510(k) · Cleared
Cleared'Rapid Drug Screen' 8 Panel-Low Volume
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
American Bio Medica Corp. · DKZ Enzyme Immunoassay, Amphetamine
Cleared on Aug 27, 1998 after a 105-day review. FDA’s file for it is a 510(k) statement, which names no predicate.
- Decided
- Aug 27, 1998
- Review
- 105 days
- Class
- Class II
- Product code
- DKZ
- Recalls
- None
- Predicates
- 0
FDA’s file for it is a 510(k) statement, which names no predicate.
Sources: FDA 510(k) database (decision fields), the FDA 510(k) statement (indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in DKZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new DKZ device
- #1 K233019 AllTest Multi-Drug Rapid Test Cup; AllTest…score 14.6 · 8 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against?
FDA’s file for it is a 510(k) statement, which names no predicate.
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
No FDA recall names this device.
No ancestor or descendant has a Class I recall.
How does its category compare?
All DKZ devicesDKZ Enzyme Immunoassay, Amphetamine: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · DKZ
1in 2024
2026 to Oct 25, outlined.
Median 510(k) review, days · DKZ
180in 2019
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · DKZ
1in 2017
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · DKZ
2in 2025
MAUDE reports naming product code DKZ, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 0 | — | 1 | 0 |
| 2008 | 2 | — | 3 | 12 |
| 2009 | 2 | — | 5 | 1 |
| 2010 | 4 | 233 | 0 | 1 |
| 2011 | 6 | 123 | 4 | 2 |
| 2012 | 5 | 55 | 0 | 0 |
| 2013 | 8 | 103 | 1 | 1 |
| 2014 | 8 | 128 | 3 | 9 |
| 2015 | 5 | 107 | 0 | 8 |
| 2016 | 5 | 215 | 0 | 8 |
| 2017 | 0 | — | 1 | 0 |
| 2018 | 3 | 128 | 0 | 0 |
| 2019 | 3 | 180 | 0 | 0 |
| 2020 | 1 | — | 0 | 0 |
| 2021 | 0 | — | 0 | 3 |
| 2022 | 0 | — | 0 | 4 |
| 2023 | 2 | — | 0 | 7 |
| 2024 | 1 | — | 0 | 1 |
| 2025 | 0 | — | 0 | 2 |
| 2026 | 0 | — | 0 | — |
What is it indicated for?
“Rapid Drug Screen” $-Panel low volume test is a one-step, lateral flow immunoassay for the simultaneous detection of eight abused substances in urine (each assay occupies a separate channel). “Rapid Drug Screen” 8-Panel low volume test is intended for use in the qualitative detection of the following 8 drugs of abuse in human urine at the following levels: d-Amphetamine 750 ng/ml Barbiturates 300 ng/m! Benzodiazepines 300 ng/ml Benzoyl ecognine ) 225…
Read the full statement (219 words)
“Rapid Drug Screen” $-Panel low volume test is a one-step, lateral flow immunoassay for the simultaneous detection of eight abused substances in urine (each assay occupies a separate channel). “Rapid Drug Screen” 8-Panel low volume test is intended for use in the qualitative detection of the following 8 drugs of abuse in human urine at the following levels:
d-Amphetamine 750 ng/ml
Barbiturates 300 ng/m!
Benzodiazepines 300 ng/ml
Benzoyl ecognine ) 225 ng/ml
Cannabinoids
(11-nor-9-carboxy-delta-9-THC) 50 ng/ml
Methamphetamine 1000 ng/ml
Opiates (codeine) 225 ng/ml
(morphine-3-glucuronide) 225 ng/ml
Phencyclidine (PCP) 19 ng/ml
“Rapid Drug Screen” 8-Panel low volume test is intended for use by professional laboratories and physicians offices in a clinical setting. The assays are easy to perform, but should not be used without proper supervision. This immunoassay is a simplified qualitative screening method that provides only a preliminary analytical test result for use in determining the need for additional or confirmatory testing (i.e., gas chromatography/mass spectrometry (GC/MS)).
“Rapid -Drug Screen” 8-Panel low volume test provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a more confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.
Indications for use read by OCR from a scanned FDA 510(k) statement, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K981705 · printed Nov 6, 2026
Cite this page
510(k) Report. “K981705: 'Rapid Drug Screen' 8 Panel-Low Volume.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K981705
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Page last updated Nov 3, 2026