, Chemistry, Micro, For Clinical Use: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · JJF
FDA has decided no JJF device since 2007.
Median 510(k) review, days · JJF
Too few JJF 510(k)s in any year since 2007 to give a median.
Recalls per year · JJF
2in 2018
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · JJF
2in 2025
MAUDE reports naming product code JJF, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 0 | — | 2 | 10 |
| 2008 | 0 | — | 0 | 2 |
| 2009 | 0 | — | 0 | 3 |
| 2010 | 0 | — | 0 | 21 |
| 2011 | 0 | — | 0 | 19 |
| 2012 | 0 | — | 3 | 8 |
| 2013 | 0 | — | 0 | 1 |
| 2014 | 0 | — | 0 | 3 |
| 2015 | 0 | — | 0 | 0 |
| 2016 | 0 | — | 0 | 0 |
| 2017 | 0 | — | 0 | 0 |
| 2018 | 0 | — | 2 | 0 |
| 2019 | 0 | — | 0 | 0 |
| 2020 | 0 | — | 0 | 0 |
| 2021 | 0 | — | 0 | 0 |
| 2022 | 0 | — | 0 | 1 |
| 2023 | 0 | — | 0 | 1 |
| 2024 | 0 | — | 0 | 0 |
| 2025 | 0 | — | 0 | 2 |
| 2026 | 0 | — | 0 | — |
What is it indicated for?
This notification is for an in vitro diagnostic device, which is an automated chemistry analyzer, micro type, intended for clinical use in conjunction with certain materials to measure a variety of analytes, including applications in clinical chemistry, monitoring drugs of abuse and in therapeutic drug monitoring.
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K973628 · printed Nov 6, 2026
Cite this page
510(k) Report. “K973628: Vitalab Flexor/Vitalab Selectra 2/Vitalab Viva.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K973628
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026